X SERIES 620-2231011-01-66Recalls & reported problems

Zoll Medical Corporation · MAIN ASSY, X SERIES MONITOR/DEFIBRILLATOR 12 LEAD/W INTERP, PACING, NIBP, SPO2, SPCO, CPR EXPANSION PACK AND ETCO2, DMST, REFURB · Class 3

All models of X SERIES (163)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll X Series monitor/defibrillator is a portable automated external defibrillator used in hospitals and clinics. The 34 reports mostly describe the unit failing to power on, showing a blank or white screen, rebooting on its own, or losing patient readings during routine checks and testing; 1 death and no injuries were reported, and most events involved no patient. Reports dropped sharply, from 8 in the previous 12 months to 2 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00847946020477), received between Aug 11, 2020 and Aug 18, 2026 — classified as 1 death, 33 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Power Up9
  • No Display/Image8
  • Unable to Obtain Readings7
  • Reset Problem3
  • Defibrillation/Stimulation Problem2
  • Failure of Device to Self-Test2
  • Failure to Analyze Signal1
  • Failure to Charge1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • Insufficient Information2
  • Facial Paralysis1
  • Necrosis1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2026Failure of Device to Self-Test

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50305858 ↗
  • MalfunctionAug 10, 2026Reset Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE RESTART/REBOOT ITSELF MULTIPLE TIMES.COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: Facial Paralysis

    MAUDE report 50192806 ↗
  • MalfunctionAug 7, 2025Use of Device Problem; Reset Problem; Use of Device Problem; Reset Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DO NOT USE ICON WAS DISPLAYED AND RESTART/REBOOT ITSELF MULTIPLE TIMES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22742262 ↗
  • MalfunctionJul 22, 2025Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22563985 ↗
  • MalfunctionJul 17, 2025No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY WAS COMPLETELY WHITE UPON POWERING ON. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22532820 ↗
  • MalfunctionJun 3, 2025No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22133414 ↗
  • MalfunctionApr 29, 2025Mechanics Altered; Mechanics Altered

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE SWITCHES SETTINGS ON IT'S OWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21928722 ↗
  • MalfunctionFeb 6, 2025Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A MALE 91-YEAR-OLD PATIENT, THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODE PADS AND DISPLAYED A "CHECK PADS" MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 21323722 ↗
  • MalfunctionJan 31, 2025Reset Problem; Reset Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE CONTINUOUSLY RESTART/REBOOT ITSELF. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21285260 ↗
  • MalfunctionJan 10, 2025Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21124039 ↗
  • MalfunctionSep 13, 2024Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20212397 ↗
  • MalfunctionJul 10, 2024Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19709411 ↗
  • MalfunctionJun 19, 2024No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY WAS WHITE WITH BLACK LINES AND WAS NOT LEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19569091 ↗
  • MalfunctionJan 31, 2024Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST FOR DEFIB AND PACER FUNCTIONS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18613884 ↗
  • MalfunctionNov 13, 2023Defibrillation/Stimulation Problem; Output Problem; Defibrillation/Stimulation Problem; Output Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION AND WOULD NOT GO INTO SYNC MODE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18118323 ↗
  • MalfunctionOct 26, 2023Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE FAILED TO DISCHARGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 18010815 ↗
  • MalfunctionSep 20, 2023No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17780616 ↗
  • MalfunctionJun 26, 2023No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17199069 ↗
  • MalfunctionMay 23, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16984659 ↗
  • MalfunctionMay 4, 2023No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16872705 ↗
  • MalfunctionFeb 21, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "ECG DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16408548 ↗
  • MalfunctionFeb 10, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO MONITOR A PATIENT (AGE & GENDER UNKNOWN) THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16357692 ↗
  • MalfunctionJan 26, 2023Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16248036 ↗
  • MalfunctionDec 20, 2022No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S SCREEN TURNED COMPLETELY WHITE AND WAS ILLEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16014512 ↗
  • DeathDec 7, 2022Failure to Charge; Unable to Obtain Readings; Failure to Charge; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A 75-YEAR-OLD MALE PATIENT, THE DEVICE WAS UNABLE TO DETECT THE ATTACHED ELECTRODES PADS. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED.

    Patient problems: Necrosis

    MAUDE report 15932451 ↗