Sechrist Industries, Inc. 3500CP-GRecalls & reported problems
Sechrist Industries, Inc. · MIXER, GENERAL · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00899660002345), received between Feb 17, 2021 and Jan 8, 2026 — classified as 1 death, 10 malfunctions.
The most frequently reported specific device problem is “No Audible Alarm” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Audible Alarm5
- Device Contamination with Body Fluid2
- Gas Output Problem1
- Infusion or Flow Problem1
- No Flow1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Hypoxia1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 8, 2026No Audible Alarm
CUSTOMER REPORTED THAT THE AIR AND OXYGEN ALARM NOT ACTIVE WHEN LOW PRESSURE. NO PATIENT INVOLVEMENT OR INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24010742 ↗MalfunctionDec 9, 2025No Flow; No Flow
CUSTOMER REPORTED STAFF WAS ATTEMPTING TO OXYGENATE THE PATIENT WHILE HE/SHE WAS ON BYPASS AND WERE UNABLE TO. STAFF SWAPPED THE DEVICE FOR ANOTHER ONE AND IT WORKED APPROPRIATELY. NO PATIENT HARM OCCURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23759634 ↗MalfunctionNov 24, 2025No Audible Alarm; No Audible Alarm
CUSTOMER REPORTED THAT THE MIXER WON'T ALARM AND HAS A HISSING SOUND COMING FROM IT. NO PATIENT INVOLVEMENT OR INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23638158 ↗MalfunctionOct 20, 2025Device Contamination with Body Fluid; Device Contamination with Body Fluid
IT WAS REPORTED THAT BLOOD HAS ENTERED THE GAS MIXER. NO HARM TO ANY PERSON WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23342249 ↗MalfunctionJul 30, 2025Output Problem; Output Problem
CUSTOMER REPORTING, "THE OXYGEN BLENDER WAS NOT PROVIDING 100% OXYGEN, AS IT WAS SET TO, BUT A MUCH LOWER FIO2." NO PATIENT INVOLVEMENT OR INJURY REPORTED. NO OTHER INFORMATION PROVIDED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22670049 ↗MalfunctionJul 29, 2025No Audible Alarm; No Audible Alarm
CUSTOMER REPORTED THE UNIT HAS NO ALARM WHEN ONE OF THE SUPPLY LINES IS MISSING. THIS ISSUE WAS DISCOVERED DURING SET-UP AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22654861 ↗MalfunctionApr 18, 2025No Audible Alarm; No Audible Alarm
CUSTOMER REPORTED THE UNIT DOES NOT ALARM WHEN IT IS DISCONNECTED FROM AIR OR O2. NO PATIENT INCIDENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21868146 ↗MalfunctionJan 21, 2025Gas Output Problem; Gas Output Problem
INCIDENT REPORTED TO SECHRIST BY EUROPEAN DISTRIBUTOR ON 12/20/2024. CUSTOMER REPORTING THAT THE MIXER WAS NOT DELIVERING GAS. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21203726 ↗MalfunctionJan 15, 2025Device Contamination with Body Fluid; Device Contamination with Body Fluid
CUSTOMER REPORTED THE DEVICE HAS BLOOD INSIDE, DURING AN ACCIDENTAL SITUATION THAT MAKES THE DEVICE UNFEASIBLE TO USE. NO PATIENT INJURY WAS REPORTED. THE DEVICE HAS BEEN DECOMMISSIONED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21160393 ↗MalfunctionOct 30, 2024No Audible Alarm; No Audible Alarm
CUSTOMER IS REPORTING THAT THERE UNIT IS NOT ALARMING. NO PATIENT INVOLVEMENT OR INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20571712 ↗DeathFeb 17, 2021Infusion or Flow Problem; Infusion or Flow Problem
THIS SUPPLEMENTAL REPORT IS REQUIRED TO REPORT DEVICE EVALUATION RESULTS. CORRECTION: D8
Patient problems: Hypoxia
MAUDE report 11340465 ↗
Other Mixer, Breathing Gases, Anesthesia Inhalation devices
- Blender 03800ACarefusion Corporation · Mixer, Breathing Gases, Anesthesia Inhalation1 death23 reports
- Bird Blender 03920ACarefusion Corporation · Mixer, Breathing Gases, Anesthesia Inhalation17 reports
- Hyperinflation AMBU 900 000 614Westmed, Inc. · Mixer, Breathing Gases, Anesthesia Inhalation6 reports
- Sechrist Industries, Inc. 3500CP-G24Sechrist Industries, Inc. · Mixer, Breathing Gases, Anesthesia Inhalation4 reports
- Blender 10040ACarefusion Corporation · Mixer, Breathing Gases, Anesthesia Inhalation4 reports
- Precision Medical Blender PM5300Precision Medical, Inc. · Mixer, Breathing Gases, Anesthesia Inhalation3 reports
- Sechrist Industries, Inc. 3500CP-G4Sechrist Industries, Inc. · Mixer, Breathing Gases, Anesthesia Inhalation3 reports
- Blender 10041ACarefusion Corporation · Mixer, Breathing Gases, Anesthesia Inhalation2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.