S-11 surgical handpiece 1:1 30056000Recalls & reported problems
W & H Dentalwerk Bürmoos GmbH · Surgical Handpiece · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI J021300560001), received between Jun 27, 2023 and Feb 25, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Temperature Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Temperature Problem2
Patient problems
- Burn(s)1
- Full thickness (Third Degree) Burn1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 25, 2025Temperature Problem; Temperature Problem
DURING A SURGERY (WISDOM TOOTH) CORONORADICULAR SECTION WITH BULB BUR ON THE 48 TOOTH INCLUDED, A PATIENT WAS DEEPLY BURNED ON THE 3RD LEVEL (LOWER LIP). RISK OF PERMANENT INJURY WITH AESTHETIC DAMAGE FOR THE PATIENT. UPDATE ON 06/02/2025: THE SURGEON WHO PERFORMED THE SURGERY ON THE VICTIM CONFIRMED THAT THE INJURY IS REVERSIBLE. A COSMETIC SURGERY IS NECESSARY TO REMEDY THE INJURY AND THERE IS NO DOUBT THAT IT WILL COMPLETELY SCAR.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 21457856 ↗InjuryJun 27, 2023Temperature Problem; Temperature Problem
PATIENT WAS HAVING REMOVAL OF HER WISDOM TEETH, WHICH REQUIRES REMOVAL OF BONE AND SECTIONING OF TEETH WITH A HIGH-SPEED ROTARY INSTRUMENT. WE USE A SURGICAL STRAIGHT HANDPIECE WITH A MOTOR THAT RUNS AT STANDARD 50,000 RPM (MFR: W&H ELCOMED). A 702 CARBIDE BUR WAS USED WITH COPIOUS NORMAL SALINE IRRIGATION. I ALWAYS USE MY FINGERS TO HOLD THE LIP / CHEEK OUT TO NOT ONLY GET MORE LIGHT INTO THE FIELD, BUT ALSO TO AVOID TRAUMA FROM THE BUR. A FEW MINS AFTER SECTIONING TOOTH #47, MY ASSISTANT NOTICED A DEEP (3RD DEGREE) BURN TO THE LEFT LIP COMMISSURE, EVIDENTLY AS A RESULT OF THE SHAFT OF THE HANDPIECE HAVING CONTACT WITH TISSUES WHILE IN USE. THIS IS EXPECTED. WE USE SURGICAL HANDPIECES EVERYDAY, BUT I HAVE NEVER ENCOUNTERED A BURN LIKE THIS BEFORE. I DID NOT FEEL ANY HEAT GENERATED BY THE HANDPIECE ITSELF, AND WE DIDN'T ACTUALLY EVEN SEE THE BURN UNTIL A MIN OR 2 LATER, ALMOST AS IF IT WAS A DELAYED REACTION TO WHATEVER HEAT OR VIBRATION WAS BEING GENERATED BY THE HANDPIECE. TO CLARITY IT WAS NOT THE BUR, BUT RATHER THE SHAFT OF THE HANDPIECE VS, MOTOR ITSELF. I COULDN'T TELL WHICH WAS RESPONSIBLE. THE BURN WAS FULL THICKNESS INTO MUSCLE. THOUGH THERE WAS NOT EVEN ANY BLEEDING FROM
Patient problems: Burn(s)
MAUDE report 17212302 ↗
FDA clearance and classification
The FDA regulates product code KMW (handpiece, rotary bone cutting) under 21 CFR 872.4120 as a Class II device. 21 CFR 872.4120 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the S-11 surgical handpiece 1:1 30056000
Has the S-11 surgical handpiece 1:1 30056000 been recalled?
No FDA recall record in our data names this exact device identifier (DI J021300560001), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the S-11 surgical handpiece 1:1 30056000?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jun 27, 2023 and Feb 25, 2025. In those reports, the most common specific device problem is “Temperature Problem” (2); the most common patient problems are “Burn(s)” (1) and “Full thickness (Third Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the S-11 surgical handpiece 1:1 30056000?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the S-11 surgical handpiece 1:1 30056000, and is it still sold?
The FDA's device database lists W & H Dentalwerk Bürmoos GmbH as the company responsible for it (brand S-11 surgical handpiece 1:1, model 30056000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Handpiece, Rotary Bone Cutting devices
- OsteoMed 450-0241Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING9 reports
- Surgical Straight Handpiece SGS-E2SNakanishi Inc. · Handpiece, Rotary Bone Cutting8 reports
- OsteoMed 450-0034Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING6 reports
- OsteoMed 450-0390Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- OsteoMed 450-0021-00Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- S-12 surgical handpiece 1:2 30061000W & H Dentalwerk Bürmoos GmbH · Handpiece, Rotary Bone Cutting3 reports
- OsteoMed 450-0084Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- SGA-E2SNakanishi Inc. · HANDPIECE, ROTARY BONE CUTTING2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page