Skip to content

OsteoMed 450-0034Recalls & reported problems

Osteomed LLC · Lever Arm Motor Unit · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00845694015646), received between Feb 10, 2023 and Apr 30, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Power Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Power Problem6

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2024Power Problem; Power Problem

    A SERVICE REQUEST FOR THE DEVICE WAS RECEIVED WITH NO PRODUCT COMPLAINT MADE. ON 12 APRIL 2024, THE SERVICE REQUEST WORK ORDER DEVICE DIAGNOSIS WAS COMPLETED. DURING THE SERVICE REQUEST WORK ORDER, IT WAS NOTED "UNIT IS IS CONTINUOUSLY RUNNING DUE TO A MALFUNCTIONING FLEX CIRCUIT IN THE CABLE ASSEMBLY ". AS A RESULT OF THESE FINDINGS OF THE SERVICE REQUEST WORK ORDER, THIS REPORT IS BEING SUBMITTED. ADDITIONALLY, AS A RESULT OF THE SERVICE REQUEST WORK ORDER, THE CUSTOMER WAS CONTACTED TO DETERMINE IF ANY SURGERY OR PATIENT WAS IMPACTED. THE CUSTOMER RESPONDED THAT THIS ISSUE WAS NOTED IN THEIR WAREHOUSE PRIOR TO ANY SURGERY WITH NO PATIENT, USER, OR SURGERY INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19212762 ↗
  • MalfunctionApr 30, 2024Power Problem; Power Problem

    A SERVICE REQUEST FOR THE DEVICE WAS RECEIVED WITH NO PRODUCT COMPLAINT MADE. ON (B)(6) 2024, THE SERVICE REQUEST WORK ORDER DEVICE DIAGNOSIS WAS COMPLETED. DURING THE SERVICE REQUEST WORK ORDER, IT WAS NOTED "UNIT IS IS CONTINUOUSLY RUNNING DUE TO A MALFUNCTIONING FLEX CIRCUIT IN THE CABLE ASSEMBLY ". AS A RESULT OF THESE FINDINGS OF THE SERVICE REQUEST WORK ORDER, THIS REPORT IS BEING SUBMITTED. ADDITIONALLY, AS A RESULT OF THE SERVICE REQUEST WORK ORDER, THE CUSTOMER WAS CONTACTED TO DETERMINE IF ANY SURGERY OR PATIENT WAS IMPACTED. THE CUSTOMER RESPONDED THAT THIS ISSUE WAS NOTED IN THEIR WAREHOUSE PRIOR TO ANY SURGERY WITH NO PATIENT, USER, OR SURGERY INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19212678 ↗
  • MalfunctionJan 4, 2024Power Problem; Power Problem

    A SERVICE REQUEST FOR THE DEVICE WAS RECEIVED WITH NO PRODUCT COMPLAINT MADE. ON 12 DECEMBER 2023, THE SERVICE REQUEST WORK ORDER DEVICE DIAGNOSIS WAS COMPLETED. DURING THE SERVICE REQUEST WORK ORDER, IT WAS NOTED "UNIT IS DIFFICULT TO INSERT AND REMOVE FROM THE CONSOLE DUE TO DAMAGED CABLE CONNECTOR. MOTOR UNIT IS CONTINUOUSLY RUNNING DUE TO A MALFUNCTION FLEX CIRCUIT IN THE CABLE ASSEMBLY ". AS A RESULT OF THESE FINDINGS OF THE SERVICE REQUEST WORK ORDER, THIS REPORT IS BEING SUBMITTED. ADDITIONALLY, AS A RESULT OF THE SERVICE REQUEST WORK ORDER, THE CUSTOMER WAS CONTACTED TO DETERMINE IF ANY SURGERY OR PATIENT WAS IMPACTED. THE CUSTOMER RESPONDED THAT THIS ISSUE WAS NOTED IN THEIR WAREHOUSE PRIOR TO ANY SURGERY WITH NO PATIENT, USER, OR SURGERY INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18448136 ↗
  • MalfunctionOct 2, 2023Power Problem; Power Problem

    SERVICE REQUEST FOR THE DEVICE WAS RECEIVED WITH NO PRODUCT COMPLAINT MADE. ON (B)(6) 2023. THE SERVICE REQUEST WORK ORDER DEVICE DIAGNOSIS WAS COMPLETED. DURING THE SERVICE REQUEST WORK ORDER, IT WAS NOTED THE DEVICE WAS RUNNING CONTINUOUSLY DUE TO A MALFUNCTIONING CABLE ASSEMBLY, AND MOTOR UNIT ALSO HAS WORN AND DIRTY ROTOR ASSEMBLY". AS A RESULT OF THESE FINDINGS OF THE SERVICE REQUEST WORK ORDER, THIS REPORT IS BEING SUBMITTED. ADDITIONALLY, AS A RESULT OF THE SERVICE REQUEST WORK ORDER, THE CUSTOMER WAS CONTACTED TO DETERMINE IF ANY SURGERY OR PATIENT WAS IMPACTED. THE CUSTOMER RESPONDED THAT THIS ISSUE WAS NOTED IN THEIR WAREHOUSE POST SURGERY WITH NO PATIENT, USER, OR SURGERY INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17857713 ↗
  • MalfunctionApr 12, 2023Power Problem; Power Problem

    A SERVICE REQUEST FOR THE DEVICE WAS RECEIVED WITH NO PRODUCT COMPLAINT MADE. ON (B)(6) 2023, THE SERVICE REQUEST WORK ORDER DEVICE DIAGNOSIS WAS COMPLETED. DURING THE SERVICE REQUEST WORK ORDER, IT WAS NOTED THE DEVICE WAS RUNNING CONTINUOUSLY DUE TO A MALFUNCTIONING CABLE ASSEMBLY. AS A RESULT OF THESE FINDINGS OF THE SERVICE REQUEST WORK ORDER, THIS REPORT IS BEING SUBMITTED. ADDITIONALLY, AS A RESULT OF THE SERVICE REQUEST WORK ORDER, THE CUSTOMER IS BEING CONTACTED TO OBTAIN ADDITIONAL INFORMATION REGARDING THESE FINDINGS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16728879 ↗
  • MalfunctionFeb 10, 2023Power Problem; Power Problem

    A SERVICE REQUEST FOR THE DEVICE WAS RECEIVED WITH NO PRODUCT COMPLAINT MADE. ON 09 JANUARY 2023, THE SERVICE REQUEST WORK ORDER DEVICE DIAGNOSIS WAS COMPLETED. DURING THE SERVICE REQUEST WORK ORDER, IT WAS NOTED THE DEVICE WAS RUNNING CONTINUOUSLY DUE TO A MALFUNCTIONING FLEX CIRCUIT IN THE CABLE ASSEMBLY. AS A RESULT OF THESE FINDINGS OF THE SERVICE REQUEST WORK ORDER, THIS REPORT IS BEING SUBMITTED. ADDITIONALLY, AS A RESULT OF THE SERVICE REQUEST WORK ORDER, THE CUSTOMER IS BEING CONTACTED TO OBTAIN ADDITIONAL INFORMATION REGARDING THESE FINDINGS.

    Patient problems: Insufficient Information

    MAUDE report 16359149 ↗

Other HANDPIECE, ROTARY BONE CUTTING devices

All HANDPIECE, ROTARY BONE CUTTING devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.