R SERIES 30000006001310012Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN, R SERIES,NELLCOR, 3/5 LD, STD ECG, AC PWR, PACE, SPO2, NIBP, CF/USB, DMST · Class 3

All models of R SERIES (94)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll R Series defibrillator is a non-wearable automated external defibrillator. Reports mostly describe devices that would not power up, showed distorted or unreadable screens, or gave pad and defibrillation error messages during routine checks or testing. There are 23 reports that name its identifier, with no deaths and no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00847946002725), received between Mar 29, 2017 and Jun 22, 2020 — classified as 23 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message5
  • Failure to Power Up5
  • No Display/Image5
  • Defibrillation/Stimulation Problem3
  • Failure of Device to Self-Test2
  • Unable to Obtain Readings2
  • Display or Visual Feedback Problem1
  • Erratic or Intermittent Display1

Patient problems

  • No Patient Involvement23

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 22, 2020No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10179929 ↗
  • MalfunctionJun 9, 2020Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "POOR PAD CONTACT" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10134338 ↗
  • MalfunctionJun 3, 2020Failure to Power Up; Failure of Device to Self-Test; Failure to Power Up; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF-TEST FOR PACER FUNCTION AND WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10116141 ↗
  • MalfunctionFeb 11, 2020Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9695213 ↗
  • MalfunctionFeb 3, 2020No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY INTERMITTENTLY SHOWED VERTICAL COLORED LINES AND WAS NOT LEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9660999 ↗
  • MalfunctionJan 8, 2020Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "CHECK PADS" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9563617 ↗
  • MalfunctionDec 27, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 76" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9525459 ↗
  • MalfunctionNov 6, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "DEFIB DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT AND WAS ABLE TO SHOCK THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9285931 ↗
  • MalfunctionAug 22, 2019No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY GREY AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8922717 ↗
  • MalfunctionMar 22, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8445392 ↗
  • MalfunctionOct 25, 2018Display or Visual Feedback Problem; Display or Visual Feedback Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY WAS DISTORTED AND WAS NOT LEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8005598 ↗
  • MalfunctionJul 26, 2018Erratic or Intermittent Display; Erratic or Intermittent Display

    COMPLAINANT ALLEGED THAT A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7722093 ↗
  • MalfunctionMay 31, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7557371 ↗
  • MalfunctionMar 16, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INTERMITTENTLY DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7346033 ↗
  • MalfunctionFeb 13, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO POWER ON. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7267148 ↗
  • MalfunctionJan 3, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7159557 ↗
  • MalfunctionNov 13, 2017Device Displays Incorrect Message; Device Displays Incorrect Message

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "CHECK PADS" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7024117 ↗
  • MalfunctionOct 16, 2017Device Displays Incorrect Message; Device Displays Incorrect Message

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "DEFIB FAULT 72" AND "DEFIB FAULT 76" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6952170 ↗
  • MalfunctionOct 10, 2017No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6937257 ↗
  • MalfunctionSep 15, 2017Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST WHEN USING EXTERNAL PADDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6870298 ↗
  • MalfunctionAug 7, 2017Device Displays Incorrect Message; Device Displays Incorrect Message

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "SYSTEM FAULT 36" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6770665 ↗
  • MalfunctionJul 25, 2017No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED WHITE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6740993 ↗
  • MalfunctionMar 29, 2017Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6443100 ↗