t:slim Insulin Pump, Reconditioned 007444Recalls & reported problems
Tandem Diabetes Care, Inc. · Insulin Pump, Reconditioned 4YR Warranty, t:slim · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This reconditioned Tandem t:slim insulin pump delivers insulin. Many of the 284 reports describe the pump stopping, blocked flow, and battery or charging problems, and some describe high blood sugar in users. The reports include no deaths and 40 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 284
- Deaths reported
- 0
- Injuries reported
- 40
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 284 adverse event reports that list this exact device (DI 00852162004378), received between Jul 6, 2018 and Oct 11, 2024 — classified as 40 injuries, 244 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (55 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Pumping Stopped55
- Obstruction of Flow36
- Battery Problem31
- Incorrect Measurement29
- Charging Problem27
- Adverse Event Without Identified Device or Use Problem22
- Device Displays Incorrect Message22
- Device Sensing Problem16
Patient problems
- No Consequences Or Impact To Patient150
- No Known Impact Or Consequence To Patient52
- No Clinical Signs, Symptoms or Conditions46
- Hyperglycemia25
- Hypoglycemia10
- Elevated Ketones/ Diabetic Ketoacidosis5
- Insufficient Information2
- Loss of consciousness2
Recent adverse event reports
The 25 most recent of 284 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 11, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20427392 ↗MalfunctionDec 28, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE EFFECT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT; THEREFORE THE ALLEGATION COULD NOT BE CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18410044 ↗MalfunctionJun 9, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PUMP WAS NOT WORKING PROPERLY. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT; THEREFORE THE ALLEGATION COULD NOT BE CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17097215 ↗MalfunctionJul 13, 2022Failure to Sense; Failure to Sense
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS INTERMITTENTLY UNRESPONSIVE. CUSTOMER DECLINED TROUBLESHOOTING FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15000211 ↗MalfunctionJun 14, 2022Failure to Sense; Obstruction of Flow; Failure to Sense; Obstruction of Flow
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE AND THAT AN OCCLUSION ALARM OCCURRED. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14685627 ↗InjuryApr 27, 2022Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
IT WAS REPORTED THAT THE PUMP WAS NOT DELIVERING INSULIN AS INTENDED WHICH RESULTED IN CUSTOMER BECOMING UNCONSCIOUS. MULTIPLE ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO FOLLOW-UP WITH THE CUSTOMER ON THE REPORTED ISSUE, BUT NO RESPONSE WAS RECEIVED.
Patient problems: Insufficient Information; Loss of consciousness
MAUDE report 14218799 ↗InjuryApr 13, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. CUSTOMER CHANGED THE PUMP SUPPLIES TO ADDRESS THE ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS "HIGH" ON THE CONTINUOUS GLUCOSE MONITOR (CGM), WITH MODERATE KETONES. ELEVATED BLOOD GLUCOSE LEVELS WERE ADDRESSED BY MANUAL INSULIN INJECTION. REPORTEDLY, THE CUSTOMER REVERTED TO AN ALTERNATE METHOD INSULIN THERAPY.
Patient problems: Hyperglycemia
MAUDE report 14101499 ↗MalfunctionFeb 4, 2022Failure to Sense; Device Sensing Problem; Failure to Sense; Device Sensing Problem
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE AND THE WAKE BUTTON WAS STICKING. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13453762 ↗MalfunctionDec 28, 2021Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. REPORTED THE CUSTOMER IS NOT PERFORMING THE AIR REMOVAL STEP PER USER GUIDE. CLINCIAL SUPPORT INFORMED CUSTOMER THAT NOT PERFORMING THE AIR REMOVAL STEP PER USER GUIDE IS OFF LABLE. CUSTOMER CHANGED THE PUMP SUPPLIES AND RESUMED INSULIN DELIVERY. CUSTOMER BLOOD GLUOCASE WAS 300 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13095439 ↗InjuryDec 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED IN THE INTENSIVE CARE UNIT DUE TO A BLOOD GLUCOSE (BG) LEVEL OF 1341MG/DL. THE CONTACT WAS UNABLE TO PROVIDE ADDITIONAL INFORMATION REGARDING THE REPORTED ISSUE. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT; HOWEVER, NO RESPONSE WAS RECEIVED FROM THE CUSTOMER.
Patient problems: Hyperglycemia
MAUDE report 13082539 ↗InjuryDec 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED; DETAILS AND NATURE OF HOSPITALIZATION WERE NOT PROVIDED. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT; HOWEVER, NO RESPONSE WAS RECEIVED FROM THE CUSTOMER.
Patient problems: Insufficient Information
MAUDE report 13065095 ↗MalfunctionDec 2, 2021Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO BLOOD GLUCOSE. CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12919029 ↗MalfunctionNov 18, 2021Obstruction of Flow; Date/Time-Related Software Problem; Battery Problem; Obstruction of Flow; Date/Time-Related Software Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY, PUMP TIME AND DATE WERE INCORRECT, AND MULTIPLE INTERMITTENT OCCLUSION ALARMS OCCURRED. NO FURTHER INFORMATION WAS OBTAINED AS CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED IMPACT TO BLOOD GLUCOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12837015 ↗MalfunctionOct 1, 2021Failure to Sense; Failure to Sense
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS CRACKED AND UNRESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED TO TROUBLESHOOT WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12561263 ↗MalfunctionSep 22, 2021Incorrect Measurement; Battery Problem; Charging Problem; Incorrect Measurement; Battery Problem; Charging Problem
IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY AND COULD NOT BE CHARGED WITHOUT HOLDING THE CHARGING CABLE IN PLACE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 168-270 MG/DL. REPORTEDLY, THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12516163 ↗InjurySep 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER WENT TO THE EMERGENCY ROOM AND WAS SUBSEQUENTLY HOSPITALIZED DUE TO A LOW BLOOD GLUCOSE (BG) LEVEL OF 12 MG/DL. CAUSE OF LOW BG WAS UNKNOWN. CUSTOMER WAS TREATED WITH SUGAR WATER AND INTRAVENOUS FLUIDS OF SALINE AND INSULIN. REPORTEDLY, LOW BG RESULTED IN UNCONSCIOUSNESS AND NERVE DAMAGE IN CUSTOMER'S FOOT. CUSTOMER WAS RELEASED FROM THE HOSPITAL 15 DAYS LATER.
Patient problems: Hypoglycemia; Loss of consciousness; Nerve Damage
MAUDE report 12516148 ↗MalfunctionSep 7, 2021Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO BLOOD GLUCOSE. CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12434524 ↗MalfunctionAug 27, 2021Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO ADVERSE IMPACT TO BLOOD GLUCOSE. CUSTOMER DECLINED FURTHER TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12378920 ↗MalfunctionAug 26, 2021Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS APPROXIMATELY 160 MG/DL. CUSTOMER DECLINED FURTHER TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT TO RESOLVE THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12373820 ↗InjuryAug 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER REQUIRED ASSISTANCE TO TREAT A LOW BLOOD GLUCOSE LEVEL AND AN ELEVATED BLOOD GLUCOSE LEVEL (VALUE NOT PROVIDED). NO ADDITIONAL INFORMATION WAS PROVIDED. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT; HOWEVER, NO RESPONSE WAS RECEIVED FROM THE CUSTOMER.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 12345329 ↗MalfunctionAug 19, 2021Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO BLOOD GLUCOSE. CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12344413 ↗MalfunctionAug 18, 2021Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. SYSTEM CHECK WAS PERFORMED WITH TANDEM TECHNICAL SUPPORT AND THE ROOT CAUSE COULD NOT BE DETERMINED. CUSTOMER CHANGED PUMP SUPPLIES AND SUCCESSFULLY RESUMED INSULIN DELIVERY. THERE WAS NO ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12337971 ↗MalfunctionAug 12, 2021Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING. ADDITIONALLY, THE WAKE BUTTON WAS UNRESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE. REPORTEDLY, THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12314074 ↗MalfunctionJul 30, 2021Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 305 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12252802 ↗MalfunctionJul 28, 2021Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT THE PUMP STOPPED DELIVERING INSULIN. CUSTOMER REVERTED TO MANUAL INJECTIONS FOR INSULIN THERAPY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT AND DECLINED TO PROVIDE FURTHER INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12239673 ↗
Other Pump, Infusion, Insulin devices
- Omnipod DASH Pod POD-BLE-11-520Insulet Corporation · Pump, Infusion, Insulin1 recall30 deaths56,209 reports
- Omnipod ZXP420Insulet Corporation · Pump, Infusion, Insulin1 recall17 deaths37,832 reports
- Omnipod ZXP420Insulet Corporation · Pump, Infusion, Insulin70 deaths20,092 reports
- t:slim X2 Insulin Pump 1000103Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin1 recall1 death7,327 reports
- t:slim Insulin Pump, Reconditioned Replacement 007262Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin2 deaths5,253 reports
- t:flex Insulin Pump, Replacement, Refurbished 007473Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin2 deaths2,373 reports
- t:slim Insulin Pump 007435Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin7 deaths2,104 reports
- Accu-Chek Combo 05027250001Roche Diabetes Care GmbH · Pump, Infusion, Insulin1,830 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.