E SERIES 80001710020124013Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN, ACLS-1, 12 LEAD, SPO2/ETCO2/NIBP, BT/GPS, PACE, AC, NO RAILS, DMST, REFURB · Class 3

All models of E SERIES (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll E Series defibrillator is a wheeled, non-wearable automated external defibrillator used for cardiac resuscitation. The 28 reports mostly describe the unit failing to power up during routine shift checks or biomed testing, along with blank screens, fault messages, and switching problems between modes. All reports involve no patient, with no deaths and no injuries. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00847946015435), received between Aug 6, 2015 and Oct 12, 2020 — classified as 28 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Power Up17
  • No Display/Image2
  • Pacing Problem2
  • Unexpected Shutdown2
  • Defibrillation/Stimulation Problem1
  • Device Displays Incorrect Message1
  • Device Operational Issue1
  • Failure of Device to Self-Test1

Patient problems

  • No Patient Involvement23
  • No Consequences Or Impact To Patient2
  • No Information2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 12, 2020Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 72" AND "PACER FAULT 116" MESSAGES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10662897 ↗
  • MalfunctionMay 18, 2020Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10067217 ↗
  • MalfunctionMar 16, 2020Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9837286 ↗
  • MalfunctionJul 22, 2019Mechanical Problem; Mechanical Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO SWITCH MODES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8813870 ↗
  • MalfunctionJul 12, 2019No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND BLANKED OUT. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8784961 ↗
  • MalfunctionMay 30, 2019Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8655623 ↗
  • MalfunctionMay 30, 2019Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8654280 ↗
  • MalfunctionFeb 5, 2019Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8306169 ↗
  • MalfunctionJan 2, 2019Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE RESTART/REBOOT ITSELF MULTIPLE TIMES. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8209290 ↗
  • MalfunctionJan 2, 2019Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) IN CARDIAC ARREST, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THE CLINICIAN PLUGGED THE DEVICE INTO AC TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1220908-2018-03771 FOR A SIMILAR EVENT REPORTED FROM THE SAME CUSTOMER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8209063 ↗
  • MalfunctionDec 12, 2018Pacing Problem; Pacing Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "PACER FAULT 117" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8155561 ↗
  • MalfunctionNov 21, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8095632 ↗
  • MalfunctionNov 8, 2018Intermittent Loss of Power; Intermittent Loss of Power

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8052772 ↗
  • MalfunctionOct 29, 2018No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED UP WITH A WHITE SCREEN AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8015927 ↗
  • MalfunctionSep 24, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7902525 ↗
  • MalfunctionApr 3, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7392102 ↗
  • MalfunctionMar 20, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7353746 ↗
  • MalfunctionMar 12, 2018Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7330850 ↗
  • MalfunctionFeb 26, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7297791 ↗
  • MalfunctionFeb 26, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7297537 ↗
  • MalfunctionJan 2, 2018Device Operational Issue; Device Operational Issue

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE POWERED UP IN THE INCORRECT MODE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7157853 ↗
  • MalfunctionOct 20, 2017Failure to Power Up; Failure to Power Up

    SUBSEQUENT TO PATIENT USE, THE COMPLAINANT ALLEGED, THE DEVICE INTERMITTENTLY WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6966147 ↗
  • MalfunctionSep 1, 2017Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6839595 ↗
  • MalfunctionMay 10, 2017Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6556687 ↗
  • MalfunctionApr 6, 2017Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6464913 ↗

Other Automated external defibrillators (non-wearable) devices