Skip to content

Sechrist Industries, Inc. 3500CP-G4Recalls & reported problems

Sechrist Industries, Inc. · MIXER,GENERAL,GERMAN · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00899660002796), received between Oct 30, 2024 and Feb 18, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Body Fluid” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Body Fluid2
  • No Audible Alarm1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2025Device Contamination with Body Fluid; Device Contamination with Body Fluid

    EU DISTRIBUTOR INFORMING SECHRIST OF CUSTOMER REPORTING " BLOOD INSIDE AND SEVERAL COMPONENTS HEAVILY CORRODED." NO PATIENT/ USER INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21406894 ↗
  • MalfunctionDec 11, 2024Device Contamination with Body Fluid; Device Contamination with Body Fluid

    DURING PM SERVICE BY A THIRD PARTY, THE MIXER WAS DISCOVERED TO BE CONTAMINATED WITH BLOOD ON THE INSIDE AND THEREFORE DETERMINED TO BE DEFECTIVE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20904359 ↗
  • MalfunctionOct 30, 2024No Audible Alarm; No Audible Alarm

    CUSTOMER REPORTED, "GAS SHORTAGE ALARM NOT WORKING. AS THE ALARM IS NOT WORKING, DOES THIS MEAN THERE IS NO AUDIBLE SIGNAL? IF SO, COULD IT BE THAT THE FOLLOWING OCCURRED AND THEREFORE THE ALARM WAS MISSING: WARNING: THE GAS MIXER AUDIBLE ALARM MAY NOT FUNCTION WHEN BOTH AIR AND OXYGEN INLET PRESSURES ARE LESS THAN THE MINIMUM SPECIFIED INLET PRESSURE OF 30 PSIG."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20571130 ↗

Other Mixer, Breathing Gases, Anesthesia Inhalation devices

All Mixer, Breathing Gases, Anesthesia Inhalation devices →

Sources and updates