Skip to content

Oracle Health Blood Bank PathNet Blood Bank Transfusion 2018Recalls & reported problems

Cerner Corporation · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00853023006173), received between Feb 5, 2021 and Feb 5, 2021 — classified as 1 death.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Computer Software Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 5, 2021Computer Software Problem; Computer Software Problem

    THIS REPORT DOCUMENTS INFORMATION RELATED TO AN ISSUE IDENTIFIED WITH FUNCTIONALITY INCLUDED IN CERNER'S HNA® MILLENNIUM PATHNET® BLOOD BANK TRANSFUSION (CLASS II, PRODUCT CODE = MMH). THIS ISSUE INVOLVES CERNER'S MILLENNIUM USERS THAT UTILIZE THE COMBINE FUNCTION (HNA COMBINE.EXE) TO COMPLETE AN ENCOUNTER MOVE FOR ONE PATIENT WITH MULTIPLE ABO/RH RESULTS TO A COMBINED PATIENT WITH THE SAME ABO/RH RESULTS. WHEN THIS ISSUE OCCURS, THE ABO/RH IS REMOVED FROM THE COMBINED PATIENT. IF A PATIENT DOES NOT HAVE AN ABO/RH RECORDED IN THE SYSTEM, THE SYSTEM DOES NOT ALLOW USERS TO COMPLETE A CROSSMATCH OR DISPENSE A BLOOD PRODUCT FOR THAT PATIENT. THIS COULD LEAD TO A DELAY IN DISPENSING BLOOD PRODUCTS FOR A PATIENT IF BLOOD BANK PERSONNEL ARE NOT AWARE THAT THE ABO/RH HAS BEEN REMOVED. A NEW ABO/RH RESULT WOULD NEED TO BE ENTERED ON THE COMBINED PATIENT BEFORE THEY CAN DISPENSE BLOOD PRODUCTS VIA THE SYSTEM TO THE PATIENT. CERNER HAS RECEIVED COMMUNICATION ON AN ADVERSE PATIENT EVENT AS A RESULT OF THIS ISSUE. A PATIENT WAS DELAYED A BLOOD PRODUCT TRANSFUSION BECAUSE OF THIS ISSUE. THE PATIENT DID RECEIVE BLOOD PRODUCTS AND WAS PRONOUNCED DECEASED LATER SAME DAY BUT IS UNKNOWN IF THE ISSUE

    Patient problems: Insufficient Information

    MAUDE report 11288175 ↗

Other Blood Establishment Computer Software And Accessories devices

All Blood Establishment Computer Software And Accessories devices →