X SERIES 620-2221511-01Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,X SERIES MONITOR/DEFIBRILLATOR W/PACING, 12 LEAD W/INTERP, SPO2, NIBP, IBP/TEMP, CPR EXPANSION PACK, ETCO2, DMST · Class 3

All models of X SERIES (163)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll X Series is a monitor and defibrillator used in emergency care. Many of its 453 reports describe a blank screen, failure to power up or pass self-tests, and unexpected shutdowns, and many were found during routine testing. The reports list 2 deaths, no injuries and 451 malfunctions, and no recalls name this device. Reports fell to 64 in the most recent year from 103 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
456
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 456 adverse event reports that list this exact device (DI 00847946019181), received between Apr 19, 2018 and Sep 29, 2026 — classified as 2 deaths, 454 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (107 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 0 reportsJan 2024: 8 reportsFeb 2024: 7 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 2 reportsJun 2024: 7 reportsJul 2024: 8 reportsAug 2024: 3 reportsSep 2024: 11 reportsOct 2024: 2 reportsNov 2024: 4 reportsDec 2024: 3 reportsJan 2025: 5 reportsFeb 2025: 9 reportsMar 2025: 2 reportsApr 2025: 9 reportsMay 2025: 15 reportsJun 2025: 13 reportsJul 2025: 14 reportsAug 2025: 8 reportsSep 2025: 9 reportsOct 2025: 12 reportsNov 2025: 6 reportsDec 2025: 3 reportsJan 2026: 7 reportsFeb 2026: 5 reportsMar 2026: 9 reportsApr 2026: 9 reportsMay 2026: 7 reportsJun 2026: 4 reportsJul 2026: 6 reportsAug 2026: 3 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image107
  • Unable to Obtain Readings86
  • Failure to Power Up74
  • Failure of Device to Self-Test62
  • Unexpected Shutdown58
  • Defibrillation/Stimulation Problem29
  • Pacing Problem21
  • Reset Problem15

Patient problems

  • No Clinical Signs, Symptoms or Conditions350
  • Insufficient Information49
  • No Patient Involvement33
  • No Consequences Or Impact To Patient10
  • No Information6
  • Cardiac Arrest5
  • Arrhythmia1
  • Bradycardia1

Recent adverse event reports

The 25 most recent of 456 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938280 ↗
  • MalfunctionSep 28, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915993 ↗
  • MalfunctionSep 24, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882908 ↗
  • MalfunctionSep 18, 2026Unexpected Shutdown

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE INAPPROPRIATELY SHUT DOWN.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50818163 ↗
  • MalfunctionSep 15, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF-TEST FOR DEFIB FUNCTION.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50775735 ↗
  • MalfunctionSep 11, 2026Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE PROMPTED A "DO NOT USE" ERROR MESAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50735802 ↗
  • MalfunctionSep 4, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50616386 ↗
  • MalfunctionSep 4, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50616382 ↗
  • MalfunctionAug 27, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED SELF TEST.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478215 ↗
  • MalfunctionAug 13, 2026Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50239840 ↗
  • MalfunctionAug 3, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "ECG DISABLED" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50114107 ↗
  • MalfunctionJul 27, 2026Intermittent Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50036541 ↗
  • MalfunctionJul 27, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50018380 ↗
  • MalfunctionJul 20, 2026Unexpected Shutdown

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25877904 ↗
  • MalfunctionJul 20, 2026Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25870234 ↗
  • MalfunctionJul 14, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25813116 ↗
  • MalfunctionJul 9, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "ECG DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25762032 ↗
  • MalfunctionJun 29, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST FOR DEFIB FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25651534 ↗
  • MalfunctionJun 26, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25635661 ↗
  • MalfunctionJun 23, 2026Unexpected Shutdown

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25598251 ↗
  • MalfunctionJun 19, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25563178 ↗
  • MalfunctionMay 28, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25315969 ↗
  • MalfunctionMay 27, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25299536 ↗
  • MalfunctionMay 26, 2026Pacing Problem; Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "PACE DEFIB DEVICE FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25276688 ↗
  • MalfunctionMay 22, 2026Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25256017 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →