Zio ECG Utilization Software (ZEUS) System FG0401USRecalls & reported problems
Irhythm Technologies, Inc. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00850043907046), received between Dec 14, 2023 and Aug 20, 2024 — classified as 3 injuries, 17 malfunctions.
The most frequently reported specific device problem is “Inaccurate Information” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Inaccurate Information17
- Failure to Transmit Record6
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Unspecified Heart Problem20
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
InjuryAug 20, 2024Adverse Event Without Identified Device or Use Problem; Inaccurate Information; Adverse Event Without Identified Device or Use Problem; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE PATIENT RECEIVED A PACEMAKER DUE TO THE INITIAL FALSE POSITIVE INTERPRETATION. IT IS UNKNOWN WHETHER THE PACEMAKER WILL BE REMOVED.
Patient problems: Unspecified Heart Problem
MAUDE report 20006481 ↗InjuryJul 18, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE PATIENT RECEIVED A PACEMAKER DUE TO THE INITIAL INTERPRETATION.
Patient problems: Unspecified Heart Problem
MAUDE report 19778852 ↗MalfunctionJun 27, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 19628844 ↗MalfunctionJun 18, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 19563910 ↗MalfunctionMay 8, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 19271064 ↗MalfunctionApr 19, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND WAS TREATING IT. THERE WERE NO DELAYS IN TREATMENT, AND NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 19145790 ↗MalfunctionApr 10, 2024Failure to Transmit Record; Failure to Transmit Record
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HEALTHCARE PROVIDER (HCP) WAS IMMEDIATELY NOTIFIED, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND WAS TREATING IT. THERE WERE NO DELAYS IN TREATMENT, AND NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 19085481 ↗MalfunctionApr 5, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND OPTED NOT TO TREAT FOR THE ARRYTHMIA DUE TO THE HISTORY OF IT. THERE WERE NO DELAYS IN TREATMENT, AND NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 19051909 ↗InjuryMar 23, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HEALTHCARE PROVIDER (HCP) WAS IMMEDIATELY NOTIFIED, AND IRHYTHM LEARNED THAT THE PATIENT HAD RECEIVED A PACEMAKER DUE TO THE INITIAL INTERPRETATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Unspecified Heart Problem
MAUDE report 18965348 ↗MalfunctionMar 23, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HEALTHCARE PROVIDER (HCP) WAS IMMEDIATELY NOTIFIED, BUT IT WAS UNCLEAR WHETHER THE PATIENT WOULD RECEIVE TREATMENT. DESPITE SEVERAL ATTEMPTS TO GATHER MORE INFORMATION, NO ADDITIONAL DETAILS WERE OBTAINED. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18965339 ↗MalfunctionMar 12, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE PATIENT¿S CARE TEAM IS CONSIDERING THE TREATMENT OPTIONS. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18888056 ↗MalfunctionMar 7, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND WAS TREATING IT. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18862180 ↗MalfunctionMar 6, 2024Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THERE WERE NO DELAYS IN CARE, AND NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18851559 ↗MalfunctionFeb 8, 2024Failure to Transmit Record; Inaccurate Information; Failure to Transmit Record; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND WAS TREATING IT. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18672902 ↗MalfunctionFeb 5, 2024Failure to Transmit Record; Inaccurate Information; Failure to Transmit Record; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. IRHYTHM COMPLETED THREE GOOD FAITH ATTEMPTS TO THE ACCOUNT TO OBTAIN ADDITIONAL INFORMATION, BUT WITH NO RESULT. IT IS UNKNOWN WHETHER THE PATIENT RECEIVED ANY TREATMENT FOR THE ARRHYTHMIA.
Patient problems: Unspecified Heart Problem
MAUDE report 18643264 ↗MalfunctionFeb 2, 2024Failure to Transmit Record; Failure to Transmit Record
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT TRANSMITTED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED A POTENTIAL ALGORITHM SENSITIVITY ISSUE WITH THE DEVICE. THE HEALTHCARE PROVIDER (HCP) WAS IMMEDIATELY NOTIFIED, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA. THERE WERE NO DELAYS IN TREATMENT, AND NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18636754 ↗MalfunctionJan 31, 2024Failure to Transmit Record; Inaccurate Information; Failure to Transmit Record; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. IRHYTHM COMPLETED THREE GOOD FAITH ATTEMPTS TO THE ACCOUNT TO OBTAIN ADDITIONAL INFORMATION, BUT WITH NO RESULT. IT IS UNKNOWN WHETHER THE PATIENT RECEIVED ANY TREATMENT FOR THE ARRHYTHMIA.
Patient problems: Unspecified Heart Problem
MAUDE report 18619512 ↗MalfunctionJan 3, 2024Failure to Transmit Record; Failure to Transmit Record
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT TRANSMITTED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED A POTENTIAL ALGORITHM SENSITIVITY ISSUE WITH THE DEVICE. THE HEALTHCARE PROVIDER (HCP) WAS IMMEDIATELY NOTIFIED OF THE PATIENT'S ARRHYTHMIA, AND IRHYTHM WAS INFORMED THAT THE PATIENT IS NOT COMPLIANT WITH HIS MEDICAL RECOMMENDATIONS. THERE WERE NO DELAYS IN TREATMENT, AND NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18440054 ↗MalfunctionDec 19, 2023Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION PROVIDED TO THE PHYSICIAN WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND WAS TREATING IT. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18354849 ↗MalfunctionDec 14, 2023Inaccurate Information; Inaccurate Information
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT COMMUNICATED DURING THE WEAR PERIOD. THE INVESTIGATION REVEALED THAT A PRELIMINARY ECG INTERPRETATION WAS MISCLASSIFIED. FOLLOWING THE WEAR PERIOD AND WHILE COMPILING THE FINAL REPORT, THE INTERPRETATION WAS AMENDED. THE HCP WAS NOTIFIED IMMEDIATELY, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND WAS TREATING IT. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Patient problems: Unspecified Heart Problem
MAUDE report 18331020 ↗
Other Computer, Diagnostic, Programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page