Bird Blender 03920ARecalls & reported problems

Carefusion Corporation · MICBLNDR,W/O HOSES,DOM,LO · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10846446002355), received between Nov 19, 2019 and May 23, 2024 — classified as 1 injury, 16 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem12
  • Contamination1
  • Failure to Align1
  • High Readings1
  • Leak/Splash1
  • Output above Specifications1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • No Information2
  • No Patient Involvement1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 23, 2024High Readings; High Readings

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE BLENDER WAS CONSTANTLY DELIVERING 100% FIO2. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19382864 ↗
  • MalfunctionJan 23, 2024Failure to Align; Failure to Align

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE BIRD¿ LOW FLOW AIR/O2BLENDER HAS A PROBLEM WITH THE KNOB (HAS NOT LIMIT) AND NEEDS OVERHAUL. AS OF THIS TIME, THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18567719 ↗
  • MalfunctionOct 27, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE MICBLNDR HAS AN ISSUE WITH FIO2 (FRACTION OF INSPIRED OXYGEN) BEING INACCURATE. CUSTOMER DID NOT SPECIFY HOW INACCURATE OR AT WHAT SET POINTS. FURTHERMORE, THERE WAS NO INFORMATION REGARDING PATIENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18026382 ↗
  • MalfunctionMar 10, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE KNOB OF THE BLENDER HAS DAMAGED DURING PATIENT USE. FURTHERMORE, THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16524942 ↗
  • MalfunctionFeb 27, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE CONTROL KNOB OF THE BLENDER WAS BROKEN PRIOR TO PATIENT USE FROM WHERE IT SHOULD BE IN RELATION TO THE SET THE FIO2 (FRACTION OF INSPIRED OXYGEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16448963 ↗
  • MalfunctionJan 11, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL PROBLEM WITH BLENDER WHERE IT IS NOT PULLING O2 UNTIL IT IS SET OVER 25%. FURTHERMORE, THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16135687 ↗
  • MalfunctionDec 14, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT 'LOW SUPPLY' ALARM DOES NOT COME ON AND SLIDER ON THE AIR SIDE WAS NOT MOVING ON THE MICROBLENDER. THERE IS NO INFORMATION OF PATIENT INVOLVEMENT ASSOCIATED WITH THE EVENT AS OF THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15979444 ↗
  • MalfunctionOct 12, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE MICROBLENDER IS LEAKING MORE THAN USUAL FROM THE BLEED FLOW. AT THIS TIME THERE WAS NO INFORMATION OF PATIENT INVOLVEMENT REPORTED FROM THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15589241 ↗
  • MalfunctionDec 16, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL BLENDER IS READING HIGH AND OUT OF SPEC AT A DIAL SETTING OF 21%. THE REPORTER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13023359 ↗
  • MalfunctionDec 16, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL BLENDER IS READING HIGH AND OUT OF SPEC AT A DIAL SETTING OF 21%. THE REPORTER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13023350 ↗
  • MalfunctionDec 16, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LOW FLOW MICROBLENDER VENTILATOR IS READING HIGH AND OUT OF SPECIFICATION AT A DIAL SETTING OF 21%. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13023279 ↗
  • MalfunctionDec 15, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LOW FLOW MICROBLENDER VENTILATOR IS READING HIGH AND OUT OF SPECIFICATION AT A DIAL SETTING OF 21%. AS OF THIS TIME THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13014770 ↗
  • MalfunctionDec 15, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LOW FLOW MICROBLENDER VENTILATOR IS READING HIGH AND OUT OF SPECIFICATION AT A DIAL SETTING OF 21%. AS OF THIS TIME THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13014743 ↗
  • MalfunctionMar 3, 2021Contamination; Contamination

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LOW FLOW MICROBLENDER HAS OXYGEN CONTAMINATION OF MEDICAL AIR. AS OF THIS TIME THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11405317 ↗
  • MalfunctionApr 1, 2020Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED THAT THE 3920 LOW FLOW MICRO BLENDER SERIES FIO2 IS OUT OF SPECS. AS OF THIS TIME THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT IN THIS REPORTED EVENT.

    Patient problems: No Information

    MAUDE report 9909035 ↗
  • MalfunctionJan 17, 2020Output above Specifications; Output above Specifications

    THE CUSTOMER REPORTED THE BIRD BLENDER WAS SET TO TWENTY-ONE PERCENT (21%) OF FRACTION OF INSPIRED OXYGEN (FIO2). THE CUSTOMER REPORTED UPON FURTHER INVESTIGATION, THE DEVICE WAS GIVING OFF 87.5% FIO2.

    Patient problems: No Information

    MAUDE report 9602610 ↗
  • MalfunctionNov 19, 2019Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE THAT THE MICROBLENDER FAILED TO ALARM WHEN ONE GAS WAS LOST. THE CUSTOMER ADVISED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9343467 ↗

Other Mixer, Breathing Gases, Anesthesia Inhalation devices

All Mixer, Breathing Gases, Anesthesia Inhalation devices →