R SERIES 10000000001110012Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN,R SERIES,MAN-1,3/5 LD,STD ECG,AC PWR,NO SPO2,W/PACE,CF/USB, ENG, REFURB · Class 3
All models of R SERIES (94)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll R Series automated external defibrillator is a portable device used to deliver lifesaving shocks. The 28 reports mostly describe problems found during routine testing and maintenance: blank, distorted or unreadable screens, shock output out of specification, the unit failing its self-test, and being unable to read a heart rhythm through the electrode pads. None involved a patient, and no deaths or injuries were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00847946001964), received between Mar 21, 2017 and Sep 22, 2020 — classified as 28 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- No Display/Image10
- Defibrillation/Stimulation Problem6
- Failure of Device to Self-Test3
- Unable to Obtain Readings3
- Device Operational Issue1
- Display or Visual Feedback Problem1
- Energy Output Problem1
- Failure to Power Up1
Patient problems
- No Patient Involvement28
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10569701 ↗MalfunctionSep 16, 2020Output Problem; Output Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10541776 ↗MalfunctionApr 22, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY SHOWED COLORED LINES AND WAS NOT LEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9986256 ↗MalfunctionJan 24, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9628515 ↗MalfunctionDec 30, 2019Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9530515 ↗MalfunctionDec 5, 2019Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9419086 ↗MalfunctionDec 3, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9405956 ↗MalfunctionNov 22, 2019Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ECG FAULT 7" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9361706 ↗MalfunctionSep 27, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9130750 ↗MalfunctionAug 27, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY WAS DISTORTED AND NOT LEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8935238 ↗MalfunctionAug 27, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8934512 ↗MalfunctionAug 26, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8932059 ↗MalfunctionJun 28, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8743421 ↗MalfunctionJun 24, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "DEFIB FAULT 76" AND "DEFIB DISABLED MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8726519 ↗MalfunctionApr 23, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8540908 ↗MalfunctionMar 23, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY SHOWED VERTICAL LINES AND WAS NOT LEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8446580 ↗MalfunctionMar 14, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 76" AND "DEFIB DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8421416 ↗MalfunctionDec 10, 2018Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8147479 ↗MalfunctionNov 16, 2018Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8080677 ↗MalfunctionNov 8, 2018Display or Visual Feedback Problem; Failure of Device to Self-Test; Display or Visual Feedback Problem; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT ALSO INDICATES THE DEVICE FAILED SELF TEST. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8050940 ↗MalfunctionOct 30, 2018No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY SHOWED VERTICAL LINES AND WAS NOT LEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8018662 ↗MalfunctionMay 7, 2018No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY SHOWED HORIZONTAL LINES AND WAS NOT LEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7490016 ↗MalfunctionNov 7, 2017Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7006975 ↗MalfunctionJul 19, 2017Energy Output Problem; Energy Output Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6724320 ↗MalfunctionJun 14, 2017Improper Alarm; Improper Alarm
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION
Patient problems: No Patient Involvement
MAUDE report 6641652 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page