PROPAQ MD 300-1011301-01-66Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN ASSY, 12LD, SPO2, C02, NIBP, IBP, PROPAQ MD, REFURB · Class 3
All models of PROPAQ MD (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll Propaq MD is a portable automated external defibrillator used to restart hearts during cardiac emergencies. The 38 reports mostly describe the unit failing its own self-test, not powering up or powering off on its own, or powering on with no display — several occurring while staff were trying to treat a patient, with no reports of deaths or injuries. Most reports come from servicing and routine shift checks rather than patient care. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 38
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 00847946017651), received between Aug 12, 2015 and May 18, 2023 — classified as 38 malfunctions.
The most frequently reported specific device problem is “Failure of Device to Self-Test” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure of Device to Self-Test9
- Failure to Power Up6
- Loss of Power5
- Reset Problem4
- Unexpected Shutdown4
- Device Displays Incorrect Message3
- Defibrillation/Stimulation Problem2
- Device Stops Intermittently1
Patient problems
- No Patient Involvement23
- No Information12
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2023Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16955125 ↗MalfunctionJun 24, 2020Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF-TEST FOR DEFIB FUNCTION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 10189900 ↗MalfunctionJun 5, 2020No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10124030 ↗MalfunctionFeb 18, 2020Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST FOR DEFIB AND PACER FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9720383 ↗MalfunctionFeb 10, 2020Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST FOR DEFIB AND PACER FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9689874 ↗MalfunctionFeb 7, 2020Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9682057 ↗MalfunctionNov 27, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9385239 ↗MalfunctionOct 29, 2019Failure to Deliver Shock/Stimulation; Energy Output Problem; Failure to Deliver Shock/Stimulation; Energy Output Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE FAILED TO DISCHARGE AND INTERNALLY DUMPED THE ENERGY AFTER CHARGING UP TO DEFIBRILLATE THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9251080 ↗MalfunctionOct 11, 2019Reset Problem; Reset Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE RESTART/REBOOT ITSELF MULTIPLE TIMES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 9181066 ↗MalfunctionSep 24, 2019Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN) DURING FLIGHT, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 9110810 ↗MalfunctionAug 13, 2019Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 8889155 ↗MalfunctionAug 7, 2019Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE FAILED SELF-TEST FOR DEFIB FUNCTION. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 8869350 ↗MalfunctionMay 22, 2019Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE SHUTS OFF. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8633646 ↗MalfunctionMay 10, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8600300 ↗MalfunctionApr 22, 2019Reset Problem; Reset Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE RESTART/REBOOT ITSELF MULTIPLE TIMES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8536885 ↗MalfunctionApr 11, 2019Reset Problem; Reset Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE RESTARTS/REBOOTS BY SELF MULTIPLE TIMES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8505934 ↗MalfunctionMar 14, 2019Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8422116 ↗MalfunctionMar 8, 2019Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DEFIB OUTPUT WAS INCORRECT. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8404929 ↗MalfunctionOct 25, 2018Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF-TEST FOR DEFIB AND PACER FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8005085 ↗MalfunctionSep 18, 2018Pacing Problem; Pacing Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO PACE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7885280 ↗MalfunctionSep 14, 2018Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST FOR DEFIB FUNCTION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7877842 ↗MalfunctionSep 13, 2018Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "SHORT DETECTED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7874044 ↗MalfunctionAug 20, 2018Failure to Read Input Signal; Failure to Read Input Signal
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7798477 ↗MalfunctionAug 2, 2018Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "SHORT DETECTED" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7744989 ↗MalfunctionJun 19, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE AND GENDER UNKNOWN), THE DEVICE DISPLAYED A "SHORT DETECTED" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN USED OTHER METHODS TO CONTINUE TREATING THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1220908-2018-01466 FOR A SIMILAR EVENT REPORTED FROM WITH SAME DEVICE.
Patient problems: No Information
MAUDE report 7612885 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page