Omnipod ZXP420 (REF 19191)Recalls & reported problems

Insulet Corporation · Omnipod Pod · Class 2

All models of Omnipod (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is an Omnipod pod. Its reports mostly describe pods that bend, shift or fail to connect or deploy, often with high blood glucose. There are 37,832 reports, including 17 deaths and 2,855 injuries. One recall, started in May 2026, is open: cannulas damaged during manufacturing can leak insulin around the pod. Reports dropped to 123 in the last 12 months, from 367 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37,831
Deaths reported
17
Injuries reported
2,856
Recalls naming it
1

Summary

The FDA's MAUDE database contains 37,831 adverse event reports that list this exact device (DI 10385081120036), received between Jul 25, 2018 and Sep 2, 2026 — classified as 17 deaths, 2,856 injuries, 34,958 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (12,016 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 116 reportsJan 2024: 119 reportsFeb 2024: 11 reportsMar 2024: 60 reportsApr 2024: 55 reportsMay 2024: 73 reportsJun 2024: 75 reportsJul 2024: 77 reportsAug 2024: 71 reportsSep 2024: 64 reportsOct 2024: 65 reportsNov 2024: 43 reportsDec 2024: 49 reportsJan 2025: 31 reportsFeb 2025: 29 reportsMar 2025: 34 reportsApr 2025: 18 reportsMay 2025: 20 reportsJun 2025: 47 reportsJul 2025: 33 reportsAug 2025: 16 reportsSep 2025: 18 reportsOct 2025: 29 reportsNov 2025: 40 reportsDec 2025: 12 reportsJan 2026: 20 reportsFeb 2026: 21 reportsMar 2026: 5 reportsApr 2026: 4 reportsMay 2026: 6 reportsJun 2026: 5 reportsJul 2026: 4 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026119
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 20, 2026Class IOpen, Classified

    External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

    Z-2605-2026 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Twisted/Bent12,016
  • Unintended Movement7,797
  • Loss of or Failure to Bond4,706
  • Insufficient Device Problem Information4,403
  • Activation, Positioning or Separation Problem3,355
  • Failure to Fire3,022
  • Difficult to Insert2,472
  • Fluid/Blood Leak2,328

Patient problems

  • Hyperglycemia26,608
  • No Clinical Signs, Symptoms or Conditions5,569
  • No Information4,164
  • Pain815
  • Elevated Ketones/ Diabetic Ketoacidosis647
  • Skin Inflammation/ Irritation505
  • Vomiting487
  • Irritation341

Recent adverse event reports

The 25 most recent of 37,831 reports, in the reporter's own words as published by the FDA.

  • InjurySep 2, 2026Insufficient Device Problem Information

    It was reported by the patient's spouse that, following removal of an initial Pod due to persistent hyperglycemia, a second Pod was applied. The patient's blood glucose levels remained above 400 mg/dL while wearing the Pod between 5 and 24 hours. Symptoms reported include headache and nausea, As a result, the patient sought medical attention on (b)(6) 2026 at (b)(6), where diagnosis of diabetic ketoacidosis (DKA) was made. The patient was treated in the Intensive Care Unit (ICU) with intravenous insulin. The patient remained hospitalized for five days, and was discharged on (b)(6) 2026. The first Pod reported is captured in complaint (b)(4).

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Headache; Hyperglycemia; Nausea

    MAUDE report 50571569 ↗
  • InjurySep 2, 2026Insufficient Device Problem Information

    It was reported by the patient's spouse that the patient was hospitalized with diabetic ketoacidosis (DKA) on 29 June 2026. The patient's blood glucose levels rose to above 400 mg/dL while wearing the Pod between 25 and 36 hours. The patient experienced headache and nausea. The reporter stated that the Pod had stopped delivering insulin however. After a second Pod was reportedly unsuccessful in lowering blood glucose levels, the patient sought medical attention at (b)(6), where diagnosis of diabetic ketoacidosis was made. The patient was treated in the Intensive Care Unit (ICU) for two days with intravenous insulin and remained hospitalized for five days. The patient was discharged on (b)(6) 2026. The Pod was removed prior to hospitalization. The second Pod reported is captured in complaint (b)(4).

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Headache; Hyperglycemia; Nausea

    MAUDE report 50567675 ↗
  • MalfunctionAug 19, 2026Premature Activation

    It was reported the needle mechanism had fully deployed before the Pod was able to be used. No impact to the patient's glucose level was reported. The Pod was discarded, and a new Pod was placed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50328894 ↗
  • MalfunctionAug 18, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50307971 ↗
  • MalfunctionAug 3, 2026Loss of or Failure to Bond; Unintended Movement

    It was reported that the patient's blood glucose level rose to greater than 250 mg/dL, while wearing the Pod between 5 and 24 hours. The Pod reportedly detached from the abdomen, causing the cannula to dislodge. As treatment, a new pod was placed.

    Patient problems: Hyperglycemia

    MAUDE report 50116922 ↗
  • InjuryJul 9, 2026Insufficient Device Problem Information

    IT WAS REPORTED BY THE PATIENT'S PARTNER THAT THE PATIENT HAD BEEN HOSPITALIZED WITH HYPERGLYCEMIA ON (B)(6) 2026. THE PATIENT'S BLOOD GLUCOSE LEVELS REACHED 53 MMOL/L (954 MG/DL) WHILE WEARING THE POD BETWEEN 24 AND 36 HOURS ON THE ARM; THE PATIENT'S USUAL BLOOD GLUCOSE RANGE WAS REPORTED TO BE APPROXIMATELY 7 MMOL/L (126 MG/DL). THE PATIENT HAD REPORTEDLY BEEN FEELING UNWELL THROUGHOUT THE DAY AND EXPERIENCED VOMITING, DIZZINESS, AND WEAKNESS. THE PATIENT'S PARTNER REPORTED THAT THE PATIENT WAS ADMITTED TO THE INTENSIVE CARE UNIT (ICU) ON THE EVENING OF (B)(6) 2026. THE REPORTED DIAGNOSIS WAS HIGH GLUCOSE LEVELS. TREATMENT REPORTEDLY INCLUDED ADMINISTRATION OF INTRAVENOUS FLUIDS AND INSULIN. THE PATIENT WAS STILL HOSPITALIZED AT THE TIME OF REPORTING, AND THEREFORE A DISCHARGE DATE WAS NOT AVAILABLE. ADDITIONAL INFORMATION REGARDING THE DATE OF DISCHARGE, HOSPITALIZATION OUTCOME, AND ANY FURTHER MEDICAL ASSESSMENT OF THE RELATIONSHIP BETWEEN THE EVENT AND THE OMNIPOD SYSTEM IS BEING REQUESTED. THE PATIENT CONTINUED TO WEAR THE POD AT THE TIME OF THE REPORT.

    Patient problems: Dizziness; Fatigue; Hyperglycemia; Vomiting

    MAUDE report 25762696 ↗
  • InjuryJul 9, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL DROPPED TO AROUND 40 MG/DL WHILE WEARING THE POD. THE PATIENT REPORTED THEY WERE INVOLVED IN A CAR ACCIDENT DUE TO LOW BLOOD GLUCOSE LEVELS BETWEEN LATE (B)(6) 2029 AND EARLY (B)(6) 2019. THE PATIENT WAS TRANSPORTED TO A HOSPITAL WHERE THE PATIENT STAYED FOR 20 HOURS. THE PATIENT REPORTED THEY WERE NOT INJURED, BUT ONLY WOKE AFTER THEY REACHED THE HOSPITAL. THE PATIENT'S POD WAS REMOVED AS IT WAS DAMAGED IN THE ACCIDENT, THE PATIENT'S BLOOD GLUCOSE LEVEL WAS MANAGED USING MANUAL INSULIN INJECTION USING SYRINGES WHILE IN THE HOSPITAL. THE PATIENT'S LOW BLOOD GLUCOSE LEVELS WERE TREATED USING 'LIQUID SUGAR' THROUGH INJECTION. THE PATIENT IS FROM THE UNITED STATES AND WAS IN AN REMOTE ISLAND IN GREECE DURING THE INCIDENT.

    Patient problems: Hypoglycemia; Loss of consciousness

    MAUDE report 25760268 ↗
  • MalfunctionJul 1, 2026Failure to Fire

    IT WAS REPORTED BY LEGAL GUARDIAN THAT THE PINK SLIDE DID NOT MOVE FORWARD, IS NOT VISIBLE IN THE VIEWING WINDOW AND WHEN THE POD WAS REMOVED THE CANNULA WAS NOT VISIBLE. NO IMPACT TO THE PATIENT'S GLUCOSE LEVEL WAS REPORTED. THE POD WAS NOT WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25679866 ↗
  • MalfunctionJul 1, 2026Failure to Fire

    IT WAS REPORTED BY LEGAL GUARDIAN THAT THE PINK SLIDE DID NOT MOVE FORWARD, IS NOT VISIBLE IN THE VIEWING WINDOW AND WHEN THE POD WAS REMOVED THE CANNULA WAS NOT VISIBLE. NO IMPACT TO THE PATIENT'S GLUCOSE LEVEL WAS REPORTED. THE POD WAS NOT WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25679802 ↗
  • MalfunctionJun 18, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 28 MMOL/L (504 MG/DL) WHILE WEARING THE POD FOR AN UNKNOWN DURATION. THE DEVICE WAS ORIGINALLY REPORTED FOR POD CUSTOMER NOT SURE DELIVERING INSULIN. INVESTIGATION OF THE DEVICE FOUND A TEAR IN THE EXPOSED PORTION OF THE CANNULA AND THE TIP OF THE NEEDLE TO BE DULL.

    Patient problems: Hyperglycemia

    MAUDE report 25556356 ↗
  • MalfunctionJun 12, 2026Difficult to Insert

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 20 MMOL/L (360 MG/DL) WHILE WEARING THE POD FOR AN UNSPECIFIED AMOUNT OF TIME. THE PATIENT REPORTED BEING UNSURE IF THE CANNULA WAS PROPERLY SEATED IN THE INFUSION SITE (HIP/BUTTOCKS). AS TREATMENT, A NEW POD WAS APPLIED.

    Patient problems: Hyperglycemia

    MAUDE report 25487585 ↗
  • InjuryJun 12, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT SOUGHT MEDICAL ATTENTION ON (B)(6) 2026 WITH SKIN IRRITATION WITH PURULENT DISCHARGE AT THE INFUSION SITE (LEG) WHILE WEARING THE POD BETWEEN 25 AND 36 HOURS. THE PATIENT¿S BLOOD GLUCOSE LEVELS ROSE TO 400 MG/DL AT THIS TIME. AS TREATMENT, THE PATIENT WAS PRESCRIBED AN UNSPECIFIED TOPICAL OINTMENT.

    Patient problems: Hyperglycemia; Purulent Discharge; Skin Inflammation/ Irritation

    MAUDE report 25483039 ↗
  • MalfunctionJun 9, 2026Loss of or Failure to Bond; Unintended Movement

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO GREATER THAN 250 MG/DL WHILE WEARING THE POD WORN LESS THAT 1 HOUR. THE ADHESIVE COMPLETELY SEPARATED FROM THE (LEG) CAUSING THE CANNULA TO DISLODGE. AS TREATMENT, A NEW POD WAS INSERTED.

    Patient problems: Hyperglycemia

    MAUDE report 25451752 ↗
  • MalfunctionJun 9, 2026Loss of or Failure to Bond; Unintended Movement

    NOT APPLICABLE AS THE EVENT IS NOT REPORTABLE.

    Patient problems: Hyperglycemia

    MAUDE report 25451740 ↗
  • MalfunctionMay 13, 2026Premature Activation

    IT WAS REPORTED THE NEEDLE MECHANISM HAD FULLY DEPLOYED BEFORE THE POD WAS ABLE TO BE USED. NO IMPACT TO THE PATIENT'S GLUCOSE LEVEL WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25163023 ↗
  • MalfunctionMay 12, 2026Failure to Fire

    IT WAS REPORTED BY THE MOTHER OF A PATIENT THAT THE NEEDLE DID NOT MOVE FORWARD WHEN THE POD WAS ACTIVATED; THIS INDICATES A NEEDLE MECHANISM FAILURE. THE POD WAS NOT WORN. NO IMPACT TO THE PATIENT'S GLUCOSE LEVEL WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25148415 ↗
  • MalfunctionMay 8, 2026Loss of or Failure to Bond; Unintended Movement

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO GREATER THAN 250 MG/DL WHILE WEARING THE POD. THE ADHESIVE COMPLETELY SEPARATED CAUSING THE CANNULA TO DISLODGE. AS TREATMENT, A NEW POD WAS INSERTED.

    Patient problems: Hyperglycemia

    MAUDE report 25131268 ↗
  • InjuryMay 8, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT, ON (B)(6) 2024, THE PATIENT WAS HOSPITALISED AT (B)(6), INCLUDING ADMISSION TO THE INTENSIVE CARE UNIT, WITH A DIAGNOSIS OF A BASAL GANGLIA HAEMORRHAGE. THE PATIENT WAS USING OMNIPOD EROS SYSTEM AT THE TIME OF THE EVENT BUT IT WAS IMMEDIATELY DISPOSED OF ON ADMISSION. THE NAME OF THE HEALTHCARE PROFESSIONAL WHO TREATED THE PATIENT WAS NOT RECALLED BY THE COMPLAINANT. TREATMENT THE PATIENT RECEIVED INCLUDED THROMBO ASS, CEREBOKAN AND PRONERV. THE PATIENT'S WIFE STATED THE EVENT WAS NOT LINKED TO A DEVICE FAILURE BUT LIKELY A COMPLICATION FROM A LONGSTANDING DIAGNOSIS OF DIABETES. THE PATIENT RESUMED OMNIPOD SYSTEM USE IN (B)(6) 2025. AS A RESULT OF THE HAEMORRHAGE, THE PATIENT HAS ONGOING COGNITIVE IMPAIRMENT AND IMPAIRED MOTOR FUNCTION. THE PATIENT WAS HOSPITALISED FOR A TOTAL OF 37 DAYS ((B)(6) 2025).

    Patient problems: Cognitive Changes; Stroke/CVA

    MAUDE report 25124631 ↗
  • MalfunctionMay 4, 2026Loss of or Failure to Bond; Unintended Movement

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO GREATER THAN 250 MG/DL WHILE WEARING THE POD BETWEEN 5 AND 24 HOURS. THE ADHESIVE COMPLETELY SEPARATED FROM THE (ABDOMEN) CAUSING THE CANNULA TO DISLODGE. AS TREATMENT, A NEW POD WAS INSERTED.

    Patient problems: Hyperglycemia

    MAUDE report 25075880 ↗
  • InjuryMay 1, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION AT THE POD¿S INSERTION SITE WHILE WEARING THE POD. THE PATIENT REPORTED THE INFUSION SITE TO BE VERY RED WITH BLISTERING, ITCHINESS, AND DRAINAGE CONSISTING OF WOUND FLUID, WITH NO BLEEDING NOTED. DUE TO CONCERN FOR INFECTION AND A KNOWN SEVERE ADHESIVE ALLERGY, THE PATIENT REMOVED THE POD AND SOUGHT MEDICAL ATTENTION FROM AN INTERNIST. UPON EXAMINATION, THE INTERNIST CLEANED THE SITE WITH ALCOHOL AND DIAGNOSED A SKIN INFECTION ASSOCIATED WITH AN ADHESIVE REACTION. THE PATIENT WAS PRESCRIBED ORAL ANTIBIOTICS FLUTICASON. FOR FUTURE POD USE, THE INTERNIST PRESCRIBED FLUTICASONE SPRAY TO APPLY TO THE SKIN PRIOR TO POD PLACEMENT TO REDUCE ALLERGIC REACTIONS. THE PATIENT REPORTS THIS TREATMENT HAS EFFECTIVELY PREVENTED FURTHER INFECTIONS, WITH ONLY MILD REDNESS OCCURRING IF THE SPRAY LAYER IS TOO THIN. A NEW POD WAS PLACED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Erythema; Itching Sensation; Purulent Discharge; Skin Infection; Skin Inflammation/ Irritation; Urticaria

    MAUDE report 25058440 ↗
  • MalfunctionApr 1, 2026Failure to Fire

    IT WAS REPORTED BY XXX THAT THE NEEDLE DID NOT MOVE FORWARD WHEN THE POD WAS ACTIVATED; THIS INDICATES A NEEDLE MECHANISM FAILURE. THE POD WAS NOT WORN. NO IMPACT TO THE PATIENT'S GLUCOSE LEVEL WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24757836 ↗
  • MalfunctionApr 1, 2026Fluid/Blood Leak; Material Twisted/Bent

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO BETWEEN 16.7 MMOL/L AND 22.2 MMOL/L (300 MG/DL AND 400 MG/DL) WHILE WEARING THE POD BETWEEN 5 AND 24 HOURS. THE PATIENT REPORTED INSULIN LEAKING WHILE WEARING THE POD. WHEN REMOVED FROM THE INFUSION SITE (ABDOMEN), THE POD'S CANNULA WAS FOUND BENT. AS TREATMENT, A NEW POD WAS APPLIED.

    Patient problems: Hyperglycemia

    MAUDE report 24757693 ↗
  • MalfunctionApr 1, 2026Material Twisted/Bent

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO GREATER THAN 13.9 MMOL/L (250 MG/DL) WHILE WEARING THE POD BETWEEN 5 AND 24 HOURS. WHEN REMOVED FROM THE INFUSION SITE (ABDOMEN), THE POD'S CANNULA WAS FOUND BENT. AS TREATMENT, A NEW POD WAS APPLIED.

    Patient problems: Hyperglycemia

    MAUDE report 24757638 ↗
  • MalfunctionApr 1, 2026Fluid/Blood Leak; Material Twisted/Bent

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 22.2 MMOL/L (400 MG/DL) WHILE WEARING THE POD BETWEEN 5 AND 24 HOURS. THE PATIENT REPORTED INSULIN LEAKING WHILE WEARING THE POD. WHEN REMOVED FROM THE INFUSION SITE (ABDOMEN), THE POD'S CANNULA WAS FOUND BENT. AS TREATMENT, A NEW POD WAS APPLIED.

    Patient problems: Hyperglycemia

    MAUDE report 24757279 ↗
  • MalfunctionMar 23, 2026Difficult to Insert; Material Twisted/Bent

    IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 30 MMOL/L (540 MG/DL) WHILE WEARING THE POD BETWEEN 1 AND 4 HOURS. THE PATIENT REPORTED BEING UNSURE IF THE CANNULA WAS PROPERLY SEATED IN THE INFUSION SITE (UNSPECIFIED). WHEN THE POD WAS REMOVED THE CANNULA WAS FOUND TO BE BENT AND THERE WAS BLEEDING FROM THE INFUSION SITE. AS TREATMENT, A NEW POD WAS APPLIED.

    Patient problems: Hyperglycemia

    MAUDE report 24663798 ↗

FDA clearance and classification

The FDA regulates product code LZG (pump, infusion, insulin) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Omnipod ZXP420

Has the Omnipod ZXP420 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2605-2026 (Class I), began on May 20, 2026: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Omnipod ZXP420?

The FDA's MAUDE database lists 37,831 adverse event reports for this device, received between Jul 25, 2018 and Sep 2, 2026. In those reports, the most common specific device problems are “Material Twisted/Bent” (12,016), “Unintended Movement” (7,797) and “Loss of or Failure to Bond” (4,706); the most common patient problems are “Hyperglycemia” (26,608), “Pain” (815) and “Elevated Ketones/ Diabetic Ketoacidosis” (647). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Omnipod ZXP420?

17 reports are classified as deaths, 2,856 as injuries and 34,958 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Omnipod ZXP420 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Omnipod ZXP420, and is it still sold?

The FDA's device database lists Insulet Corporation as the company responsible for it (brand Omnipod, model ZXP420). Its record shows it as in commercial distribution.