X SERIES 620-2431011-01Recalls & reported problems
Zoll Medical Corporation · Defibrillator ,MAIN ASSY, X SERIES MONITOR/DEFIBRILLATOR 12 LEAD/W INTERP, PACING, VOICE, NIBP, SPO2, SPCO, CPR EXPANSION PACK AND ETCO2, DMST · Class 3
All models of X SERIES (163)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zoll X Series defibrillator is a non-wearable automated external defibrillator used in hospitals and clinics. Most of the 37 reports describe the screen turning completely white and becoming unreadable, along with units that would not power up or would not deliver a shock. There are no deaths and no injuries reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 37
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00847946019457), received between Dec 4, 2018 and Jan 5, 2026 — classified as 37 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (26 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image26
- Failure to Power Up5
- Unable to Obtain Readings2
- Battery Problem1
- Defibrillation/Stimulation Problem1
- Failure to Deliver Shock/Stimulation1
- Pacing Problem1
- Unexpected Shutdown1
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- No Patient Involvement6
- Insufficient Information3
- Cardiac Arrest1
- Inadequate Osseointegration1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 23965392 ↗MalfunctionJan 2, 2026Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A 46 YEAR OLD MALE PATIENT, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 23945775 ↗MalfunctionDec 23, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICES SCREEN TURNED COMPLETELY WHITE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: Inadequate Osseointegration
MAUDE report 23876888 ↗MalfunctionDec 22, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23863457 ↗MalfunctionAug 13, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22795080 ↗MalfunctionJul 30, 2025Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WOULD NOT POWER UP. ANOTHER DEVICE WAS USED. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22665595 ↗MalfunctionJul 2, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22381983 ↗MalfunctionOct 24, 2024Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO PACE A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20526808 ↗MalfunctionAug 2, 2024Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Cardiac Arrest
MAUDE report 19892951 ↗MalfunctionJun 3, 2024No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19453551 ↗MalfunctionMar 20, 2024No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18943983 ↗MalfunctionMar 20, 2024Unexpected Shutdown; Unexpected Shutdown
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18941835 ↗MalfunctionFeb 12, 2024No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18690502 ↗MalfunctionSep 21, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17787730 ↗MalfunctionSep 21, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17787724 ↗MalfunctionSep 20, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17779973 ↗MalfunctionJul 19, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17352791 ↗MalfunctionJun 22, 2023Failure to Power Up; Failure to Power Up
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17181947 ↗MalfunctionMay 23, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16983749 ↗MalfunctionFeb 24, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16437394 ↗MalfunctionJan 11, 2023No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16137344 ↗MalfunctionOct 21, 2022No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15652409 ↗MalfunctionSep 15, 2022Failure to Power Up; Battery Problem; Failure to Power Up; Battery Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WOULD INTERMITTENTLY NOT POWER UP . COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15432938 ↗MalfunctionMar 24, 2022No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13873656 ↗MalfunctionOct 5, 2021Use of Device Problem; Use of Device Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DID NOT OPERATE CORRECTLY. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 12577996 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page