Quick summary

This Zoll R Series defibrillator is an external device used to restart the heart in emergencies. Its 149 reports describe the unit being unable to get readings, failing its self-test, or having a blank screen, and some mention it failing to discharge or advising no shock for a shockable rhythm. The reports include 3 deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
152
Deaths reported
3
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 152 adverse event reports that list this exact device (DI 00847946017408), received between Aug 22, 2017 and Sep 15, 2026 — classified as 3 deaths, 149 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (30 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 4 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 3 reportsSep 2024: 4 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 5 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 5 reportsJan 2026: 0 reportsFeb 2026: 6 reportsMar 2026: 5 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 4 reportsJul 2026: 4 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unable to Obtain Readings30
  • Defibrillation/Stimulation Problem26
  • Grounding Malfunction23
  • No Display/Image17
  • Failure to Power Up16
  • Failure to Deliver Shock/Stimulation13
  • Pacing Problem8
  • Failure of Device to Self-Test7

Patient problems

  • No Clinical Signs, Symptoms or Conditions114
  • No Patient Involvement22
  • Insufficient Information10
  • Death2
  • No Consequences Or Impact To Patient2
  • Burn(s)1
  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 152 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Pacing Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO PACE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772573 ↗
  • MalfunctionSep 8, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50669304 ↗
  • MalfunctionSep 3, 2026Failure to Deliver Shock/Stimulation; Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50592376 ↗
  • MalfunctionAug 18, 2026No Display/Image; Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50308850 ↗
  • MalfunctionJul 29, 2026Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50071037 ↗
  • MalfunctionJul 21, 2026Failure to Capture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25893284 ↗
  • MalfunctionJul 20, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25876767 ↗
  • MalfunctionJul 2, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25693452 ↗
  • MalfunctionJun 15, 2026Failure to Deliver Shock/Stimulation; Activation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED TO DISCHARGE AND THE DEVICE DID NOT RESPOND TO THE FRONT PANEL CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25503316 ↗
  • MalfunctionJun 12, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25485551 ↗
  • MalfunctionJun 8, 2026Incorrect Interpretation of Signal

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A MALE PATIENT (AGE UNKNOWN), THE DEVICE PROMPTED "NO SHOCK ADVISED" FOR A RHYTHM CLINICIANS BELIEVED TO BE SHOCKABLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 25425707 ↗
  • MalfunctionJun 2, 2026Incorrect Interpretation of Signal

    COMPLAINANT ALLEGED WHILE ATTEMPTING TO TREAT A PATIENT DUE TO ARRYTHMIA (AGE & GENDER UNKNOWN), THE DEVICE PROMPTED "NO SHOCK ADVISED" FOR A RHYTHM CLINICIANS BELIEVED TO BE SHOCKABLE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25365115 ↗
  • MalfunctionMay 20, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25228957 ↗
  • MalfunctionMay 1, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25057746 ↗
  • MalfunctionApr 10, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24839711 ↗
  • MalfunctionMar 25, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24686992 ↗
  • MalfunctionMar 23, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24664994 ↗
  • MalfunctionMar 16, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ECG FAULT" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24606067 ↗
  • MalfunctionMar 5, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TESTING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24519995 ↗
  • MalfunctionMar 3, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24495996 ↗
  • MalfunctionFeb 24, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24429025 ↗
  • MalfunctionFeb 17, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24364448 ↗
  • MalfunctionFeb 16, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24343844 ↗
  • MalfunctionFeb 11, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND CLINICAL DATA COULD NOT BE SEEN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24316651 ↗
  • MalfunctionFeb 6, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TESTING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24279600 ↗