Bigfoot Unity Diabetes Management System ASM-10027Recalls & reported problems
Bigfoot Biomedical, Inc. · Bigfoot Unity Cap for Humalog-Lispro-Lyumjev · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00850003506111), received between Jul 24, 2025 and Jul 25, 2025 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (13 reports).
1 other recall cover devices cleared under the same FDA submission (K202145); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch13
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22618292 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22612890 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22610716 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22610578 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22610108 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22606788 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22606726 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22606478 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22606429 ↗MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22593174 ↗MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22592643 ↗MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22592582 ↗MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22588447 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K202145. One submission often covers many models, so check the product and lot details in each record.
- Z-1997-2025 · May 21, 2025 · Open, ClassifiedBigfoot Unity Diabetes Management System SW-300689-01 Bigfoot Inject v1 Pen Cap LCAP Firmware 1.3.0 Release Configuration, Rev H SW-300689-02 Bigfoot Inject v1 Pen Cap RCAP Firmwar
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems devices
- FreeStyle Libre 2 71992Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 death23,063 reports
- FreeStyle Libre 3 72081Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems2 recalls16,326 reports
- FreeStyle Libre 2 72000Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems453 reports
- FreeStyle Libre 3 72080Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 recall274 reports
- FreeStyle Libre 2 71951Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems2 recalls72 reports
- FreeStyle Libre 3 App 72075-01Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 recall23 reports
- Bigfoot Unity Diabetes Management System ASM-10046Bigfoot Biomedical, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems1 recall12 reports
- Bigfoot Unity Diabetes Management System FG-300173Bigfoot Biomedical, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page