Rotation Medical Tendon Staples 2504Recalls & reported problems
Smith & Nephew, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Smith & Nephew tendon anchor staple is implanted to secure tendon repair during shoulder surgery. The 57 reports mostly describe staples breaking during surgery, sometimes leaving fragments inside the patient that had to be removed, along with anchors arriving unsterile or with packaging and connection problems. There are 28 injuries and no deaths. One open recall, started in August 2024, says the outer foil pouch may not seal properly and can compromise sterility.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 57
- Deaths reported
- 0
- Injuries reported
- 28
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 57 adverse event reports that list this exact device (DI 00854501006067), received between Sep 9, 2019 and Jun 13, 2026 — classified as 28 injuries, 29 malfunctions.
The most frequently reported specific device problem is “Break” (17 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 12, 2024Class IIOpen, Classified
Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility
Z-3096-2024 · Root cause: Packaging · Recall details · FDA record ↗
Most reported problems
Device problems
- Break17
- Adverse Event Without Identified Device or Use Problem14
- Delivered as Unsterile Product8
- Connection Problem6
- Packaging Problem4
- Material Separation3
- Crack2
- Device Dislodged or Dislocated2
Patient problems
- No Clinical Signs, Symptoms or Conditions28
- Inflammation10
- No Known Impact Or Consequence To Patient8
- Pain8
- Unspecified Infection6
- Blood Loss2
- Fluid Discharge2
- Foreign Body In Patient2
Recent adverse event reports
The 25 most recent of 57 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 13, 2026Break
IT WAS REPORTED THAT, DURING A PROCEDURE, EVERY TENDON ANCHOR BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25495661 ↗MalfunctionDec 8, 2025Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE TENDON ANCHORS 8 BROKE WHEN THEY WERE DEPLOYED. THE BROKEN PIECES WERE REMOVED FROM THE PATIENT WITH AN ARTERY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A SIMILAR S+N BACK-UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23744083 ↗MalfunctionOct 31, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF PROCEDURE, IN AN ATTEMPT TO AVOID WASTING THE REGENETEN IMPLANT, THE SURGEON TRIED TO STAPLE IT USING A TENDON ANCHOR; HOWEVER, THE IMPLANT WAS FIXATED OFF THE INTENDED REPAIR SITE AND THE SURGEON DECIDED TO REMOVE IT COMPLETELY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACKUP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23442200 ↗MalfunctionMay 14, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE TENDON ANCHORS HAD WATER DAMAGE TO THE BOX AND TO THE ACTUAL PRODUCT INSIDE OF THE PACKAGE. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22019731 ↗InjuryOct 29, 2024Break; Break
IT WAS REPORTED THAT DURING AN SHOULDER CUFF AUGMENTATION, BOX OF EIGHT (8) TENDON ANCHORS BROKE. THREE 3 BROKE DURING ASSEMBLY WHEN OPENING THE BOX, AND THE OTHER FIVE (5) IMPLANTS BROKE DURING APPLICATION INSIDE THE PATIENT. THE BROKEN IMPLANTS REMAINED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 20561580 ↗MalfunctionAug 20, 2024Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TENDON ANCHORS (QUALITY HOLD ITEM) WERE USED IN THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20005827 ↗MalfunctionJul 16, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT, DURING SET UP, A TENDON ANCHOR PUCK WAS NOT VACUUM SEALED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19747434 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT, DURING A ROTATOR CUFF REPAIR, SEVEN (7) TENDON ANCHORS BROKE WHEN TRYING TO RETRIEVE THEM FROM THE PUCK IT COMES IN AND THE REGENETEN PROCEDURE HAD TO BE ABANDONED. THE DEVICES DID NOT BREAK INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING A BACK UP DEVICE INSTEAD. THERE WAS A SIGNIFICANT SURGICAL DELAY OF MORE THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19713529 ↗MalfunctionJul 1, 2024Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem
IT WAS REPORTED THAT, DURING SET UP, SEVEN TENDOR ANCHOR PUCKS WERE NOT VACUUM SEALED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19650346 ↗MalfunctionJul 1, 2024Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem
IT WAS REPORTED THAT, DURING SET UP, TWO TENDOR ANCHOR PUCKS WERE NOT VACUUM SEALED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19648597 ↗MalfunctionJun 27, 2024Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem
IT WAS REPORTED THAT (3) THREE TENDON ANCHORS' PACKAGES WERE NOT SEALED AND WERE PUNCTURED. THE EVENT OCCURRED AFTER REMOVING PACKAGING FROM OUTER BOX, FOR TREATMENT OF A PARTIAL ROTATOR CUFF TEAR. THE PROCEDURE WAS COMPLETED USING A BACKUP DEVICE WITH A DELAY OF LESS THAN 2 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19624859 ↗MalfunctionJun 27, 2024Delivered as Unsterile Product; Unsealed Device Packaging; Delivered as Unsterile Product; Unsealed Device Packaging
IT WAS REPORTED THAT, TWO (2) TENDON ANCHORS WERE NOT VACUUMED SEALED COMPLETELY. THE AIR COULD NOT BE SQUEEZED OUT, IT WAS VACUUMED SEALED JUST NOT COMPLETELY. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19624805 ↗MalfunctionMay 21, 2024Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT, DURING A SHOULDER RCR PROCEDURE, THREE (3) DIFFERENT LOTS OF TENDON ANCHORS COULD NOT BE LOADED IN THE INSERTION DEVICE. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE INSTEAD. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19360804 ↗MalfunctionMar 15, 2024Break; Break
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, FIVE (5) OUT OF EIGHT (8) TENDON ANCHORS, BROKE AFTER BEING IMPLANTED USING THE TENDON STAPLE. ALL THE ANCHORS WERE REMOVED FROM THE PATIENT AS WELL AS THE REGENETEN PATCH. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Perforation
MAUDE report 18918479 ↗InjuryNov 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A RIGHT ACHILLES TENDON ENDOSCOPIC DEBRIDEMENT, EXCISION OF HAGLUND¿S LESION AND REGENETEN OPERATION WHERE TENDON ANCHORS WERE USED, THE PATIENT HAD SERIOUS DISCHARGE OVER THE WOUND, EVEN AFTER THE WOUNDS WERE HEALED AND STO WAS DONE. A CULTURE WAS TAKING FROM THE WOUND AND IT WAS NEGATIVE FOR ANY HARMFUL PATHOGEN. WHEN THE DISCHARGE WAS NOTICED, THE PATIENT WAS PROVIDED WITH ORAL AUGMENTIN, BUT THE PATIENT COULD NOT TOLERATE IT AND THIS LET TO VOMITING. THEREFORE, THE PATIENT WAS ADMINISTER 3 DAYS OF IV AND 5 DAYS OF ORAL TREATMENT. AFTER 3 WEEKS THE WOUND HAD DRIED UP, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Fluid Discharge; Unspecified Infection
MAUDE report 18092780 ↗InjuryNov 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A REGENETEN IMPLANTATION PROCEDURE USING THE TENDON ANCHORS IN 2022, THE PATIENT HAD RICE BODIES AND EXPERIENCED PAIN, A WASHOUT WAS PERFORMED ON (B)(6) 2023. THE CURRENT PATIENT STATUS IMPROVED SINCE WASHOUT.
Patient problems: Foreign Body Reaction; Pain; Synovitis
MAUDE report 18048674 ↗InjuryMay 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, APPROXIMATELY, A MONTH AND HALF AFTER AN ARTHROSCOPY WHERE TENDON ANCHORS WERE USED, A RAISED SACK APPEARED ON THE INTRA LATERAL PORTAL OF THE PATIENT. SYNOVIAL FLUID WAS DRAINING OUT OF THE SACK WITH SOME CHUNKS. WHEN THE SACK WAS DECOMPRESSED AND THEN ABDUCTED, PARTICLES COULD BE SEEN. IT IS UNKNOWN HOW THE INJURY WAS TREATED OR THE PATIENT'S OUTCOME.
Patient problems: Fluid Discharge; Swelling/ Edema
MAUDE report 16950282 ↗InjuryApr 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, FIVE WEEKS AFTER A DOUBLE ROW REPAIR ON A MASSIVE ROTATOR CUFF TEAR SURGERY, THE PATIENT EXPERIENCED INFLAMMATION AND DRAINING PORTALS. THE SURGEON ADMINISTERED ANTIBIOTICS FOR TWO DAYS BUT IT GOT WAY WORSE. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY IN WHICH 40CC OF PUSS WAS DRAINED AND THE SUBACROMIAL SPACE WAS CLEANED OUT. DURING THIS SURGERY THE SURGEON SAW THE TENDON ANCHORS FLOATING INSIDE THE SUBACROMIAL SPACE AS WELL. THE CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Inflammation; Medical device site infection
MAUDE report 16773111 ↗InjuryFeb 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT 4 MONTHS AFTER A SHOULDER PROCEDURE IN WHICH A REGENETEN DEVICE WAS IMPLANTED, THE PATIENT RETURN TO THE PHYSICIAN WITH STIFFNESS, PAIN AND POOR MOBILITY. THE PHYSICIAN DECIDED TO MRI THE PATIENT AND FROM THE IMAGING IT TURNED OUT THAT THE PATIENT HAD CALCIUM DEPOSITS IN THEIR SHOULDER, PATIENT WAS TREATED WITH AN INJECTION AFTERWARDS. PATIENT OUTCOME IS UNKNOWN.
Patient problems: Calcium Deposits/Calcification; Muscular Rigidity; Pain
MAUDE report 16440968 ↗InjuryFeb 23, 2023Break; Break
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR, FIVE (5) TENDON ANCHORS FRACTURED; 4 ANCHORS WHILE LOADING THEM INTO THE INSERTER AND 1 ANCHOR INSIDE OF THE PATIENT AND IT WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS FINISHED WITH 5 MINUTES OF DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16433464 ↗InjuryFeb 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A PROCEDURE, THE TENDON ANCHOR GENERATED AN ADVERSE REACTION, AND CAUSED AN INFLAMMATORY RESPONSE WHEN USING IT. THE DEVICE WAS EXPLANTED THE SAME DAY. IT IS UNKNOWN IF THERE WAS A DELAY OR IF A BACKUP DEVICE WAS AVAILABLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Inflammation
MAUDE report 16397257 ↗InjuryFeb 1, 2023Break; Break
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, WHEN INSERTING THE PUCKS INTO THE TENDON, THE TENDON ANCHORS THEMSELVES WOULD BREAK. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED, AND WHETHER IT CAUSED OR NOT A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16279454 ↗InjuryDec 21, 2022Break; Break
IT WAS REPORTED THAT DURING AN ACL, A TENDON ANCHOR DID NOT CLIP AND GOT BROKEN WHEN IMPLANTED TO THE PATCH. THE BROKEN ANCHOR WAS REMOVED WITH A GRASPER. SURGERY RESUMED AFTER A NON-SIGNIFICANT DELAY, WITH A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16024705 ↗InjuryOct 14, 2022Break; Break
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY, SIX (6) TENDON ANCHORS BROKE WHILE UPLOADING AND IMPLEMENTATION. SURGERY WAS RESUMED WITH A BACK-UP DEVICE, BUT IT IS UNKNOWN WHETHER IT CAUSED A SURGICAL DELAY OR NOT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15611502 ↗InjuryOct 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED AN INFLAMMATION 6 WEEKS AFTER A SURGERY WHERE 2 REGENETEN DEVICES WERE USED. THE PATIENT WAS BROUGHT BACK TO SURGERY AT 8 WEEKS, NO INFECTION WAS FOUND BUT THE DOCTOR FELT THE REGENETEN WAS ¿MUSHY¿ AND DECIDED TO REMOVE IT. HE WAS NOT SURE IF THERE WAS AN ISSUE WITH REGENETEN, BUT HE IS ALSO NOT SURE IT WAS INCORPORATING LIKE IT SHOULD. FIXATION COULD HAVE ALSO BEEN AN ISSUE AS THE DISTAL PART OF THE IMPLANT MAY NOT HAVE BEEN FIXATED WELL. NO ADDITIONAL COMPLICATIONS WERE REPORTED.
Patient problems: Inflammation
MAUDE report 15577573 ↗
Other Staple, Implantable devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page