R SERIES 30000000301110013Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,R SERIES, MAIN, PLUS, 3/5 LD, AC PWR, STD ECG, PACE, ETCO2, CF/USB, DMST · Class 3
All models of R SERIES (94)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll R Series defibrillator is a non-wearable automated external defibrillator used for shock and pacing. Most of the 28 reports describe problems found during routine checks or biomedical testing rather than during patient care: units powering down unexpectedly, failing to read the heartbeat through the pads, shock output out of specification, and fault messages on the display. Reports describe no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00847946002381), received between Mar 29, 2017 and Oct 13, 2020 — classified as 28 malfunctions.
The most frequently reported specific device problem is “Output below Specifications” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Output below Specifications4
- Grounding Malfunction3
- Unexpected Shutdown3
- Defibrillation/Stimulation Problem2
- Failure to Deliver Shock/Stimulation2
- Failure to Power Up2
- No Display/Image2
- Unable to Obtain Readings2
Patient problems
- No Patient Involvement24
- No Consequences Or Impact To Patient2
- No Information2
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 13, 2020Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10670988 ↗MalfunctionOct 12, 2020Output below Specifications; Output below Specifications
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIBER OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10662881 ↗MalfunctionOct 9, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10656297 ↗MalfunctionOct 9, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10656229 ↗MalfunctionSep 21, 2020Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "DEFIB FAULT 72", DEFIB FAULT 76" AND "PACER FAULT 115" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10563592 ↗MalfunctionJun 26, 2020Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INTERMITTENTLY FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10201099 ↗MalfunctionMay 22, 2020Output below Specifications; Output below Specifications
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10082859 ↗MalfunctionApr 30, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 72" AND "DEFIB FAULT 76" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10014315 ↗MalfunctionApr 8, 2020Output below Specifications; Output below Specifications
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9940483 ↗MalfunctionApr 8, 2020Output below Specifications; Output below Specifications
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9940456 ↗MalfunctionMar 23, 2020High impedance; High impedance
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S PATIENT IMPEDANCE WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9866420 ↗MalfunctionMar 13, 2020Output Problem; Output Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9832385 ↗MalfunctionFeb 7, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9683876 ↗MalfunctionNov 23, 2019Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INAPPROPRIATELY SHUT DOWN.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9366828 ↗MalfunctionOct 24, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9235135 ↗MalfunctionOct 22, 2019Use of Device Problem; Use of Device Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ERROR" MESSAGE. NO FURTHER INFORMATION WAS AVAILABLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9221397 ↗MalfunctionAug 7, 2019Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ECG FFFF" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8870825 ↗MalfunctionApr 22, 2019Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8537703 ↗MalfunctionNov 29, 2018Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8115503 ↗MalfunctionNov 15, 2018Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A MALE PATIENT (AGE UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8074937 ↗MalfunctionOct 28, 2018Incorrect Interpretation of Signal; Incorrect Interpretation of Signal
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE ISSUED A "SHOCK ADVISED" PROMPT FOR A HEART RHYTHM THEY BELIEVED WAS NON-SHOCKABLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 8011430 ↗MalfunctionMay 11, 2018Failure to Discharge; Failure to Discharge
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7508396 ↗MalfunctionOct 25, 2017No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6975968 ↗MalfunctionOct 20, 2017Device Operational Issue; Device Operational Issue
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT IN CARDIAC ARREST (AGE & GENDER UNKNOWN), THE ONE STEP COMPLETE PADS WOULD NOT CONNECT TO THE DEFIBRILLATOR'S ASSOCIATED ONE STEP PACING CABLE. COMPLAINANT INDICATED THAT THE CLINICIAN SWITCHED TO A SET OF STAT PADS TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 6966551 ↗MalfunctionJul 25, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "CHECK PADS" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6741053 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page