AED PLUS 20000000100011010Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN ASSY, PS SERIES, LCD, NO VOICE RCDG, ENGLISH · Class 3

All models of AED PLUS (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll AED Plus defibrillator is the subject of 1,062 adverse event reports, which mention 16 deaths and 3 injuries. Most reports describe the unit failing to power up or self-test, or showing a unit failed message, often during functional testing with no patient involved. Some also describe unexpected shutdowns and readings that could not be obtained. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,072
Deaths reported
16
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,072 adverse event reports that list this exact device (DI 00847946002152), received between Aug 3, 2015 and Sep 28, 2026 — classified as 16 deaths, 3 injuries, 1,053 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (590 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 1 reportsJan 2024: 10 reportsFeb 2024: 6 reportsMar 2024: 8 reportsApr 2024: 5 reportsMay 2024: 9 reportsJun 2024: 7 reportsJul 2024: 8 reportsAug 2024: 9 reportsSep 2024: 12 reportsOct 2024: 8 reportsNov 2024: 6 reportsDec 2024: 7 reportsJan 2025: 8 reportsFeb 2025: 5 reportsMar 2025: 6 reportsApr 2025: 7 reportsMay 2025: 4 reportsJun 2025: 5 reportsJul 2025: 14 reportsAug 2025: 5 reportsSep 2025: 6 reportsOct 2025: 14 reportsNov 2025: 8 reportsDec 2025: 7 reportsJan 2026: 5 reportsFeb 2026: 5 reportsMar 2026: 8 reportsApr 2026: 8 reportsMay 2026: 4 reportsJun 2026: 10 reportsJul 2026: 1 reportsAug 2026: 13 reportsSep 2026: 12 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Power Up590
  • Failure of Device to Self-Test290
  • Incorrect Interpretation of Signal33
  • Unable to Obtain Readings23
  • Unexpected Shutdown23
  • Failure to Cycle17
  • Use of Device Problem16
  • Failure to Analyze Signal13

Patient problems

  • No Patient Involvement507
  • No Clinical Signs, Symptoms or Conditions477
  • Insufficient Information32
  • No Information22
  • No Consequences Or Impact To Patient15
  • Necrosis7
  • Death4
  • Cardiac Arrest2

Recent adverse event reports

The 25 most recent of 1,072 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915840 ↗
  • MalfunctionSep 24, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882701 ↗
  • MalfunctionSep 22, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50855322 ↗
  • MalfunctionSep 21, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50839852 ↗
  • MalfunctionSep 10, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50714000 ↗
  • MalfunctionSep 8, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50669918 ↗
  • MalfunctionSep 8, 2026Failure of Device to Self-Test

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50669916 ↗
  • MalfunctionSep 4, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50612330 ↗
  • MalfunctionSep 2, 2026Failure to Power Up; Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50567603 ↗
  • MalfunctionSep 2, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50567504 ↗
  • MalfunctionSep 2, 2026Failure of Device to Self-Test

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50566396 ↗
  • MalfunctionSep 1, 2026Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50541936 ↗
  • MalfunctionAug 28, 2026Incorrect Interpretation of Signal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50498482 ↗
  • MalfunctionAug 26, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50453588 ↗
  • MalfunctionAug 25, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50432492 ↗
  • MalfunctionAug 20, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50349973 ↗
  • MalfunctionAug 20, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50346235 ↗
  • MalfunctionAug 17, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50275278 ↗
  • MalfunctionAug 17, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50275047 ↗
  • MalfunctionAug 14, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50254678 ↗
  • MalfunctionAug 11, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50208871 ↗
  • MalfunctionAug 10, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50193768 ↗
  • MalfunctionAug 10, 2026Failure of Device to Self-Test

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50192331 ↗
  • MalfunctionAug 4, 2026Incorrect Interpretation of Signal

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE ISSUED AN "ADVISE TO SHOCK PATIENT WITH RIGOR" PROMPT.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 50133581 ↗
  • MalfunctionAug 3, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50114112 ↗