M SERIES 40010211100123010Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN, MNL-1 BIPH/SPO2/AC/PC/EL/RC/SM/CDM · Class 3
All models of M SERIES (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll M Series defibrillator is a non-wearable automated external defibrillator used to shock and pace patients in emergencies. The 75 reports mostly describe problems found during routine testing and biomedical checks — units that would not power up, would not deliver a shock, could not be set to the right energy level, or showed error messages on the screen. One report mentions a death; no injuries were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 75
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 75 adverse event reports that list this exact device (DI 00847946003623), received between Jul 31, 2015 and Dec 22, 2021 — classified as 1 death, 74 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure to Power Up17
- Device Displays Incorrect Message10
- Defibrillation/Stimulation Problem9
- Failure to Discharge6
- Failure to Select Signal5
- Grounding Malfunction5
- Pacing Problem4
- Failure of Device to Self-Test3
Patient problems
- No Patient Involvement66
- No Information4
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
- Death1
- Not Applicable1
Recent adverse event reports
The 25 most recent of 75 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13061792 ↗MalfunctionDec 20, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO DISCHARGE VIA EXTERNAL PADDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT NUMBER (B)(4) FOR A SIMILAR REPORT FROM THE SAME CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13047724 ↗MalfunctionApr 26, 2021Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 72", "DEFIB DISABLED", AND "PACER DISABLED" MESSAGES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11728683 ↗MalfunctionJul 14, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10270801 ↗MalfunctionJun 1, 2020Pacing Problem; Pacing Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S PACER RATE COULD NOT BE PROPERLY ADJUSTED. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10106211 ↗MalfunctionApr 8, 2020Failure to Select Signal; Failure to Select Signal
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO SELECT ENERGY LEVEL. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9940537 ↗MalfunctionApr 8, 2020Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9935761 ↗MalfunctionApr 8, 2020Failure to Charge; Failure to Select Signal; Failure to Charge; Failure to Select Signal
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO CHARGE AND WAS UNABLE TO SELECT ENERGY LEVEL. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9934055 ↗MalfunctionMar 23, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9864741 ↗MalfunctionMar 10, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9812531 ↗MalfunctionFeb 3, 2020Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED THE EARTH GROUND RESISTANCE TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9661322 ↗MalfunctionJan 10, 2020Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "DEFIB DISABLED" MESSAGE AND "PACER DISABLED" MESSAGES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9576026 ↗MalfunctionDec 20, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9505767 ↗MalfunctionDec 13, 2019Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED THE EARTH GROUND RESISTANCE TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9466028 ↗MalfunctionNov 27, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9384046 ↗MalfunctionNov 13, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9316652 ↗MalfunctionSep 23, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9107242 ↗MalfunctionJun 6, 2019Grounding Malfunction; Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED THE EARTH GROUND RESISTANCE TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8675088 ↗MalfunctionMay 15, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING INCOMING TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY YELLOW AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8613165 ↗MalfunctionMay 3, 2019Use of Device Problem; Use of Device Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S SELECTOR SWITCH WAS NOT FUNCTIONING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8579338 ↗MalfunctionApr 17, 2019Failure to Select Signal; Failure to Select Signal
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO SELECT ENERGY LEVEL. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8522952 ↗MalfunctionApr 16, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8520181 ↗MalfunctionApr 9, 2019Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8497287 ↗MalfunctionMar 29, 2019Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "POOR PAD CONTACT" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8464415 ↗MalfunctionMar 14, 2019Failure to Charge; Failure to Charge
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO CHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8421431 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record