t:slim X2 Insulin Pump with Basal-IQ 1000892Recalls & reported problems
Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Basal-IQ, Replacement, 6.0 · Class 3
All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Basal-IQ (14)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Tandem t:slim X2 insulin pump with Basal-IQ delivers insulin automatically. Reports most often describe the pump stopping, incorrect on-screen messages, blocked flow and charging or battery problems. The device appears in 5,216 reports, including 1 death and 494 injuries. It has 1 open recall, started in May 2022, covering memory, battery display and touchscreen faults, which a software update can address. Reports dropped to 6 in the last 12 months from 28 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 5,216
- Deaths reported
- 1
- Injuries reported
- 494
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5,216 adverse event reports that list this exact device (DI 00853052007288), received between Nov 9, 2018 and Jun 8, 2026 — classified as 1 death, 494 injuries, 4,721 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (858 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 24, 2022Class IIOpen, Classified
Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.
Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Pumping Stopped858
- Device Displays Incorrect Message643
- Obstruction of Flow537
- Incorrect Measurement526
- Charging Problem487
- Battery Problem396
- Failure to Sense372
- Adverse Event Without Identified Device or Use Problem355
Patient problems
- No Consequences Or Impact To Patient1,959
- No Clinical Signs, Symptoms or Conditions1,878
- No Known Impact Or Consequence To Patient903
- Hypoglycemia261
- Hyperglycemia215
- Elevated Ketones/ Diabetic Ketoacidosis40
- Loss of consciousness6
- Insufficient Information5
Recent adverse event reports
The 25 most recent of 5,216 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT PUMP SCREEN WAS DARK AND WOULD NOT TURN ON DESPITE MULTIPLE ATTEMPTS TO WAKE IT. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25429495 ↗MalfunctionMay 11, 2026Pumping Stopped
IT WAS REPORTED THAT PUMP MALFUNCTION OCCURRED WITHOUT ANY NOTABLE EVENTS BEFOREHAND. CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS 438 MG/DL. CUSTOMER HAD NOT TAKEN ANY ACTION TO LOWER BLOOD GLUCOSE BEFORE CONTACTING SUPPORT, AND TECHNICAL SUPPORT ASSISTED CUSTOMER WITH RESETTING PUMP USING PROVIDED INSTRUCTIONS; MALFUNCTION DID NOT RECUR AFTER RESET, CUSTOMER WAS ADVISED TO CONTINUE USING PUMP AND TO CALL BACK IF MALFUNCTION OCCURRED AGAIN, AND CUSTOMER ACKNOWLEDGED AND UNDERSTOOD THESE INSTRUCTIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25136034 ↗MalfunctionDec 15, 2025Pumping Stopped; Pumping Stopped
A MALFUNCTION ON THE PUMP WAS REPORTED BY THE CALLER, WITH NO NOTABLE EVENTS OCCURRING PRIOR TO THE INCIDENT. THERE WAS NO ADVERSE IMPACT ON THE CUSTOMER'S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT RESOLVED THE ISSUE BY REPLACING THE PUMP, AND THE CUSTOMER SWITCHED TO AN IN-WARRANTY PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23804595 ↗MalfunctionNov 26, 2025Break; Charging Problem; Break; Charging Problem
IT WAS REPORTED THAT THE PUMP'S USB PORT WAS DAMAGED RESULTING IN DIFFICULTIES CHARGING THE PUMP. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE; THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23658580 ↗MalfunctionNov 24, 2025Failure to Charge; Failure to Charge
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE ALERT CLEARED, AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23640610 ↗MalfunctionNov 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
CALLER REPORTED A CGM ERROR IDENTIFIED AS ERROR 42. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT PROVIDED GUIDANCE FOR A COMPLETE PUMP RESET, ASSISTING THE CALLER THROUGH THE PROCESS, AND THE SENSOR SESSION WAS SUCCESSFULLY RESTARTED WITHOUT RECURRING ERROR MESSAGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23453070 ↗MalfunctionOct 15, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23299298 ↗MalfunctionSep 16, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. THERE WAS NO REPORTED ADVERSE IMPACT. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23066865 ↗MalfunctionSep 15, 2025Date/Time-Related Software Problem; Date/Time-Related Software Problem
AN ISSUE WAS REPORTED INVOLVING AN INCORRECT TIME SETTING ON A PUMP, WHICH THE CALLER DID NOT UNDERSTAND THE CAUSE OF. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL AS PREVIOUS SETTINGS DID NOT RESULT IN ABNORMAL LEVELS. MULTIPLE ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO FOLLOW-UP WITH THE CUSTOMER ON THE REPORTED ISSUE, BUT NO RESPONSE WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23051746 ↗MalfunctionJul 21, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
CUSTOMER REPORTED THAT PUMP WAS GIVING FALSE NO DELIVERY ALARMS AND NO PRIME ALARMS, WHICH RESULTED IN UNNECESSARY PRIMING AND INTERRUPTED INSULIN DELIVERY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER OPTED OUT OF TROUBLESHOOTING AND REQUESTED DOCUMENTATION FOR INSURANCE PURPOSES; NO FURTHER ACTION WAS TAKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22555542 ↗MalfunctionJul 17, 2025Failure to Charge; Failure to Charge
CUSTOMER REPORTED THAT THE CURRENT PUMP DID NOT FULLY CHARGE THE BATTERY PROPERLY AND ISSUED A BATTERY ERROR WHILE CHARGING, PREVENTING THE PUMP FROM REACHING 100% CHARGE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FURTHER ASSISTANCE FROM CTS, WAS CURRENTLY OUT OF WARRANTY, AND IN THE PROCESS OF ACQUIRING A NEW DEVICE. NO ADDITIONAL OUTCOME INFORMATION WAS RECEIVED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22534409 ↗MalfunctionJul 16, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE. THE CUSTOMER DECLINED FOLLOW UP WITH TANDEM TECHNICAL SUPPORT AND FURTHER INFORMATION WAS UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22521616 ↗MalfunctionJul 16, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
A MALFUNCTION WAS REPORTED BY THE CUSTOMER, WHERE THE PUMP SCREEN WAS FROZEN AND UNRESPONSIVE TO INPUT. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER PERFORMED A PUMP RESET WITH GUIDANCE FROM TECHNICAL SUPPORT, AND THE ISSUE WAS RESOLVED, ALLOWING THE PUMP SCREEN TO FUNCTION NORMALLY AGAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22521382 ↗MalfunctionJul 15, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING THE LOAD SEQUENCE. CUSTOMER LOADED SEVERAL NEW CARTRIDGES TO RESOLVE THE ISSUE HOWEVER THE CARTRIDGE ALARM RECURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22507806 ↗MalfunctionMay 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A CARTRIDGE DID NOT FIT ONTO THE PUMP. REPORTEDLY, THERE WAS AN O-RING FROM A PREVIOUS CARTRIDGE ON THE PNEUMATIC TAP. THE CUSTOMER REMOVED THE O-RING AND LOADED A NEW CARTRIDGE TO ADDRESS THE EVENT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21979913 ↗MalfunctionMar 28, 2025Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT WITH TANDEM TECHNICAL SUPPORT; THEREFORE, NO ADDITIONAL INFORMATION COULD BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21721908 ↗MalfunctionMar 28, 2025Charging Problem; Charging Problem
IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21719477 ↗MalfunctionMar 26, 2025Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21699297 ↗MalfunctionMar 19, 2025Filling Problem; Incorrect Measurement; Filling Problem; Incorrect Measurement
IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER THE USER FILLED THE CARTRIDGE WITH 225 UNITS OF INSULIN DURING THE LOAD SEQUENCE. REPORTEDLY, CUSTOMER DECLINED TO LOAD A NEW CARTRIDGE ON THE PUMP. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21646673 ↗MalfunctionFeb 24, 2025Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21446755 ↗MalfunctionJan 6, 2025Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER PRESSED THE WAKE BUTTON MULTIPLE TIMES AND APPLIED MORE FORCE TO THE BUTTON TO RESOLVE THE ISSUE. THE CUSTOMER REVERTED TO ALTERNATE METHOD OF INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21084116 ↗MalfunctionDec 30, 2024Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE WAKE BUTTON WAS ¿OFTEN UNRESPONSIVE¿. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21035858 ↗MalfunctionDec 13, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. THERE WAS NO REPORTED ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20929399 ↗MalfunctionDec 12, 2024Failure to Sense; Failure to Sense
IT WAS REPORTED THAT THE P UMP TOUCHSCREEN WAS UNRESPONSIVE. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE PUMP WAS RESET, AND THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20920993 ↗MalfunctionDec 10, 2024Difficult to Insert; Fitting Problem; Difficult to Insert; Fitting Problem
IT WAS REPORTED THAT THE RACK PUSHROD ON THE PUMP WAS DAMAGED, WHICH LED TO DIFFICULTIES LOADING CARTRIDGES. REPORTEDLY, THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. CUSTOMER¿S BLOOD GLUCOSE VALUE WAS 266 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20898184 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P180008 on Jun 21, 2018. Submitted by Tandem Diabetes Care, Inc. as “t:slim X2 Insulin Pump with Basal-IQ Technology”. The FDA review took 115 days. The FDA has since approved 6 supplements (changes to the device or its labeling), most recently on Dec 19, 2019. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
FDA definition of this device type: An automated insulin dosing system that temporarily suspends or reduces insulin infusion from an insulin pump based upon specified thresholds of measured glucose levels.
Common questions about the t:slim X2 Insulin Pump with Basal-IQ 1000892
Has the t:slim X2 Insulin Pump with Basal-IQ 1000892 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1811-2022 (Class II), began on May 24, 2022: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the t:slim X2 Insulin Pump with Basal-IQ 1000892?
The FDA's MAUDE database lists 5,216 adverse event reports for this device, received between Nov 9, 2018 and Jun 8, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (858), “Device Displays Incorrect Message” (643) and “Obstruction of Flow” (537); the most common patient problems are “Hypoglycemia” (261), “Hyperglycemia” (215) and “Elevated Ketones/ Diabetic Ketoacidosis” (40). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Basal-IQ 1000892?
1 report is classified as a death, 494 as injuries and 4,721 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the t:slim X2 Insulin Pump with Basal-IQ 1000892 FDA approved?
It was approved through premarket approval (PMA) P180008 on Jun 21, 2018.
Is the t:slim X2 Insulin Pump with Basal-IQ 1000892 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the t:slim X2 Insulin Pump with Basal-IQ 1000892, and is it still sold?
The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Basal-IQ, model 1000892). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Automated Insulin Dosing , Threshold Suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,651 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,105 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page