R SERIES 30001000001110013Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN, MAN-1, PLUS, 3/5 LD, STD ECG, W/PACE, AC PWR, W/PACE, CF/USB, DOMESTIC · Class 3

All models of R SERIES (94)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll R Series defibrillator is an external device used to restart the heart in emergencies. Its 149 reports describe the unit being unable to detect electrode pads or get readings, failing its self-test, or showing an unreadable screen, and some mention it failing to deliver a shock. The reports include 1 injury and 2 deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
150
Deaths reported
2
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 150 adverse event reports that list this exact device (DI 00847946017170), received between Jul 11, 2017 and Sep 15, 2026 — classified as 2 deaths, 1 injury, 147 malfunctions.

The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 6 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 4 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 2 reportsAug 2026: 3 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defibrillation/Stimulation Problem36
  • Unable to Obtain Readings23
  • No Display/Image20
  • Failure to Power Up18
  • Failure of Device to Self-Test11
  • Failure to Deliver Shock/Stimulation11
  • Pacing Problem10
  • Unexpected Shutdown10

Patient problems

  • No Clinical Signs, Symptoms or Conditions101
  • No Patient Involvement29
  • Insufficient Information8
  • No Consequences Or Impact To Patient4
  • No Information4
  • Administration Site Reaction1
  • Death1
  • Necrosis1

Recent adverse event reports

The 25 most recent of 150 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Sparking; Packaging Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE AND GENDER UNKNOWN), SPARK WAS OBSERVED FROM THE ELECTRODE PADS AND IT WAS NOTED THAT THE PADS HAD A HOLE IN THEM. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50775734 ↗
  • MalfunctionAug 21, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50371554 ↗
  • MalfunctionAug 17, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY SHOWED HORIZONTAL LINES AND WAS NOT LEGIBLE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50275069 ↗
  • MalfunctionAug 14, 2026Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50253416 ↗
  • MalfunctionJul 29, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50074035 ↗
  • MalfunctionJul 20, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25870199 ↗
  • MalfunctionJun 19, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25563212 ↗
  • MalfunctionJun 15, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25503051 ↗
  • MalfunctionMay 12, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150510 ↗
  • MalfunctionMay 11, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25135172 ↗
  • MalfunctionApr 29, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ERROR FFFF" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25027976 ↗
  • MalfunctionApr 14, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872757 ↗
  • MalfunctionMar 17, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB DISABLED" AND "DEFIB FAULT 76" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24616613 ↗
  • MalfunctionFeb 20, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ERROR FFFF" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24402157 ↗
  • MalfunctionFeb 2, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24233429 ↗
  • MalfunctionJan 21, 2026Mechanical Problem; Mechanics Altered

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED UP IN THE INCORRECT MODE AND WAS UNABLE TO SWITCH MODES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24134043 ↗
  • MalfunctionOct 22, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23360472 ↗
  • MalfunctionOct 20, 2025Grounding Malfunction; Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TESTING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23336684 ↗
  • MalfunctionOct 16, 2025Grounding Malfunction; Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TESTING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23315959 ↗
  • MalfunctionOct 15, 2025Pacing Problem; Unable to Obtain Readings; Defibrillation/Stimulation Problem; Pacing Problem; Unable to Obtain Readings; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ERROR FFFF," "DEFIB DISABLED," AND "PACER DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23294980 ↗
  • MalfunctionSep 3, 2025Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22952301 ↗
  • MalfunctionJul 7, 2025Use of Device Problem; Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, CRITICAL ERROR CODES WERE OBSERVED IN THE ERROR LOG. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22411844 ↗
  • MalfunctionJun 18, 2025Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22285742 ↗
  • MalfunctionMay 30, 2025Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22114862 ↗
  • MalfunctionMay 19, 2025Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22046194 ↗

Other Automated external defibrillators (non-wearable) devices