PROPAQ MD 320-2421312-01Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN ASSY, PROPAQ MONITOR/DEFIBRILLATOR 12 LEAD/W INTERP, PACING, NIBP, SPO2, IBP, TEMP CPR EXPANSION PACK AND ETCO2, W/CP2, DMST · Class 3

All models of PROPAQ MD (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00847946020163), received between Aug 17, 2020 and Dec 17, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Shutdown4
  • Failure of Device to Self-Test3
  • Unable to Obtain Readings3
  • High impedance2
  • Pacing Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient2
  • Insufficient Information1
  • No Patient Involvement1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 17, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST FOR DEFIB AND PACER FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23826252 ↗
  • MalfunctionDec 1, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST FOR "DEFIB" AND "PACER" FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23681015 ↗
  • MalfunctionDec 16, 2024Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT NUMBERS 1220908-2024-04447, 1220908-2024-04446, 1220908-2024-04500 FOR SIMILAR REPORTS FROM THE SAME CUSTOMER.

    Patient problems: Insufficient Information

    MAUDE report 20941002 ↗
  • MalfunctionDec 16, 2024Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A 16-YEAR-OLD MALE PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT NUMBERS 1220908-2024-04446, 1220908-2024-04499, 1220908-2024-04500 FOR SIMILAR REPORTS FROM THE SAME CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20940884 ↗
  • MalfunctionDec 16, 2024Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT NUMBERS 1220908-2024-04447, 1220908-2024-04499, 1220908-2024-04500 FOR SIMILAR REPORTS FROM THE SAME CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20938878 ↗
  • MalfunctionDec 16, 2024Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE AND GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THE CLINICIAN CHANGED THE BATTERY IN THE DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT NUMBERS 1220908-2024-04447, 1220908-2024-04446, 1220908-2024-04499 FOR SIMILAR REPORTS FROM THE SAME CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20938846 ↗
  • MalfunctionJul 22, 2024Unable to Obtain Readings; Failure of Device to Self-Test; Unable to Obtain Readings; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19795020 ↗
  • MalfunctionMay 31, 2023Pacing Problem; Pacing Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "PACER DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17031991 ↗
  • MalfunctionFeb 8, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" ERROR MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16330101 ↗
  • MalfunctionSep 11, 2020Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "ECG DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 10523530 ↗
  • MalfunctionAug 17, 2020High impedance; High impedance

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) THE DEVICE DISPLAYED AN "UNKNOWN FAULT" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10417233 ↗
  • MalfunctionAug 17, 2020High impedance; High impedance

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED AN UNKNOWN ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10415350 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →