X SERIES 610-2231011-01Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN ASSY, X SERIES MONITOR/DEFIBRILLATOR 12 LEAD/W INTERP, PACING, NIBP, SPO2, SPCO, CPR EXPANSION PACK AND ETCO2, W/CP2, DMST · Class 3
All models of X SERIES (163)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll X Series monitor and defibrillator appears in 641 adverse event reports, including 28 deaths and no injuries. Many reports describe screens going blank or white, units failing to power on or self-test, and sudden shutdowns. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 644
- Deaths reported
- 28
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 644 adverse event reports that list this exact device (DI 00847946004354), received between Aug 4, 2015 and Sep 29, 2026 — classified as 28 deaths, 616 malfunctions.
The most frequently reported specific device problem is “Unable to Obtain Readings” (143 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Unable to Obtain Readings143
- Failure to Power Up114
- Failure of Device to Self-Test77
- No Display/Image76
- Unexpected Shutdown43
- Defibrillation/Stimulation Problem36
- Reset Problem34
- Failure to Deliver Shock/Stimulation31
Patient problems
- No Clinical Signs, Symptoms or Conditions261
- No Patient Involvement182
- No Information59
- Insufficient Information51
- No Consequences Or Impact To Patient48
- Death17
- Cardiac Arrest13
- Necrosis6
Recent adverse event reports
The 25 most recent of 644 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Unable to Obtain Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933453 ↗MalfunctionSep 22, 2026Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP.\\COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853979 ↗MalfunctionSep 16, 2026Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING TRAINING BY ZOLL STAFF, THE DEVICE FAILED SELF-TEST FOR DEFIB FUNCTION.\\COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50785443 ↗MalfunctionAug 17, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50286708 ↗MalfunctionAug 10, 2026Smoking
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE SMOKED.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50192836 ↗MalfunctionJun 23, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25600243 ↗MalfunctionJun 18, 2026Unable to Obtain Readings
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A 30-YEAR-OLD MALE PATIENT, THE DEVICE DISPLAYED "ECG DISABLED" AND "ECG MONITORING FAILURE" MESSAGES. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25547257 ↗MalfunctionJun 15, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25502924 ↗MalfunctionJun 8, 2026Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A FEMALE PATIENT (AGE & UNKNOWN), THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25429775 ↗MalfunctionMay 28, 2026Pacing Problem
COMPLAINANT ALLEGED WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WAS UNABLE TO PACE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25318718 ↗MalfunctionMay 1, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25058091 ↗MalfunctionApr 24, 2026Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24981648 ↗MalfunctionMar 11, 2026Failure to Charge
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT AN 88-YEAR-OLD MALE PATIENT, THE DEVICE FAILED TO CHARGE TO SET ENERGY. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 24575158 ↗MalfunctionMar 6, 2026Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24532842 ↗MalfunctionMar 3, 2026Unexpected Shutdown
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A 57-YEAR-OLD MALE PATIENT, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 24500720 ↗MalfunctionFeb 4, 2026Failure to Charge
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED TO CHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24256807 ↗MalfunctionFeb 2, 2026Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST FOR PACER FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24234565 ↗MalfunctionJan 16, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24096720 ↗MalfunctionJan 9, 2026Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A 75-YEAR-OLD FEMALE PATIENT, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24020814 ↗MalfunctionJan 6, 2026Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23974747 ↗MalfunctionDec 23, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICES SCREEN TURNED COMPLETELY WHITE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23876307 ↗MalfunctionDec 23, 2025Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23876167 ↗MalfunctionNov 7, 2025Failure to Analyze Signal; Failure to Analyze Signal
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE EXPERIENCED AN ECG MONITORING FAILURE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23497978 ↗MalfunctionNov 6, 2025Reset Problem; Reset Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE REBOOT/RESTART MULTIPLE TIMES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23486709 ↗MalfunctionOct 21, 2025Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA THE MULTIFUNCTION CABLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23348578 ↗
FDA clearance and classification
The FDA regulates product code MKJ (automated external defibrillators (non-wearable)) under 21 CFR 870.5310 as a Class III device. 21 CFR 870.5310 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
FDA definition of this device type: This device is a non-wearable prescription use only aed. These are devices that include automated external defibrillation. Automated external defibrillators use external pad-type electrodes to sense, detect, classify and treat (with an electrical shock) ventricular fibrillation. These devices are intended to be used on suspected victims of sudden cardiac arrest. A person in cardiac arrest is unresponsive and is not breathing normally. The device can be sold with prescription only.
Common questions about the X SERIES 610-2231011-01
Has the X SERIES 610-2231011-01 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00847946004354), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the X SERIES 610-2231011-01?
The FDA's MAUDE database lists 644 adverse event reports for this device, received between Aug 4, 2015 and Sep 29, 2026. In those reports, the most common specific device problems are “Unable to Obtain Readings” (143), “Failure to Power Up” (114) and “Failure of Device to Self-Test” (77); the most common patient problems are “Death” (17), “Cardiac Arrest” (13) and “Necrosis” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the X SERIES 610-2231011-01?
28 reports are classified as deaths, 0 as injuries and 616 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the X SERIES 610-2231011-01 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the X SERIES 610-2231011-01, and is it still sold?
The FDA's device database lists Zoll Medical Corporation as the company responsible for it (brand X SERIES, model 610-2231011-01). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,163 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,012 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,384 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,719 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,582 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page