COMPREHENSIVE SHOULDER SYSTEM 113633Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00880304462632), received between Oct 28, 2019 and Jul 25, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
2 FDA recall records name this device's identifier.
5 other recalls cover devices cleared under the same FDA submission (K060692); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 7, 2022Terminated
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
Z-0676-2022 · Root cause: Process control · FDA record ↗
- Dec 8, 2017Terminated
Zimmer Biomet is conducting a medical device recall for two lots of Comprehensive Humeral stems. The lots were commingled during the manufacturing process and were etched with incorrect lot numbers. The size difference between the two stem sizes is 1.1 inches. It is visually recognizable by the user that the box label does not match the contents of the box.
Z-0979-2018 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Loosening of Implant Not Related to Bone-Ingrowth2
- Loss of or Failure to Bond1
- Separation Problem1
Patient problems
- Insufficient Information2
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
- No Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 25, 2025Separation Problem; Separation Problem
IT WAS REPORTED THAT THE PATIENT WAS INITIALLY IMPLANTED WITH A TOTAL SHOULDER ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION DUE TO UNKNOWN REASONS. DURING THE REVISION PROCEDURE, THE SURGEON NOTED THE HUMERAL TRAY WAS FRACTURED AND THE TAPER WOULD NOT DISENGAGE FROM THE STEM. THEREFORE, THE SURGEON REMOVED THE TAPER AND STEM. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22609764 ↗InjuryOct 3, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Insufficient Information
MAUDE report 17861405 ↗InjuryNov 15, 2022Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth; Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT UNDERWENT A LEFT SHOULDER REVISION SURGERY APPROXIMATELY ONE (1) WEEK AFTER UNDERGOING AN INITIAL REVISION SURGERY DUE TO AN UNKNOWN REASON.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 15797574 ↗InjuryOct 28, 2019Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Information
MAUDE report 9243769 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K060692. One submission often covers many models, so check the product and lot details in each record.
- Z-1543-2020 · Feb 17, 2020 · TerminatedComprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component o
- Z-0591-2019 · Oct 10, 2018 · TerminatedComprehensive Mini Stem, Item No. 113631:
- Z-0980-2018 · Dec 8, 2017 · TerminatedHumeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
- Z-0299-2018 · Jun 5, 2017 · TerminatedCOMP PRIMARY STEM 8MM MINI, model # 113628
- Z-1755-2015 · Apr 16, 2015 · TerminatedComprehensive Shoulder System, 4 mm, 122 mm long, Primary Shoulder Stem, Ti-6Al-4V, Standard Length Porous Plasma. Shoulder joint replacement components include humeral stems, hum
Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices
- DJO SURGICAL 509-02-032Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED364 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- TRIGEN 71118200Smith & Nephew, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED123 reports
- DJO SURGICAL 509-02-432Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED119 reports
- DELTA XTEND 130738206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED116 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
- GLOBAL UNITE 1100-12-100Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall29 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.