R SERIES 30000005201310013Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN,MAN-1,PLUS,3/5 LD,STD ECG,AC PWR,SPO2,ETCO2,PACE,NIBPCF/USB,R SERIES,DMST · Class 3
All models of R SERIES (94)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll R Series defibrillator is an automated external defibrillator. Many of the 126 reports describe the device failing to power up, showing no display, or displaying incorrect messages during routine checks, and most say no patient was involved. The reports include no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 126
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 126 adverse event reports that list this exact device (DI 00847946002695), received between Aug 7, 2015 and Oct 14, 2020 — classified as 126 malfunctions.
The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (27 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Defibrillation/Stimulation Problem27
- No Display/Image19
- Device Displays Incorrect Message15
- Unable to Obtain Readings14
- Failure to Power Up11
- Failure of Device to Self-Test8
- Failure to Deliver Shock/Stimulation4
- Grounding Malfunction4
Patient problems
- No Patient Involvement121
- No Consequences Or Impact To Patient4
- No Information1
Recent adverse event reports
The 25 most recent of 126 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 14, 2020Unable to Obtain Readings; Defibrillation/Stimulation Problem; Unable to Obtain Readings; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "CHECK PADS" AND "DEFIB PAD SHORT" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10678976 ↗MalfunctionOct 13, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 72" AND "DEFIB DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10673008 ↗MalfunctionOct 7, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10644713 ↗MalfunctionOct 2, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10620294 ↗MalfunctionSep 21, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DEVICE'S DISPLAY BLANKED OUT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10560533 ↗MalfunctionSep 3, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY SHOWED HORIZONTAL LINES AND WAS NOT LEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10490433 ↗MalfunctionAug 24, 2020Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10444473 ↗MalfunctionAug 12, 2020Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10397211 ↗MalfunctionAug 10, 2020Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10387539 ↗MalfunctionAug 7, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10382410 ↗MalfunctionAug 7, 2020Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ECG DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10380230 ↗MalfunctionJul 31, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10351157 ↗MalfunctionJul 29, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY WAS BLURRY AND CLINICAL DATA WAS UNREADABLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10338421 ↗MalfunctionJul 27, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10325905 ↗MalfunctionJul 22, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10309108 ↗MalfunctionJul 20, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10297200 ↗MalfunctionJul 6, 2020Insufficient Cooling; Insufficient Cooling
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO DISCHARGE DURING SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10234499 ↗MalfunctionJul 2, 2020High impedance; High impedance
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE PATIENT IMPEDANCE WAS INCORRECT. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10224252 ↗MalfunctionJun 11, 2020Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10143941 ↗MalfunctionJun 10, 2020Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10140241 ↗MalfunctionJun 10, 2020Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10138809 ↗MalfunctionJun 1, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10105953 ↗MalfunctionMay 22, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10082048 ↗MalfunctionMay 20, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "DEFIB FAULT 72" AND "DEFIB FAULT 76" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10076499 ↗MalfunctionMay 14, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "SYSTEM FAULT 36" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10058887 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page