t:flex Insulin Pump, Replacement, Refurbished 007473Recalls & reported problems

Tandem Diabetes Care, Inc. · Assembly, Pump Replacement, Refurbished, t:flex · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Tandem t:flex insulin pump is a refurbished replacement pump used to deliver insulin. Its 2,373 reports often describe the pump stopping, inaccurate insulin readings, blocked flow, and battery or charging problems, and mention 201 injuries and 2 deaths. They also mention high or low blood sugar, and in some cases elevated ketones. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,373
Deaths reported
2
Injuries reported
201
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,373 adverse event reports that list this exact device (DI 00852162004569), received between Jul 9, 2018 and Jan 8, 2025 — classified as 2 deaths, 201 injuries, 2,170 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (458 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 5 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement458
  • Pumping Stopped430
  • Obstruction of Flow345
  • Charging Problem253
  • Battery Problem250
  • Failure to Sense145
  • False Alarm138
  • Adverse Event Without Identified Device or Use Problem113

Patient problems

  • No Consequences Or Impact To Patient1,045
  • No Clinical Signs, Symptoms or Conditions614
  • No Known Impact Or Consequence To Patient521
  • Hyperglycemia120
  • Hypoglycemia76
  • Elevated Ketones/ Diabetic Ketoacidosis22
  • Dizziness3
  • Death2

Recent adverse event reports

The 25 most recent of 2,373 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2025Overheating of Device; Device Sensing Problem; Overheating of Device; Device Sensing Problem

    IT WAS REPORTED THAT THE PUMP WAS WARM TO THE TOUCH DESPITE BEING IN ROOM-TEMPERATURE CONDITIONS. REPORTEDLY, THE PUMP WAS CHARGING DURING THE EVENT. ADDITIONALLY, IT WAS REPORTED THAT THE WAKE BUTTON WAS NOT WORKING. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21103984 ↗
  • MalfunctionNov 6, 2024Pumping Stopped; Failure to Sense; Battery Problem; Pumping Stopped; Failure to Sense; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY RESULTING IN THE PUMP SHUTTING OFF. ADDITIONALLY, THE PUMP TOUCHSCREEN WAS UNRESPONSIVE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20620199 ↗
  • MalfunctionSep 24, 2024Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. REPORTEDLY THE CUSTOMER IS USING U-500 INSULIN. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20296589 ↗
  • InjurySep 23, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS OVER 600 MG/DL. AT THE TIME OF THE REPORT, THE CUSTOMER HAD NOT ADDRESSED BG LEVEL AND WENT TO THE EMERGENCY ROOM. MULTIPLE ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO FOLLOW-UP WITH THE CUSTOMER; HOWEVER, NO RESPONSE WAS RECEIVED, AND NO ADDITIONAL INFORMATION WAS ABLE TO BE OBTAINED.

    Patient problems: Hyperglycemia

    MAUDE report 20287563 ↗
  • MalfunctionSep 11, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT THE PUMP SHUT OFF UNEXPECTEDLY. THE CUSTOMER DECLINED FURTHER TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20189119 ↗
  • MalfunctionAug 23, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT AN UNSPECIFIED MALFUNCTION ALARM OCCURRED. REPORTEDLY, THE CUSTOMER CONTACTED THE HEALTHCARE PROVIDER TO OBTAIN ALTERNATE INSULIN THERAPY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20051954 ↗
  • MalfunctionAug 22, 2024Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. THERE WAS NO REPORTED ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVELS. CUSTOMER DECLINED TROUBLESHOOTING. NO ADDITIONAL INFORMATION WAS ABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20035277 ↗
  • MalfunctionAug 9, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19954448 ↗
  • MalfunctionAug 5, 2024Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19913071 ↗
  • MalfunctionAug 5, 2024Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE ALERT CLEARED, AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19911345 ↗
  • MalfunctionJul 19, 2024Filling Problem; Incorrect Measurement; Battery Problem; Filling Problem; Incorrect Measurement; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY, THE INSULIN GAUGE WAS INACCURATE, AND A MINIMUM FILL NOTIFICATION OCCURRED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19785023 ↗
  • MalfunctionJun 6, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19477225 ↗
  • MalfunctionMay 20, 2024Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19353227 ↗
  • MalfunctionMay 15, 2024Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP USB PORT WAS DAMAGED RESULTING IN DIFFICULTIES CHARGING THE PUMP. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19320551 ↗
  • MalfunctionApr 26, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19191878 ↗
  • MalfunctionApr 19, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. ADDITIONAL INFORMATION WAS NOT PROVIDED. THE CUSTOMER DECLINED FURTHER TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19143541 ↗
  • MalfunctionMar 18, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18921362 ↗
  • MalfunctionMar 15, 2024Failure to Sense; Battery Problem; Failure to Sense; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY AND THAT THE PUMP TOUCHSCREEN WAS UNRESPONSIVE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18913022 ↗
  • MalfunctionFeb 13, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVELS. CUSTOMER DECLINED TROUBLESHOOTING. NO ADDITIONAL INFORMATION WAS ABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18697991 ↗
  • MalfunctionFeb 9, 2024Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE CUSTOMER WAS UNABLE TO CHARGE THE PUMP USING THE TANDEM-PROVIDED WALL POWER ADAPTER. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER WAS ABLE TO SUCCESSFULLY CHARGE THE PUMP USING A SECONDARY WALL POWER ADAPTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18678361 ↗
  • MalfunctionFeb 9, 2024Incorrect Measurement; Battery Problem; Device Sensing Problem; Incorrect Measurement; Battery Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING, THE INSULIN GAUGE WAS INACCURATE AND THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18675405 ↗
  • MalfunctionJan 18, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO MANUAL INJECTIONS FOR INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18537664 ↗
  • MalfunctionDec 14, 2023Overheating of Device; Date/Time-Related Software Problem; Battery Problem; Overheating of Device; Date/Time-Related Software Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP CLOCK "LOSES" TIME. IT WAS ALSO REPORTED THAT THE PUMP WAS WARM TO THE TOUCH DESPITE BEING IN ROOM-TEMPERATURE CONDITIONS. REPORTEDLY, THE PUMP WAS CHARGING DURING THE EVENT. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18324218 ↗
  • MalfunctionNov 3, 2023Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS ¿NOT ILLUMINATING FULLY¿. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE. REPORTEDLY, THE CUSTOMER DECLINED TO PROVIDE ADDITIONAL INFORMATION, AND DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18064681 ↗
  • MalfunctionNov 2, 2023Incorrect Measurement; Date/Time-Related Software Problem; Incorrect Measurement; Date/Time-Related Software Problem

    IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE AND THAT THE PUMP TIME WAS INCORRECT, CAUSE WAS UNKNOWN. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT; THEREFORE THE ALLEGATION COULD NOT BE CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18056713 ↗

Other Pump, Infusion, Insulin devices

All Pump, Infusion, Insulin devices →