Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850000509023), received between Nov 7, 2024 and Nov 7, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Improper Flow or Infusion” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper Flow or Infusion1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2024Improper Flow or Infusion; Improper Flow or Infusion

    IT WAS REPORTED, REGISTERED NURSE (RN) PROGRAMMED THE AMBIT PUMP (SN (B)(6), PN 220546) AS ORDERED BY THE ANESTHESIOLOGIST, WITH A DOSE 6 CC AND AN INTERVAL 120 MINUTES. A SECOND RN COMPLETED INDEPENDENT VERIFICATION OF WHAT WAS PROGRAMMED AND ATTACHED THE PUMP TO THE PERIPHERAL NERVE CATHETER AND INITIATED THE BOLUS. THE PUMP DISPLAYED 6 CC AND WAS PAUSED FOR A FEW SECONDS BEFORE BEING RELEASED. THE PUMP CONTINUED TO BOLUS BEYOND THE PRESCRIBED DOSE, IT WAS IMMEDIATELY PAUSED AGAIN, AND ANESTHESIOLOGIST WAS NOTIFIED. THE AMBIT PUMP WAS THEN DISCONNECTED FROM THE PATIENT AND UPON RELEASED FROM PAUSED AND NOTED BOLUS ADDITIONAL 6 CC OF AIR. IT DID NOT ALARM. THE DOCTORS REVIEWED THE TOTAL BOLUS AND WHAT WAS PROGRAMMED SETTINGS.

    Patient problems: Insufficient Information

    MAUDE report 20629224 ↗

Other Pump, Infusion devices

All Pump, Infusion devices →