Skip to content

M SERIES CCT 40010811100124010Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN CCT,MAN-1,3/5 LEAD,SPO2,NIBP,IBP,TEMP,BIPHSC,PACE,RCD,SMR,CDMK,AC,DMST · Class 2

All models of M SERIES CCT (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll M Series defibrillator is a portable monitor with built-in blood-pressure, oxygen and temperature readings. The 72 reports mostly describe the unit not powering up, not responding to controls, or showing error and blank-screen messages, often found during routine testing or service checks rather than during patient care. Three reports mention deaths; in each, the reporters stated the patient outcome was not related to the device. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
72
Deaths reported
3
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 72 adverse event reports that list this exact device (DI 00847946004811), received between Jul 31, 2015 and Nov 15, 2022 — classified as 3 deaths, 69 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Power Up14
  • Defibrillation/Stimulation Problem8
  • Device Operational Issue8
  • No Display/Image7
  • Activation Problem5
  • Grounding Malfunction5
  • Energy Output Problem3
  • Failure of Device to Self-Test3

Patient problems

  • No Patient Involvement61
  • No Information4
  • Death2
  • No Consequences Or Impact To Patient2
  • Necrosis1
  • No Known Impact Or Consequence To Patient1
  • Premature Labor1

Recent adverse event reports

The 25 most recent of 72 reports, in the reporter's own words as published by the FDA.

  • DeathNov 15, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A 45-YEAR-OLD MALE PATIENT, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED; HOWEVER, THEY DO NOT BELIEVE THIS WAS DEVICE ASSOCIATED.

    Patient problems: Necrosis

    MAUDE report 15799627 ↗
  • MalfunctionJun 25, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "SYSTEM FAULT 36" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10196847 ↗
  • MalfunctionJun 1, 2020Activation Problem; Activation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10105947 ↗
  • MalfunctionMay 15, 2020Activation Problem; Activation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10062154 ↗
  • MalfunctionMay 14, 2020Activation Problem; Activation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10059256 ↗
  • MalfunctionApr 21, 2020Failure to Charge; Failure to Charge

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE INTERMITTENTLY FAILED TO CHARGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9980834 ↗
  • MalfunctionApr 17, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9973554 ↗
  • MalfunctionMar 5, 2020Pacing Problem; Pacing Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "PACER FAULT 122", "PACER FAULT 123" AND "PACER FAULT 126" ERROR MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9794956 ↗
  • MalfunctionFeb 14, 2020Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9711476 ↗
  • MalfunctionJan 6, 2020No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE HAS AN UNKNOWN INTERMITTENT DISPLAY PROBLEM. NO FURTHER INFORMATION WAS AVAILABLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9555444 ↗
  • MalfunctionDec 23, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DISCHARGE FAULT" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9513146 ↗
  • MalfunctionOct 16, 2019Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9198290 ↗
  • MalfunctionOct 10, 2019Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9177475 ↗
  • MalfunctionSep 30, 2019Activation Problem; Activation Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9135071 ↗
  • MalfunctionSep 11, 2019Use of Device Problem; Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO SELECT ENERGY LEVEL. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8995800 ↗
  • MalfunctionAug 27, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8934791 ↗
  • MalfunctionJun 27, 2019Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8739855 ↗
  • MalfunctionJun 20, 2019Pacing Problem; Failure to Power Up; Pacing Problem; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO PACE AND WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8718458 ↗
  • MalfunctionMay 17, 2019Use of Device Problem; Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED UP IN THE INCORRECT MODE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8620686 ↗
  • MalfunctionMay 14, 2019Activation Problem; Activation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8609422 ↗
  • MalfunctionApr 29, 2019Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) DURING CARDIOVERSION, THE DEVICE FAILED TO SHOCK THE PATIENT. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8560243 ↗
  • MalfunctionApr 29, 2019Failure to Select Signal; Failure to Select Signal

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT RESPOND TO INPUT SELECTION VIA THE FRONT PANEL MEMBRANE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: Premature Labor

    MAUDE report 8560128 ↗
  • MalfunctionApr 25, 2019Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8550882 ↗
  • MalfunctionApr 11, 2019Use of Device Problem; Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8505957 ↗
  • DeathApr 10, 2019Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "DEFIB PAD SHORT" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED.

    Patient problems: Death

    MAUDE report 8501570 ↗

Other COMPUTER, BLOOD-PRESSURE devices

All COMPUTER, BLOOD-PRESSURE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page