Skip to content

Vanguard 6200x135Recalls & reported problems

Contego Medical, Inc. · The Vanguard IEP Peripheral Balloon Angioplasty System with Integrated Embolic Protection is indicated for percutaneous transluminal angioplasty (PTA) and capture and removal of embolic material during angioplasty, for the femoral, iliac, popliteal and profunda arteries. The System is not intended for use in the renal, cerebral, coronary or carotid vasculature. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00851616007149), received between Aug 9, 2024 and Aug 9, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • Unspecified Vascular Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 9, 2024Mechanical Problem; Mechanical Problem

    ONE PATIENT INVOLVED. DR (B)(6) WAS TREATING A DIFFUSIVELY DISEASED SFA WITH SIGNIFICANT CALCIUM IN A PATIENT WITH SINGLE VESSEL RUNOFF HE DECIDED TO USE A VANGUARD 6X200MM TO PREVENT ANY DISTAL EMBOLI HE DID TWO SEPARATE INFLATIONS IN THE DISEASED VESSEL AFTER HE WAS SATISFIED WITH THE RESULT WAS FULLY DEFLATED, HE CLOSED THE FILTER AND BEGAN TO WITHDRAW THE DEVICE-AS FILTER MET THE 6FR SHEATH HE GOT SIGNIFICANT RESISTANCE COMING INTO THE SHEATH. -DR (B)(6) REPOSITIONED THE CATHETER AND SHEATH A FEW TIMES UNDER FLUORO AND WAS STILL HAVING TROUBLE PULLING CATHETER INTO THE SHEATH AT SOME POINT HE REALIZED THE FILTER WAS DEFORMED AND COULD NOT EXIT THE SHEATH. HE DECIDED TO PULL THE ENTIRE SHEATH AND VANGUARD CATHETER OUT THROUGH THE ACCESS SITE. HE WAS UNABLE TO DO SO AND HE ALSO NOTICED AT THIS POINT THE FILTER HAD DETACHED FROM THE CATHETER DR (B)(6) WENT IN AND TRIED SEVERAL MANEUVERS TO GET THE NOW DETACHED FILTER OUT OF THE VESSEL TO NO AVAIL. AT THIS POINT HE HAD DONE ENOUGH TO HAVE THE FILTER, WHICH WAS NOW JUST A WIRE PRESSED AGAINST THE WALL OF THE COMMON FEMORAL, NOT OBSTRUCTING FLOW AT ALL.-DR (B)(6) LEFT THIS AND HIS PLAN AS HE STATED WAS TO MOST LIKELY GO BACK IN WHEN

    Patient problems: Unspecified Vascular Problem

    MAUDE report 19956803 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →

Sources and updates