Skip to content

ThermaCor 1200 Infusion System PNC-1200Recalls & reported problems

Smisson-Cartledge Biomedical, LLC · Cassette with Patient Line · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00857893006066), received between Sep 22, 2020 and Sep 22, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

1 FDA recall record name this device's identifier.

2 other recalls cover devices cleared under the same FDA submission (K052055, K202461); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 18, 2021Class ITerminated

    Customers are being notified of results from a Toxicological Risk Assessment related to potential for aluminum ions to leach into warmed fluids.

    Z-1260-2021 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2020Fluid/Blood Leak; Fluid/Blood Leak

    DURING SURGERY AT (B)(6) MEMORIAL HOSPITAL, THE STAFF NOTICED THE CASSETTE STARTED LEAKING BLOOD AT THE PATIENT LINE CONNECTION TO THE CASSETTE. THE CASSETTE LOT NUMBER WAS LOT 316277. THE HOSPITAL CHANGED OVER TO A LEVEL ONE UNIT TO COMPLETE THE SURGERY. NO PATIENT HARM WAS NOTED BY THE HOSPITAL. THE TUBING WAS RETURNED TO THE DISTRIBUTOR, MEDICAL SOLUTIONS, INC (MSI). MSI FOUND NO ADHESIVE AT THE CASSETTE END OF THE PATIENT LINE WHICH CAUSED THE LEAK AT THE FITTING. RETAINS WERE REVIEWED, AND ADHESIVE WAS PRESENT AT THE CONNECTION. NO OTHER ISSUES HAVE BEEN NOTED FOR THIS LOT. THE DHR WAS REVIEWED; NO ISSUE WAS NOTICED FOR THIS LOT. THE MANUFACTURER HAS BEEN NOTIFIED OF THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10568031 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K052055, K202461. One submission often covers many models, so check the product and lot details in each record.

  • Z-1258-2021 · Feb 18, 2021 · Terminated
    ThermaCor 1200 Disposable Sets Components of the Smisson-Cartledge Biomedical ThermaCor 1200 Rapid Thermal Infusion System, used for infusion of: -crystalloid, colloid, or blood p
  • Z-1259-2021 · Feb 18, 2021 · Terminated
    ThermaCor 1200 Disposable Sets Components of the Smisson-Cartledge Biomedical ThermaCor 1200 Rapid Thermal Infusion System, used for infusion of: -crystalloid, colloid, or blood

Other Pump, infusion devices

All Pump, infusion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page