Tenex Health TX® 554-2003-001Recalls & reported problems
Tenex Health, Inc. · TX2 Procedure Pack · Class U
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Tenex Health TX ultrasonic surgical device has 213 adverse event reports, with no deaths and 14 injuries. Most describe a portion of the microtip needle separating from the handpiece during procedures, and some say the piece was retrieved or checked by X-ray or ultrasound. Several recent reports mention no patient complications, and reports have fallen over the past year. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 213
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 213 adverse event reports that list this exact device (DI 00857156006154), received between Aug 20, 2020 and Sep 24, 2026 — classified as 14 injuries, 199 malfunctions.
The most frequently reported specific device problem is “Fracture” (203 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture203
- Adverse Event Without Identified Device or Use Problem3
- Appropriate Device Problem Term/Code Not Available3
- Overheating of Device2
- Insufficient Device Problem Information1
- Material Fragmentation1
- Material Frayed1
- Material Too Soft/Flexible1
Patient problems
- No Clinical Signs, Symptoms or Conditions200
- Pain7
- Foreign Body In Patient3
- Swelling/ Edema3
- Device Embedded In Tissue or Plaque2
- Hematoma1
- Impaired Healing1
- Inadequate Pain Relief1
Recent adverse event reports
The 25 most recent of 213 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50892216 ↗MalfunctionMay 13, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. 2 ADDITIONAL HANDPIECES WERE USED TO COMPLETE THE PROCEDURE, BOTH REMAINED INTACT. AN X-RAY WAS PERFORMED AND CONFIRMED NOTHING WAS RETAINED IN THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25166513 ↗MalfunctionMar 14, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. THE PIECE WAS RETRIEVED FROM THE PATIENT, AND AN ULTRASOUND CONFIRMED NO FOREIGN BODY WAS LEFT BEHIND. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24599721 ↗MalfunctionFeb 26, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, THERE WERE 2 INSTANCES OF A PORTION OF THE MICROTIP NEEDLE SEPARATING FROM THE REST OF THE HANDPIECE. ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24464658 ↗MalfunctionFeb 20, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24408191 ↗MalfunctionFeb 20, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24408187 ↗MalfunctionFeb 13, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. ULTRASOUND CONFIRMED NOTHING WAS RETAINED IN THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24339921 ↗MalfunctionFeb 13, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24339838 ↗MalfunctionFeb 13, 2026Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. A MANUAL TENOTOMY WITH A 18G SPINAL NEEDLE WAS THEN COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24339826 ↗InjuryFeb 5, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IN THE ARTICLE "MINIMALLY INVASIVE ULTRASOUND-GUIDED PERCUTANEOUS PLANTAR FASCIOTOMY ON CHRONIC PLANTAR FASCIITIS: A RETROSPECTIVE ANALYSIS" BY ENCINAS ET AL., 2 CASES OF DELAYED WOUND HEALING AND 2 CASES OF BAXTER'S NERVE ENTRAPMENT WERE REPORTED IN PATIENTS WHO UNDERWENT A PLANTAR FASCIITIS PROCEDURE WITH THE TX SYSTEM. ALL 4 ADVERSE EVENTS REQUIRED ADDITIONAL INTERVENTION TO TREAT. PATIENTS WITH DELAYED WOUND HEALING WERE TREATED WITH LOCAL WOUND CARE AND ORAL ANTIBIOTICS. PATIENTS WITH BAXTER'S NERVE ENTRAPMENT UNDERWENT BAXTER'S NERVE DECOMPRESSION, WHICH WAS PERFORMED MORE THAN A YEAR AFTER THE INITIAL TX SYSTEM PROCEDURE. THESE ADVERSE EVENTS ARE SUMMARIZED IN THIS FORM 3500A.
Patient problems: Impaired Healing; Neuropathy
MAUDE report 24273614 ↗MalfunctionDec 18, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23846117 ↗MalfunctionDec 18, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23846116 ↗MalfunctionDec 9, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. A DIAGNOSTIC ULTRASOUND WAS PERFORMED AND CONFIRMED THAT NOTHING WAS RETAINED IN THE PATIENT. ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23759097 ↗MalfunctionNov 21, 2025Fracture; Fracture
DURING A PODIATRY PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE WHEN THE HANDPIECE WAS REMOVED FROM THE PATIENT. AN X-RAY WAS PERFORMED AND CONFIRMED THAT NOTHING WAS RETAINED IN THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23619175 ↗MalfunctionOct 28, 2025Overheating of Device; Overheating of Device
SHORTLY AFTER THE START OF A PROCEDURE WITH THE TENEX SYSTEM, THE SURGEON REPORTED FEELING THE HANDPIECE OVERHEAT IN THEIR HAND. ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23407761 ↗MalfunctionOct 22, 2025Fracture; Fracture
AT THE START OF A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE AND ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23360781 ↗MalfunctionOct 17, 2025Fracture; Fracture
TOWARDS THE END OF A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23330606 ↗MalfunctionOct 17, 2025Fracture; Fracture
TOWARDS THE END OF A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23330565 ↗MalfunctionSep 16, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE AND ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23064459 ↗MalfunctionSep 4, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22973635 ↗MalfunctionAug 25, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE AND ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22890405 ↗MalfunctionAug 20, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE AND ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22856699 ↗MalfunctionAug 18, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE AND ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22833070 ↗MalfunctionAug 11, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE AND ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22772418 ↗MalfunctionAug 4, 2025Fracture; Fracture
DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HANDPIECE AND ANOTHER HANDPIECE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22710064 ↗
Other Instrument, Ultrasonic Surgical devices
- Harmonic ACE HARH36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical4 deaths7,609 reports
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,037 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,726 reports
- Harmonic HAR1136Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2 deaths1,538 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,373 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death970 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical935 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page