Skip to content

Tornier HRS ARS342001Recalls & reported problems

Tornier, Inc. · Spacer · Class 2 · Implant

All models of Tornier HRS (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00846832092581), received between Dec 18, 2023 and Jun 4, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Loss of Osseointegration1
  • Noise, Audible1
  • Osseointegration Problem1

Patient problems

  • Inadequate Osseointegration2
  • Failure of Implant1
  • Implant Pain1
  • Rheumatoid Arthritis1
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 4, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A PATIENT EXPERIENCED A DISLOCATION DUE TO A LOOSE PROXIMAL BODY OF THE REVIVE STEM, POTENTIALLY CAUSED BY A BACKED-OUT ASSEMBLY SCREW AND LOCKING CAP. REVISION SURGERY WAS PERFORMED, REPLACING ALL COMPONENTS EXCEPT FOR THE DISTAL STEM. ADDITIONAL IMPLANTS USED INCLUDED TWO 9X20 SPACERS, A LOW OFFSET HUMERAL TRAY, AND A 36+9 POLY INSERT.

    Patient problems: Failure of Implant

    MAUDE report 19464271 ↗
  • InjuryApr 22, 2024Osseointegration Problem; Osseointegration Problem

    REMOVAL OF FLEX REVIVE STEM AND WIRES AND REPLACEMENT WITH CEMENT SPACER DUE TO DISTAL HUMERAL (LIKELY) INFECTION AND PROSTHETIC LOOSENING. THE PATIENT HAS RA AND A BONE DENSITY DEFICIENCY. THE REVIVE STEM WAS USED TO STABILIZE THE DISPLACED DISTAL HUMERUS AND CABLES USED TO SECURE THE BONE. THERE WAS EVIDENCE OF LOOSENING OF THE THE DISTAL HUMERUS ONLY AND NO EVIDENCE ON XR OF ANY SHOULDER JOINT ISSUE.

    Patient problems: Inadequate Osseointegration; Rheumatoid Arthritis; Unspecified Infection

    MAUDE report 19159479 ↗
  • InjuryDec 18, 2023Loss of Osseointegration; Noise, Audible; Loss of Osseointegration; Noise, Audible

    AS REPORTED: "PATIENT HAD A REVIVE (IMPLANT) PUT IN ON (B)(6) 2023. PATIENT CAME BACK COMPLAINING OF PAIN AND A CLICKING. DR. SCHEDULED FOR A REVISION ON (B)(6) 2023 AND IT WAS DISCOVERED THAT THE ASSEMBLY SCREW HAD COME LOOSE AND THE PROXIMAL BODY WAS ROTATING".

    Patient problems: Implant Pain; Inadequate Osseointegration

    MAUDE report 18352813 ↗

Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices

All PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices →

Sources and updates