t:slim X2 Insulin Pump 1000369Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump, Reconditioned, 4YR Warranty · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is the Tandem t:slim X2 insulin pump. The 1,640 reports often describe the pump stopping, charging problems, inaccurate readings, and flow blockages. Some mention blood sugar running too high or too low. The records include no deaths, 188 injuries and 1,452 malfunctions. There is 1 open recall from May 2022 over memory faults and an inaccurate battery display.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,640
Deaths reported
0
Injuries reported
188
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1,640 adverse event reports that list this exact device (DI 00852162004798), received between Mar 6, 2017 and Aug 26, 2025 — classified as 188 injuries, 1,452 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (295 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped295
  • Charging Problem243
  • Incorrect Measurement230
  • Obstruction of Flow192
  • Battery Problem146
  • Adverse Event Without Identified Device or Use Problem128
  • Failure to Sense96
  • Physical Resistance/Sticking93

Patient problems

  • No Consequences Or Impact To Patient723
  • No Clinical Signs, Symptoms or Conditions596
  • No Known Impact Or Consequence To Patient141
  • Hyperglycemia95
  • Hypoglycemia77
  • Elevated Ketones/ Diabetic Ketoacidosis23
  • Insufficient Information5
  • Loss of consciousness3

Recent adverse event reports

The 25 most recent of 1,640 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2025Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22897534 ↗
  • MalfunctionJul 11, 2025Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. . THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22481318 ↗
  • InjuryMay 9, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT CARTRIDGE CHANGE ERRORS OCCURRED WITH MULTIPLE CARTRIDGES DURING THE LOAD SEQUENCE. A NEW CARTRIDGE WAS LOADED TO RESOLVE THE ISSUE. CUSTOMER WENT TO THE EMERGENCY ROOM (ER) WITH A BLOOD GLUCOSE (BG) LEVEL OF 400 MG/DL. CUSTOMER WAS TREATED WITH INTRAVENOUS FLUIDS OF SALINE AND INSULIN. CUSTOMER WAS DISCHARGED LATER THE SAME DAY WITH THE ISSUE RESOLVED AND NO PERMANENT DAMAGE.

    Patient problems: Hyperglycemia

    MAUDE report 21994184 ↗
  • MalfunctionMay 8, 2025Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. THE CUSTOMER REVERTED TO MANUAL INJECTIONS FOR INSULIN THERAPY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT AND DECLINED TO PROVIDE FURTHER INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21988937 ↗
  • MalfunctionApr 25, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT A CARTRIDGE CHANGE ERROR OCCURRED. REPORTEDLY, THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21914528 ↗
  • MalfunctionJan 28, 2025Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE CUSTOMER WAS UNABLE TO CHARGE THE PUMP USING THE TANDEM-PROVIDED USB CHARGING CABLE. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER WAS ABLE TO SUCCESSFULLY CHARGE THE PUMP USING A SECONDARY USB CABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21257452 ↗
  • MalfunctionNov 8, 2024Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT AN UNEXPECTED RESET OCCURRED. THERE WAS NO REPORTED IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20646467 ↗
  • MalfunctionOct 28, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT THE "PUMP DOES NOT TURN ON." THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, AND ADDITIONAL INFORMATION COULD NOT BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20550428 ↗
  • MalfunctionJul 26, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19842870 ↗
  • InjuryJun 25, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. IT WAS ALSO REPORTED THAT THE CUSTOMER EXPERIENCED AN ELEVATED BLOOD GLUCOSE (BG) LEVEL WAS "HIGH", ALTHOUGH A SPECIFIC VALUE WAS NOT GIVEN. CUSTOMER ADMINISTERED A MANUAL INJECTION TO ADDRESS THEIR BG LEVEL. CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: Hyperglycemia

    MAUDE report 19607185 ↗
  • MalfunctionJun 11, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19512216 ↗
  • MalfunctionApr 2, 2024Crack; Charging Problem; Crack; Charging Problem

    IT WAS REPORTED THAT THE PUMP'S USB PORT WAS DAMAGED RESULTING IN DIFFICULTIES CHARGING THE PUMP. THERE WAS NO ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19019224 ↗
  • MalfunctionJan 25, 2024Failure to Sense; Failure to Sense

    IT WAS REPORTED THAT THE PUMP TOUCHSCREEN WAS UNRESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE. REPORTEDLY, THE CUSTOMER DECLINED TO PROVIDE ADDITIONAL INFORMATION, AND DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18579531 ↗
  • MalfunctionJan 11, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER CONTACTED THEIR HEALTHCARE PROVIDER TO OBTAIN ALTERNATE INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18495459 ↗
  • InjuryDec 28, 2023Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. CUSTOMER CHANGED PUMP SUPPLIES TO ADDRESS THE ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS OVER 500 MG/DL. CUSTOMER ADDRESSED BG WITH A CORRECTION BOLUS VIA THE PUMP.

    Patient problems: Hyperglycemia

    MAUDE report 18410439 ↗
  • MalfunctionDec 26, 2023Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVEL. ADDITIONAL INFORMATION WAS NOT RECEIVED. THE CUSTOMER DECLINED FURTHER TROUBLESHOOTING FROM TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18393936 ↗
  • MalfunctionDec 8, 2023Incorrect Measurement; Pumping Stopped; Charging Problem; Incorrect Measurement; Pumping Stopped; Charging Problem

    IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP SHUT OFF UNEXPECTEDLY. IT WAS ALSO REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18290926 ↗
  • MalfunctionNov 22, 2023Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18193848 ↗
  • MalfunctionNov 21, 2023Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVEL. MULTIPLE FOLLOW UP ATTEMPTS FROM TANDEM TECHNICAL SUPPORT WAS MADE TO OBTAIN ADDITIONAL INFORMATION, HOWEVER, A RESPONSE FROM THE CUSTOMER WAS NOT RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18184418 ↗
  • MalfunctionNov 9, 2023Failure to Sense; Failure to Sense

    IT WAS REPORTED THAT THE PUMP TOUCHSCREEN WAS UNRESPONSIVE. CUSTOMER'S BLOOD WAS 185 MG/DL. REPORTEDLY THE CUSTOMER REVERTED TO MANUAL INJECTIONS FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18100473 ↗
  • MalfunctionOct 25, 2023Failure to Sense; Date/Time-Related Software Problem; Battery Problem; Failure to Sense; Date/Time-Related Software Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY, THAT THE TOUCH SCREEN WAS UNRESPONSIVE AND THAT THE PUMP TIME WAS INCORRECT, CAUSE WAS UNKNOWN. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT; THEREFORE THE ALLEGATION COULD NOT BE CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18002287 ↗
  • MalfunctionOct 24, 2023Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE AND THE BATTERY GAUGE WAS FLUCTUATING. CUSTOMER LOADED A NEW CARTRIDGE TO RESOLVE THE ISSUE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17995552 ↗
  • MalfunctionOct 10, 2023Failure to Sense; Obstruction of Flow; Failure to Sense; Obstruction of Flow

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. ADDITIONALLY, IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17904869 ↗
  • MalfunctionOct 6, 2023Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT THE PUMP SHUT OFF UNEXPECTEDLY. THE CUSTOMER DECLINED A FOLLOW UP, SO NO ADDITIONAL INFORMATION WAS OBTAINED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17887526 ↗
  • MalfunctionSep 29, 2023Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS MISSING PIXELS. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17844124 ↗

FDA clearance and classification

The FDA regulates product code LZG (pump, infusion, insulin) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the t:slim X2 Insulin Pump 1000369

Has the t:slim X2 Insulin Pump 1000369 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1811-2022 (Class II), began on May 24, 2022: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump 1000369?

The FDA's MAUDE database lists 1,640 adverse event reports for this device, received between Mar 6, 2017 and Aug 26, 2025. In those reports, the most common specific device problems are “Pumping Stopped” (295), “Charging Problem” (243) and “Incorrect Measurement” (230); the most common patient problems are “Hyperglycemia” (95), “Hypoglycemia” (77) and “Elevated Ketones/ Diabetic Ketoacidosis” (23). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump 1000369?

None of the reports are classified as deaths, 188 as injuries and 1,452 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump 1000369 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump 1000369, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump, model 1000369). Its record shows it as in commercial distribution.