The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00847946004828), received between Oct 25, 2017 and Oct 9, 2020 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Activation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Activation Problem1
- Defibrillation/Stimulation Problem1
- Failure of Device to Self-Test1
- Loss of Power1
- No Display/Image1
- Use of Device Problem1
Patient problems
- No Patient Involvement4
- No Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 9, 2020No Display/Image; Activation Problem; No Display/Image; Activation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY FLICKERED AND DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10655837 ↗MalfunctionFeb 21, 2020Use of Device Problem; Use of Device Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED UP IN THE INCORRECT MODE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9738883 ↗MalfunctionSep 10, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8983369 ↗MalfunctionApr 9, 2018Loss of Power; Loss of Power
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 7406121 ↗MalfunctionOct 25, 2017Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INTERMITTENTLY FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 6975985 ↗
Other COMPUTER, BLOOD-PRESSURE devices
- M SERIES CCT 40021911100124300Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE79 reports
- M SERIES CCT 40010811100124010Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE3 deaths72 reports
- AVVIGO™+ H7492493121C0Boston Scientific Corporation · COMPUTER, BLOOD-PRESSURE1 recall0 reports
- SureSigns VS2+ 863278Philips Medical Systems Hsg · COMPUTER, BLOOD-PRESSURE40 reports
- SureSigns VSi 863276Philips Medical Systems Hsg · COMPUTER, BLOOD-PRESSURE21 reports
- M SERIES CCT 40021911100124010Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE17 reports
- SureSigns VS2+ 863279Philips Medical Systems Hsg · COMPUTER, BLOOD-PRESSURE16 reports
- M SERIES CCT 40010911100124010Zoll Medical Corporation · COMPUTER, BLOOD-PRESSURE8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page