Precision Medical Blender PM5300Recalls & reported problems

Precision Medical, Inc. · PM5300 SERIES,BLENDER,AIR/O2,LOW FLOW · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00855887006016), received between Jul 12, 2018 and Mar 2, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fire” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fire2
  • Device Damaged by Another Device1
  • No Audible Prompt/Feedback1

Patient problems

  • Hypoxia1
  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2023No Audible Prompt/Feedback; Device Damaged by Another Device; No Audible Prompt/Feedback; Device Damaged by Another Device

    NICU RN REPORTED THAT THE BLENDER GAS SUPPLY IMBALANCE ALARM WAS NOT ALARMING PROPERLY AND ADVISED THERE WAS A PATIENT INCIDENT. THE BABY WAS TRANSPORTED TO NICU WITHOUT PROPER OXYGEN RESULTING IN THE BABY BEING HYPOXIC UNTIL ARRIVAL AND STABILIZATION IN NICU. BLENDER WAS TESTED AND THE REPORTED PROBLEM WAS REPRODUCED - REMOVED DEVICE FROM SERVICE. AFTER FURTHER TESTING, AN ADDITIONAL THREE UNITS HAVE BEEN FOUND WITH THE SAME PROBLEM.

    Patient problems: Hypoxia

    MAUDE report 16469217 ↗
  • InjuryApr 8, 2021Fire; Fire

    FIRE INVOLVING PRECISION MEDICAL BLENDER CAUSED BY A HEATING ELEMENT HEAD OF THE WARMER AS DESCRIBED IN THE NARRATIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11633153 ↗
  • MalfunctionJul 12, 2018Fire; Fire

    FIRE INCIDENT WHICH OCCURRED TO AN OXYGEN BLENDER (5300) IN A NEONATAL INTENSIVE CARE UNIT. THIS INCIDENT WAS REPORTED BY THE HOSPITAL PERSONNEL TO THE (B)(6) AND A COMPLAINT SENT TO US ON (B)(6) 2018. DATE OF ACTUAL OCCURRENCE IS UNKNOWN

    Patient problems: No Patient Involvement

    MAUDE report 7682122 ↗

Other Mixer, Breathing Gases, Anesthesia Inhalation devices

All Mixer, Breathing Gases, Anesthesia Inhalation devices →