Quick summary

This Zoll X Series monitor defibrillator has 166 reports, including 2 deaths and no injuries. Reports describe the device failing its self-test, failing to power up, or shutting down unexpectedly, and in some cases failing to deliver a shock. Many note no patient was involved. No recalls name this device, and reports rose from 4 in the prior 12 months to 10 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
166
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 166 adverse event reports that list this exact device (DI 00847946006020), received between Mar 21, 2017 and Aug 25, 2026 — classified as 2 deaths, 164 malfunctions.

The most frequently reported specific device problem is “Failure of Device to Self-Test” (41 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 3 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure of Device to Self-Test41
  • Unable to Obtain Readings22
  • Failure to Power Up16
  • Loss of Power16
  • Unexpected Shutdown15
  • Pacing Problem14
  • Defibrillation/Stimulation Problem13
  • Reset Problem9

Patient problems

  • No Patient Involvement66
  • No Clinical Signs, Symptoms or Conditions59
  • No Information20
  • No Consequences Or Impact To Patient15
  • Insufficient Information5
  • Necrosis1

Recent adverse event reports

The 25 most recent of 166 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE PROMPTED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50431273 ↗
  • MalfunctionJul 9, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25769919 ↗
  • MalfunctionMay 27, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF-TEST FOR PACER FUNCTION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25299539 ↗
  • MalfunctionMay 11, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST FOR DEFIB FUNCTION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25135137 ↗
  • MalfunctionApr 9, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING INCOMING TESTING, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24835290 ↗
  • MalfunctionDec 29, 2025Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 23908808 ↗
  • MalfunctionDec 5, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23733099 ↗
  • DeathNov 24, 2025Failure to Deliver Shock/Stimulation; Delayed Charge Time; Failure to Deliver Shock/Stimulation; Delayed Charge Time

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A 77-YEAR-OLD FEMALE PATIENT, THE DEVICE WAS SLOW TO CHARGE AND FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED.

    Patient problems: Insufficient Information

    MAUDE report 23638618 ↗
  • MalfunctionNov 19, 2025No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23597983 ↗
  • MalfunctionNov 4, 2025Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23465531 ↗
  • MalfunctionOct 20, 2025Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED "ECG MONITORING FAILURE" AND "ECG DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23344023 ↗
  • MalfunctionSep 9, 2025Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT WHILE PACING A 77-YEAR-OLD MALE PATIENT, THE DEVICE DISPLAYED A "PACE DEFIB DEVICE FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004641 ↗
  • MalfunctionMay 15, 2025Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22026824 ↗
  • MalfunctionJan 20, 2025Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21191096 ↗
  • MalfunctionAug 21, 2024No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20013656 ↗
  • MalfunctionJun 28, 2024Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19637198 ↗
  • MalfunctionJun 6, 2024Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST FOR DEFIB AND PACER FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19475304 ↗
  • MalfunctionJan 31, 2024Fire; Fire

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S POWER SUPPLY CAUGHT FIRE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18614131 ↗
  • MalfunctionJan 16, 2024Failure to Capture; Failure to Capture

    COMPLAINANT ALLEGED THAT DURING TRAINING BY STAFF, THE DEVICE FAILED TO CAPTURE THE HEART RHYTHM. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18517632 ↗
  • MalfunctionAug 29, 2023Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST FOR DEFIB. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17640924 ↗
  • MalfunctionAug 14, 2023Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17541889 ↗
  • MalfunctionAug 8, 2023Pacing Problem; Pacing Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO PACE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17480828 ↗
  • MalfunctionJul 17, 2023Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17333193 ↗
  • MalfunctionJul 12, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A 60-YEAR-OLD MALE PATIENT, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17310030 ↗
  • MalfunctionJul 12, 2023Use of Device Problem; Reset Problem; Use of Device Problem; Reset Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DID NOT OPERATE CORRECTLY AND RESTART/REBOOT ITSELF. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17309999 ↗