M SERIES CCTRecalls & reported problems across all models
Zoll Medical Corporation · 10 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 191
- Deaths reported
- 3
- Injuries reported
- 0
- Recalls
- 0
We found 10 models of M SERIES CCT by Zoll Medical Corporation in the FDA's device registry (GUDID). Together they are named in 191 adverse event reports in the FDA's MAUDE database, received between Jul 31, 2015 and Nov 15, 2022 — classified as 3 deaths, 188 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (48 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Most reported M SERIES CCT problems
Device problems
- Failure to Power Up48
- Defibrillation/Stimulation Problem28
- No Display/Image24
- Device Displays Incorrect Message13
- Device Operational Issue12
- Pacing Problem12
- Activation Problem11
- Unable to Obtain Readings7
Patient problems
- No Patient Involvement175
- No Information8
- No Consequences Or Impact To Patient3
- Death2
- Necrosis1
- No Known Impact Or Consequence To Patient1
- Premature Labor1
M SERIES CCT models
10 models, most reported first. Each model has its own page with its reports, recalls and device details.
- M SERIES CCT 40021911100124300COMPUTER, BLOOD-PRESSURE79 reports
- M SERIES CCT 40010811100124010COMPUTER, BLOOD-PRESSURE3 deaths72 reports
- M SERIES CCT 40021911100124010COMPUTER, BLOOD-PRESSURE17 reports
- M SERIES CCT 40010911100124010COMPUTER, BLOOD-PRESSURE8 reports
- M SERIES CCT 40010811101124010COMPUTER, BLOOD-PRESSURE5 reports
- M SERIES CCT 60021911100124010COMPUTER, BLOOD-PRESSURE5 reports
- M SERIES CCT 40010911101124010COMPUTER, BLOOD-PRESSURE2 reports
- M SERIES CCT 40010811100024010COMPUTER, BLOOD-PRESSURE1 report
- M SERIES CCT 60010811100024010COMPUTER, BLOOD-PRESSURE1 report
- M SERIES CCT 60010811100124010COMPUTER, BLOOD-PRESSURE1 report
Recent M SERIES CCT adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
DeathNov 15, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A 45-YEAR-OLD MALE PATIENT, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED; HOWEVER, THEY DO NOT BELIEVE THIS WAS DEVICE ASSOCIATED.
Patient problems: Necrosis
MAUDE report 15799627 ↗MalfunctionOct 9, 2020No Display/Image; Activation Problem; No Display/Image; Activation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY FLICKERED AND DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10655837 ↗MalfunctionSep 22, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY WAS FLICKERING AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10567143 ↗MalfunctionJul 27, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10325907 ↗MalfunctionJul 15, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10276260 ↗MalfunctionJul 13, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10265039 ↗MalfunctionJun 25, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "SYSTEM FAULT 36" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10196847 ↗MalfunctionJun 23, 2020Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "PACER FAULT 116" AND "DEFIB DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10185236 ↗MalfunctionJun 12, 2020Pacing Problem; Pacing Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "PACER FAULT 117" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10148756 ↗MalfunctionJun 11, 2020Pacing Problem; Pacing Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "PACER FAULT 116" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10143620 ↗MalfunctionJun 1, 2020Activation Problem; Activation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10105947 ↗MalfunctionMay 15, 2020Activation Problem; Activation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10062154 ↗MalfunctionMay 15, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "DEFIB FAULT 78" AND "DEFIB FAULT 79" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10061595 ↗MalfunctionMay 14, 2020Activation Problem; Activation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10059256 ↗MalfunctionMay 12, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10048613 ↗
Other Zoll Medical Corporation products
- R SERIES94 models · 8,522 reports
- X SERIES163 models · 6,915 reports
- AED PLUS30 models · 1,966 reports · 1 recall
- Z Vent6 models · 1,204 reports · 1 recall
- Model EMV+ Vent, Army, MRI3 models · 964 reports
- E SERIES160 models · 833 reports
- ONESTEP8 models · 586 reports
- M SERIES BIPHASIC59 models · 572 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page