MAGEC® Spinal Bracing and Distraction System RA002-4545SLRRecalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Spinal bracing and distraction system, 4.5 mm offset rod with 90 mm actuator · Class 2
All models of MAGEC Spinal Bracing and Distraction System (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This NuVasive MAGEC growing rod is a magnetically driven spinal implant that lengthens the spine over time without repeat surgery. Its 60 reports mostly describe the rod refusing to lengthen or shorten, getting stuck, or showing wear, corrosion, a broken magnet, and debris inside the housing when removed. Reports describe 20 injuries and no deaths, with 40 malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 60
- Deaths reported
- 0
- Injuries reported
- 20
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 60 adverse event reports that list this exact device (DI 00856719002138), received between Feb 20, 2020 and Jul 26, 2024 — classified as 20 injuries, 40 malfunctions.
The most frequently reported specific device problem is “Defective Device” (29 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device29
- Corroded14
- Failure to Advance12
- Mechanical Jam12
- Break9
- Mechanical Problem8
- Physical Resistance/Sticking7
- Degraded5
Patient problems
- Failure of Implant56
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 60 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2024Break; Degraded; Material Discolored; Failure to Advance; Physical Resistance/Sticking; Break; Degraded; Material Discolored; Failure to Advance; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT DISTRACT OR RETRACT AND FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS WEAR, SCRATCHING, DISCOLORATION, AND MECHANICAL DAMAGE. UPON DISASSEMBLY OF THE ROD, DEBRIS WAS FOUND IN THE HOUSING TUBE AND PRESENCE OF A PIN FRACTURE THERE WAS ALSO A BROKEN MAGNET, COLD-WELDING BETWEEN THE HOUSING TUBE AND THE DISTRACTION ROD, AND THE DISTRACTION ROD WAS STUCK IN THE HOUSING TUBE. IT IS UNKNOWN WHY THE ROD WAS REMOVED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 19839660 ↗MalfunctionOct 3, 2023Failure to Advance; Failure to Advance
NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17862033 ↗MalfunctionAug 3, 2023Failure to Advance; Failure to Advance
INFORMATION WAS RECEIVED THAT DURING X-RAY IT WAS DISCOVERED THAT THE ROD DID NOT DISTRACT AT ALL. THE ROD WAS REMOVED AND REPLACED WITH A NEW ROD.
Patient problems: Failure of Implant
MAUDE report 17451651 ↗MalfunctionJun 16, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT FORCE TESTING WAS UNABLE TO BE COMPLETED DUE TO THE EXTERNAL REMOTE CONTROLLER (ERC) NOT BEING FUNCTIONAL. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17148911 ↗MalfunctionJun 16, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT THE ROD WOULD NOT EXTEND AND FORCE TESTING WAS UNABLE TO BE COMPLETED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17148906 ↗MalfunctionJun 16, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT FORCE TESTING WAS UNABLE TO BE COMPLETED DUE TO THE EXTERNAL REMOTE CONTROLLER (ERC) NOT FUNCTIONING. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17148838 ↗MalfunctionJun 16, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT THE ROD WOULD NOT EXTEND AND FORCE TESTING WAS UNABLE TO BE COMPLETED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS CORROSION ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17144901 ↗MalfunctionMay 31, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT DISTRACT AND FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS CORROSION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17035179 ↗MalfunctionMay 31, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, FORCE TESTING WAS UNABLE TO BE PERFORMED AS THE ELECTRONIC REMOTE CONTROLLER (ERC) WAS NOT FUNCTIONAL. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS CORROSION AND GALLING. IT WAS REPORTED THAT THE ROD WAS REMOVED DUE TO THE ROD NOT DISTRACTING AS THE PATIENT WAS OBESE. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17033395 ↗MalfunctionMay 31, 2023Break; Corroded; Mechanical Jam; Break; Corroded; Mechanical Jam
NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17031873 ↗MalfunctionMay 26, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17014607 ↗MalfunctionMay 26, 2023Break; Corroded; Mechanical Jam; Break; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION. IT WAS REPORTED THAT THE ROD WAS REMOVED DUE TO A BROKEN PIN AND HIGH ENERGY X-RAY EVALUATION REVEAL EVIDENCE OF A PIN FRACTURE. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17012002 ↗MalfunctionMay 26, 2023Corroded; Migration or Expulsion of Device; Mechanical Jam; Corroded; Migration or Expulsion of Device; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT EXTEND AND FORCE TESTING WAS UNABLE TO BE COMPLETED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED DUE TO DISTAL PULLOUT; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17011334 ↗MalfunctionFeb 15, 2023Failure to Advance; Material Protrusion/Extrusion; Physical Resistance/Sticking; Failure to Advance; Material Protrusion/Extrusion; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD EXERTED NO FORCE DURING TESTING. ADDITIONALLY, THE ROD WAS SEIZED TOGETHER AND WAS IMPOSSIBLE TO DISASSEMBLE AND A PORTION OF THE MAGNET WAS PROTRUDING FROM THE OUTER CASING. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 16379878 ↗MalfunctionFeb 10, 2023Break; Degraded; Mechanical Problem; Structural Problem; Failure to Advance; Break; Degraded; Mechanical Problem; Structural Problem; Failure to Advance
NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 16358858 ↗MalfunctionFeb 10, 2023Degraded; Structural Problem; Failure to Advance; Physical Resistance/Sticking; Degraded; Structural Problem; Failure to Advance; Physical Resistance/Sticking
MalfunctionFeb 10, 2023Break; Degraded; Structural Problem; Failure to Advance; Physical Resistance/Sticking; Break; Degraded; Structural Problem; Failure to Advance; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION OF THE ROD, THE RETRIEVAL CENTER FOUND THE ROD EXERTED NO FORCE DURING FORCE TESTING. ADDITIONALLY, THE ROD WAS REPORTED TO HAVE BEEN PISTONING AND THE DRIVE PIN WAS BROKEN. THE O-RING SEAL WAS REPORTED TO HAVE BEEN VERY BADLY DAMAGED, MISSHAPEN, AND COPIOUS BLACK DEBRIS WAS NOTED. THE RADIAL BEARING WAS FOUND TO BE BROKEN AND SEIZED TO THE MAGNET. THE LEAD SCREW WAS ALSO SEIZED INSIDE THE EXTENDING BAR. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 16358698 ↗MalfunctionFeb 10, 2023Break; Corroded; Degraded; Structural Problem; Failure to Advance; Detachment of Device or Device Component; Material Protrusion/Extrusion; Material Twisted/Bent; Physical Resistance/Sticking; Break; Corroded; Degraded; Structural Problem; Failure to Advance; Detachment of Device or Device Component; Material Protrusion/Extrusion; Material Twisted/Bent; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION OF THE ROD, THE RETRIEVAL CENTER FOUND THE ROD EXERTED NO FORCE DURING FORCE TESTING. ADDITIONALLY, THE EXTENDING BAR WAS REPORTED TO NOT BE PARALLEL WITH THE OUTER CASING. THE ROD WAS REPORTEDLY POSITIONING, INDICATING THAT THE DRIVE PIN WAS FRACTURED. UPON INITIAL DISASSEMBLY, THE RACE AND SOME OF THE ROLLING ELEMENTS WERE FOUND TO BE COVERED WITH SOME RUST-COLORED DEBRIS. FURTHER DISASSEMBLY REVEALED THAT THE MAGNET WAS DETACHED FROM THE LEAD SCREW AND THE DRIVE PIN WAS IN FACT SNAPPED. THE LEAD SCREW WAS PROTRUDING FROM THE OUTER CASING. IT WAS ADDITIONALLY REPORTED THAT THE LEAD SCREW WAS SEIZED WITHIN THE EXTENDING BAR AND THE O-RING HAD SNAPPED.
Patient problems: Failure of Implant
MAUDE report 16356384 ↗MalfunctionFeb 10, 2023Structural Problem; Failure to Advance; Physical Resistance/Sticking; Structural Problem; Failure to Advance; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION OF THE ROD, THE RETRIEVAL CENTER FOUND THE ROD EXERTED 35.3LBF DURING FORCE TESTING WHICH DID NOT MEET THE SPECIFICATION. ADDITIONALLY, UPON INITIAL DISASSEMBLY THE MAGNET, RADIAL BEARING, AND EXTENDING BAR WERE FOUND TO BE SEIZED TOGETHER AND THE COMPONENTS WERE UNABLE TO BE SEPARATED, THEREFORE, THE DRIVE PIN WAS UNABLE TO BE EXAMINED. IT WAS ADDITIONALLY REPORTED THAT THE O-RING SEAL WAS FOUND TO HAVE BEEN SNAPPED IN THREE PIECES. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 16353731 ↗MalfunctionOct 29, 2022Break; Corroded; Failure to Advance; Mechanical Jam; Break; Corroded; Failure to Advance; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT DISTRACT AND FORCE TESTING WAS UNABLE TO BE PERFORMED. DEVICE X-RAYS TAKEN DURING THE EVALUATION REVEALED CLEAR EVIDENCE OF PIN FRACTURE. ADDITIONALLY, IT WAS REPORTED THAT DURING SURFACE DAMAGE GRADING THERE WAS SOME GALLING AND CORROSION OBSERVED AROUND THE TELESCOPIC ROD. UPON DISASSEMBLY OF THE ROD, THE PIN FRACTURE WAS CONFIRMED AND THERE WAS DEBRIS FOUND IN THE CONSTRUCT WITHIN THE PIN HOUSING. THE REASON FOR REMOVAL WAS REPORTEDLY DUE TO THE PIN FRACTURE. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15695253 ↗MalfunctionOct 27, 2022Corroded; Failure to Advance; Mechanical Jam; Corroded; Failure to Advance; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT DISTRACT AND FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION AROUND THE TELESCOPIC ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 2 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15685380 ↗InjuryJun 8, 2022Corroded; Fracture; Naturally Worn; Scratched Material; Corroded; Fracture; Naturally Worn; Scratched Material
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD EXPERIENCING A PIN FRACTURE. DURING EXAMINATION OF THE ROD, THE RETRIEVAL CENTER FOUND CORROSION AND DISCOLORATION ON THE ROD. ADDITIONALLY, WEAR, SCRATCHES, AND OTHER SURFACE DAMAGE WAS NOTED. THERE WAS NO FURTHER REPORTED PATIENT IMPACT AND NO PATIENT OR USER INJURY WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 2 OF 2.
Patient problems: Failure of Implant
MAUDE report 14629723 ↗MalfunctionFeb 8, 2022Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED THAT AFTER FIVE YEARS OF BEING IMPLANTED, A REVISION PROCEDURE WAS PERFORMED DUE TO THE RODS ONLY DISTRACTING 3 CENTIMETERS. AFTER REMOVAL, THE ROD WOULD NOT DISTRACT WHILE UTILIZING MANUAL DISTRACTION, HOWEVER, A "CLICK" COULD BE HEARD. THERE WAS NO PATIENT INJURY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant
MAUDE report 13481564 ↗MalfunctionJul 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE ROD HAD EVIDENCE OF GALLING AND CORROSION. ADDITIONALLY IT WAS REPORTED THAT THE ROD WAS UNABLE TO BE RETRACTED.
Patient problems: Insufficient Information
MAUDE report 12144829 ↗MalfunctionJul 7, 2021Mechanical Problem; Mechanical Problem
Other Growing Rod System- Magnetic Actuation devices
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation112 reports
- MAGEC MC2-5090SNuvasive, Inc. · Growing rod system- magnetic actuation1 recall33 reports
- MAGEC® Spinal Bracing and Distraction System RA002-4545SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation82 reports
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLRNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation81 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall26 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall25 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page