M SERIES 40010211100103010Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN BSC-4 BIPH/SPO2/AC/PC/EL/RC/SM/CD · Class 3
All models of M SERIES (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00847946003609), received between Mar 27, 2017 and Oct 13, 2020 — classified as 18 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message4
- No Display/Image4
- Defibrillation/Stimulation Problem3
- Failure of Device to Self-Test2
- Failure to Power Up2
- Energy Output Problem1
- Output below Specifications1
- Pacing Problem1
Patient problems
- No Patient Involvement15
- No Information2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 13, 2020Output below Specifications; Output below Specifications
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S PACER OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10671688 ↗MalfunctionJul 21, 2020Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "PACER FAULT 116", "DEFIB DISABLED", AND "DEFIB FAULT 72" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10303018 ↗MalfunctionFeb 25, 2020Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "DEFIB PAD SHORT" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 9752207 ↗MalfunctionJul 13, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8787405 ↗MalfunctionApr 11, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8504874 ↗MalfunctionNov 16, 2018No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE'S DISPLAY WAS DISTORTED AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8077248 ↗MalfunctionNov 1, 2018Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "PACER FAULT 116" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8029176 ↗MalfunctionSep 13, 2018Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "POOR PAD CONTACT" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 7873691 ↗MalfunctionApr 23, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7452256 ↗MalfunctionFeb 9, 2018Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7258669 ↗MalfunctionNov 6, 2017No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7002697 ↗MalfunctionOct 18, 2017Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6957712 ↗MalfunctionSep 5, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6843198 ↗MalfunctionApr 21, 2017Energy Output Problem; Energy Output Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DEFIB OUTPUT WAS INCORRECT VIA EXTERNAL PADDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6510997 ↗MalfunctionApr 14, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6493232 ↗MalfunctionApr 13, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6489624 ↗MalfunctionMar 31, 2017No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAY WAS MISSING THE ECG RHYTHM. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6450306 ↗MalfunctionMar 27, 2017Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS INTERMITTENTLY UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6437396 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page