X SERIES 620-0220011-01Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN ASSY, X SERIES MONITOR/DEFIBRILLATOR 3/5 LEAD, PACING, NIBP, SPO2, CPR EXPANSION PACK, DMST · Class 3
All models of X SERIES (163)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll X Series monitor/defibrillator is a portable resuscitation unit used in hospitals and by emergency services. The 33 reports mostly describe units failing to power on, screens going blank or completely white, and self-tests failing for defibrillation and pacing, all found during routine shift checks or biomedical testing with no patient involved. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00847946018665), received between May 17, 2018 and Sep 15, 2026 — classified as 34 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Power Up14
- No Display/Image12
- Failure of Device to Self-Test5
- Defibrillation/Stimulation Problem1
- Device Displays Incorrect Message1
- Failure to Run on Battery1
- Pacing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions24
- No Patient Involvement7
- No Information2
- Intraoperative Cardiac Valve Injury1
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50779297 ↗MalfunctionAug 17, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50275168 ↗MalfunctionApr 29, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25031518 ↗MalfunctionMar 30, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24717699 ↗MalfunctionJan 15, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24081344 ↗MalfunctionNov 3, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23451473 ↗MalfunctionJul 16, 2025Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22521264 ↗MalfunctionOct 21, 2024Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED SELF TEST FOR DEFIB AND PACER FUNCTIONS. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: Intraoperative Cardiac Valve Injury
MAUDE report 20492745 ↗MalfunctionSep 19, 2024Pacing Problem; Defibrillation/Stimulation Problem; Pacing Problem; Defibrillation/Stimulation Problem
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB PACE DEVICE FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20257373 ↗MalfunctionMar 5, 2024No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18840485 ↗MalfunctionAug 31, 2023Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17660700 ↗MalfunctionAug 3, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17455123 ↗MalfunctionJul 12, 2023Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED SELF TEST FOR DEFIB FUNCTION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17309921 ↗MalfunctionMar 6, 2023Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16490875 ↗MalfunctionFeb 15, 2023Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16380743 ↗MalfunctionNov 21, 2022Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15835558 ↗MalfunctionNov 8, 2022No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15754358 ↗MalfunctionOct 17, 2022No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15616269 ↗MalfunctionSep 24, 2022Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15477098 ↗MalfunctionDec 17, 2021Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13027470 ↗MalfunctionDec 14, 2021No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13003905 ↗MalfunctionNov 19, 2021Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12848248 ↗MalfunctionNov 15, 2021No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12814078 ↗MalfunctionOct 14, 2021Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12634221 ↗MalfunctionOct 13, 2021No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12625686 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗