Quick summary
This Zoll OneStep CPR electrode pad appears in 89 adverse event reports, with 3 deaths and 6 injuries. Reports often describe failed defibrillator self-tests, pads not being detected, and sparking. Some patient problems listed include burns. There was 1 report in the past 12 months, down from 7 in the year before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 89
- Deaths reported
- 3
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 89 adverse event reports that list this exact device (DI 00847946016289), received between Mar 17, 2017 and Dec 1, 2025 — classified as 3 deaths, 6 injuries, 80 malfunctions.
The most frequently reported specific device problem is “Failure of Device to Self-Test” (27 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure of Device to Self-Test27
- Unable to Obtain Readings24
- Patient Device Interaction Problem11
- Failure to Deliver Shock/Stimulation7
- Sparking6
- Device Operational Issue3
- Arcing of Electrodes2
- Defibrillation/Stimulation Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions48
- No Patient Involvement21
- Burn(s)5
- Insufficient Information5
- No Information4
- Necrosis2
- No Consequences Or Impact To Patient2
- Death1
Recent adverse event reports
The 25 most recent of 89 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 1, 2025Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEVICE FAILED THE SELF-TEST USING THESE ONESTEP PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23680187 ↗MalfunctionJun 17, 2025Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THESE ATTACHED ELECTRODES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22231617 ↗MalfunctionMay 19, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), WHEN APPLYING THESE ELECTRODE PADS TO THE PATIENT, ONE OF THE PAD'S ADHESIVE WAS STUCK TO THE COVER. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 22046405 ↗MalfunctionMay 7, 2025Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21976789 ↗MalfunctionApr 25, 2025Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ELECTRODE PADS FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21909205 ↗MalfunctionFeb 26, 2025Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21468547 ↗MalfunctionJan 23, 2025Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEFIBRILLATOR FAILED THE SELF TEST USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21219599 ↗InjuryDec 5, 2024Sparking; Patient Device Interaction Problem; Sparking; Patient Device Interaction Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A 50-YEAR-OLD MALE PATIENT, SPARKS WERE EMITTED FROM THE ELECTRODES, AFTER REMOVING THE ELECTRODE PADS, BURNS WERE FOUND ON THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUFFERED BURNS. NO INFORMATION WAS AVAILABLE ON THE DEGREE OF BURNS.
Patient problems: Burn(s); Low Blood Pressure/ Hypotension
MAUDE report 20854879 ↗MalfunctionSep 3, 2024Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20126308 ↗MalfunctionJun 18, 2024Use of Device Problem; Use of Device Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THESE ELECTRODE PADS DID NOT WORK AS EXPECTED. THERE IS NO ADDITIONAL INFORMATION REGARDING THE MALFUNCTION AT THIS TIME. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19559991 ↗MalfunctionJun 14, 2024
CHECKED ZOLL DEFIBRILLATOR FOR ROUTINE CHECK AND DEFIBRILLATOR READING "CHECK PADS". CONNECTED ANOTHER SET OF DEFIBRILLATOR PADS AND WAS ABLE TO COMPLETE ZOLL CHECK. REMOVED PADS FROM INVENTORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19536385 ↗MalfunctionJun 3, 2024Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED THE SELF TEST USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19448394 ↗MalfunctionApr 17, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A 65-YEAR-OLD MALE PATIENT, THE ELECTRODE PADS WOULD NOT ADHERE TO THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER SET OF PADS TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19124794 ↗MalfunctionFeb 12, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEVICE FAILED TO DISCHARGE USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18685495 ↗MalfunctionFeb 9, 2024Failure to Deliver Shock/Stimulation; Failure of Device to Self-Test; Failure to Deliver Shock/Stimulation; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEVICE FAILED SELF-TEST USING THESE ELECTRODE PADS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18677116 ↗MalfunctionJan 24, 2024Sparking; Sparking
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), A SPARK WAS SEEN FROM THE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18570874 ↗MalfunctionJan 4, 2024Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18443070 ↗MalfunctionNov 8, 2023Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEVICE FAILED SELF TEST USING THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18091144 ↗MalfunctionOct 5, 2023Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED THE SELF TEST WITH THESE ELECTRODE PADS ATTACHED. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17877037 ↗InjurySep 8, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO PACE A 77-YEAR-OLD FEMALE PATIENT, AFTER REMOVING THE ELECTRODE PADS, BURNS WERE FOUND ON THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY SUSTAINED A BURN. THE CUSTOMER WAS UNABLE TO PROVIDE INFORMATION ON THE DEGREE OF THE BURN.
Patient problems: Burn(s)
MAUDE report 17714959 ↗MalfunctionJun 30, 2023Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE ASSOCIATED DEFIBRILLATOR DISPLAYED A "POOR PAD CONTACT" MESSAGE USING THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17240566 ↗MalfunctionMay 24, 2023Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED TO DISCHARGE USING THESE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16994736 ↗MalfunctionApr 20, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE ELECTRODES WOULD NOT ADHERE TO THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1218058-2023-00082 FOR A SIMILAR EVENT REPORTED FROM THE SAME CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16778080 ↗MalfunctionApr 20, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE ELECTRODES WOULD NOT ADHERE TO THE PATIENT'S SKIN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1218058-2023-00081 FOR A SIMILAR EVENT REPORTED FROM THE SAME CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16778040 ↗MalfunctionApr 11, 2023Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DETECT THESE ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16713544 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page