M SERIES BIPHASIC 60021711100123010Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN,ACLS/MNL-1,12LEAD,SPO2,ETCO2,NIBP,BIPHASIC,EL,AC,PC,RC,SM,CM,ENG · Class 3

All models of M SERIES BIPHASIC (59)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00847946013738), received between Apr 6, 2017 and Aug 27, 2020 — classified as 1 death, 9 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Power Up4
  • Defibrillation/Stimulation Problem2
  • Battery Problem1
  • Device Operational Issue1
  • Difficult or Delayed Activation1
  • Failure to Deliver Shock/Stimulation1
  • No Display/Image1
  • Unexpected Shutdown1

Patient problems

  • No Patient Involvement8
  • Death1
  • No Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2020Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10461302 ↗
  • MalfunctionMay 13, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 85" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10053924 ↗
  • MalfunctionSep 26, 2019Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9122143 ↗
  • MalfunctionJul 19, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8807662 ↗
  • MalfunctionJul 12, 2019Failure to Power Up; Failure to Power Up

    DURING TRAINING BY A ZOLL MEDICAL SALES REPRESENTATIVE, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8784942 ↗
  • DeathDec 11, 2018Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A MALE PATIENT (AGE UNKNOWN), THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED. COMPLAINANT INDICATED THAT DURING SUBSEQUENT BIOMED TESTING, THE MALFUNCTION WAS NOT DUPLICATED.

    Patient problems: Death

    MAUDE report 8152837 ↗
  • MalfunctionNov 6, 2017Device Operational Issue; Device Operational Issue

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7003927 ↗
  • MalfunctionOct 31, 2017Failure to Power Up; Difficult or Delayed Activation; Failure to Power Up; Difficult or Delayed Activation

    COMPLAINANT ALLEGED THAT THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. ONCE THE DEVICE IS ABLE TO POWER UP, THERE IS A DELAY UPON POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6989816 ↗
  • MalfunctionApr 20, 2017No Display/Image; Failure to Power Up; No Display/Image; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY USING AC POWER AND WAS UNABLE TO POWER ON USING BATTERY POWER. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6507222 ↗
  • MalfunctionApr 6, 2017Battery Problem; Battery Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) THE DEVICE'S BATTERY "SHORTED" IN THE DEVICE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Information

    MAUDE report 6464698 ↗

Other Automated external defibrillators (non-wearable) devices