ADIMS 1.0Recalls & reported problems
Csl Plasma Inc. · ADIMS is a stand-alone blood establishment computer software application (BECS) that provides for the management of plasma and whole blood donors for an RBC immunization program as well as the record and inventory management of immunogen red blood cells. The application is used to determine donor eligibility and release of the blood or blood component for transfusion as well as tracking the preparation of immunogen components (vials) for storage and distribution. Additionally, ADIMS allows for the tracking the recipient of transfused immunogen RBCs · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00860007397615), received between Jan 31, 2022 and Jan 31, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Application Program Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Application Program Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 31, 2022Application Program Problem
THE REQUIREMENT TO HAVE ONE-YEAR POST COLLECTION NEGATIVE TEST RESULTS WAS INCLUDED IN THE USER REQUIREMENT SPECIFICATIONS; BUT WAS NOT PROGRAMMED AS INTENDED IN THE APPLICATION. THE REQUIREMENT TO MARK ANY ALIQUOT WITH POSITIVE TEST RESULTS FOR DESTRUCTION IS PROGRAMMED AS INTENDED. TO THIS DATE, THERE HAVE BEEN NO POSITIVE ALIQUOTS THAT REQUIRE DESTRUCTION. A TOTAL OF 54 ALIQUOTS WERE IMPACTED. ALL ALIQUOTS HAD ACCEPTABLE TEST RESULTS AT TIME OF COLLECTION. ONE (1) YEAR POST DONATION TEST RESULTS WERE NOT AVAILABLE AT TIME OF DEGLYCEROLIZATION. EIGHT (8) ALIQUOTS WERE ADMINISTERED WITHOUT TESTING AT THE TIME OF ADMINISTRATION; BUT RECEIVED ACCEPTABLE TEST RESULTS AFTERWARDS; 46 ALIQUOTS FROM 7 DIFFERENT DONORS WERE ADMINISTERED WITH NO 1 YEAR POST DONATION TESTING RECEIVED TO DATE, BUT HAD ACCEPTABLE RESULTS AT THE TIME OF COLLECTION. FDA SAFETY REPORT ID# (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13416908 ↗
Other Blood Establishment Computer Software And Accessories devices
- Pyxis TRANSFUSIONCarefusion 303, Inc. · SOFTWARE, BLOOD BANK, STAND ALONE PRODUCTS5 reports
- Regulated Software Application (RSA) 1.0.19.0Oneblood, Inc. · Software, Blood Bank, Stand Alone Products4 reports
- ImmuLINK 0064319Immucor, Inc. · Software, Blood Bank, Stand Alone Products3 reports
- Expanse Blood BankMedical Information Technology, Inc. · Blood Establishment Computer Software And Accessories1 report
- Instrument Manager 8150011Data Innovations LLC · Blood Establishment Computer Software And Accessories1 report
- Oracle Health Blood Bank PathNet Blood Bank Transfusion 2018Cerner Corporation · Blood Establishment Computer Software And Accessories1 death1 report
- Instrument Manager 817Data Innovations LLC · Blood Establishment Computer Software And Accessories1 report
- Sunquest Blood Bank 10.0CliniSys, Inc · Blood Establishment Computer Software And Accessories1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.