M SERIES CCT 40021911100124010Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN,CCT,MAN-1,12 LEAD,SPO2,ETCO2,NIBP,IBP,TEMP,BIP,PACE,RCD,SMR,CDMK,AC,DMST · Class 2
All models of M SERIES CCT (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00847946006501), received between May 11, 2017 and Jan 17, 2020 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure to Power Up5
- No Display/Image4
- Device Displays Incorrect Message3
- Activation Problem1
- Defibrillation/Stimulation Problem1
- Failure to Discharge1
- Loss of Power1
- Output Problem1
Patient problems
- No Patient Involvement14
- No Information2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 17, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9600927 ↗MalfunctionDec 12, 2019No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY BLANKED OUT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9457509 ↗MalfunctionMay 23, 2019Activation Problem; Activation Problem
COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8635911 ↗MalfunctionFeb 21, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8358246 ↗MalfunctionFeb 14, 2019Output Problem; Output Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO ADJUST PACER RATE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8338929 ↗MalfunctionNov 12, 2018Failure to Power Up; Unexpected Shutdown; Failure to Power Up; Unexpected Shutdown
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) WHILE TRANSFERRING THE PATIENT FROM THE AIRCRAFT TO THE AMBULANCE, THE DEVICE INAPPROPRIATELY SHUT DOWN AND WOULD NOT POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8061026 ↗MalfunctionJul 12, 2018Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "SYSTEM FAULT 36", "PACER FAULT 115" AND "PACER DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7682713 ↗MalfunctionMar 14, 2018No Display/Image; Device Displays Incorrect Message; No Display/Image; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE PROMPTED A "CHECK PADS" MESSAGE AND POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7338527 ↗MalfunctionFeb 7, 2018Loss of Power; Loss of Power
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE AND GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. COMPLAINANT INDICATED THAT DURING SUBSEQUENT BIOMED TESTING, THE MALFUNCTION WAS DUPLICATED.
Patient problems: No Information
MAUDE report 7248229 ↗MalfunctionJan 18, 2018Failure to Discharge; Failure to Discharge
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION. COMPLAINANT INDICATED THAT SUBSEQUENT TESTING DID NOT DUPLICATE THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7203260 ↗MalfunctionDec 12, 2017Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO OBTAIN ECG SIGNAL. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7109604 ↗MalfunctionOct 30, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6988153 ↗MalfunctionOct 5, 2017No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE'S DISPLAY INTERMITTENTLY BLANKED OUT. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 6917047 ↗MalfunctionJul 25, 2017Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6741001 ↗MalfunctionJul 18, 2017Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT WHILE FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6721449 ↗MalfunctionMay 16, 2017Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6570977 ↗MalfunctionMay 11, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6559927 ↗
Other COMPUTER, BLOOD-PRESSURE devices
- M SERIES CCT 40021911100124300Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE79 reports
- M SERIES CCT 40010811100124010Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE3 deaths72 reports
- AVVIGO™+ H7492493121C0Boston Scientific Corporation · COMPUTER, BLOOD-PRESSURE1 recall0 reports
- SureSigns VS2+ 863278Philips Medical Systems Hsg · COMPUTER, BLOOD-PRESSURE40 reports
- SureSigns VSi 863276Philips Medical Systems Hsg · COMPUTER, BLOOD-PRESSURE21 reports
- SureSigns VS2+ 863279Philips Medical Systems Hsg · COMPUTER, BLOOD-PRESSURE16 reports
- M SERIES CCT 40010911100124010Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE8 reports
- M SERIES CCT 60021911100124010Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page