Skip to content

MOSAIQ (UDI 00858164002336)Recalls & reported problems

Elekta Solutions AB · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00858164002336), received between Nov 26, 2020 and Jan 9, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (12 reports).

1 FDA recall record name this device's identifier.

1 other recall cover devices cleared under the same FDA submission (K183034); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 20, 2021Terminated

    Malware attack

    Z-1783-2021 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem12
  • Communication or Transmission Problem1

Patient problems

  • Insufficient Information10
  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2025Computer Software Problem; Computer Software Problem

    THE CUSTOMER REQUESTED HELP ON HOW TO HANDLE LAG BETWEEN SESSIONS AND DOSE ACCUMULATION IN MOSAIQ MONITORING.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21110583 ↗
  • MalfunctionJul 12, 2024Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT POSITIONS OF THE BLADES BETWEEN THE MOSAIQ AND MONACO DRRS ARE NOT IDENTICAL.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19729041 ↗
  • MalfunctionApr 26, 2024Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT THE TREATMENT COUNT IS INCORRECT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19190839 ↗
  • MalfunctionJan 12, 2024Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT THE IMAGES RECORDED/STORED IN MOSAIQ ARE NOT RELATED TO REFERENCE COMPUTERISED TOMOGRAPHY (CT)/ CONE BEAM COMPUTED TOMOGRAPHY (CBCT).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18501261 ↗
  • MalfunctionSep 7, 2023Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT OFFSETS WERE NOT SENT TO THE ACCELERATOR AFTER MATCHING.

    Patient problems: Insufficient Information

    MAUDE report 17697098 ↗
  • MalfunctionJul 14, 2023Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED AN ISSUE WHEN EXCEEDING TABLE TARGET CALCULATED FROM CBCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17324122 ↗
  • MalfunctionMay 5, 2023Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT THE IMAGES RECORDED/STORED IN MOSAIQ ARE NOT RELATED TO REFERENCE COMPUTERISED TOMOGRAPHY (CT)/ CONE BEAM COMPUTED TOMOGRAPHY (CBCT).

    Patient problems: Insufficient Information

    MAUDE report 16874663 ↗
  • MalfunctionApr 27, 2023Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT FIELD A1.1 WAS TREATED OUTSIDE OF THE TOLERANCE POSITION. THIS DID NOT PROMPT THE USER TO OVERRIDE THE COUCH POSITION AND NO INDICATOR WAS SHOWN THAT THIS WAS TREATED OUTSIDE OF THE 1.0 TOLERANCE IN MOSAIQ.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16822098 ↗
  • MalfunctionApr 6, 2023Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT DURING TREATMENT THERE WAS A BEAM TIMER ERROR. THE BMDM WAS RESET AND THE MODALITY BECAME UNKNOWN.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16688271 ↗
  • MalfunctionJun 10, 2022Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED MOSAIQ POTENTIAL TREATMENT WITH HEXAPOD IN THE WRONG POSITION.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14654359 ↗
  • MalfunctionNov 19, 2021Computer Software Problem; Communication or Transmission Problem; Computer Software Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THE SEQ DISCONNECTED FROM LINAC DURING TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12842930 ↗
  • MalfunctionNov 26, 2020Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT QUICK NOTES DISPLAY STILL SHOWS STATUS AS, "VOIDED NOTES AS ACTIVE."

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10911007 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K183034. One submission often covers many models, so check the product and lot details in each record.

  • Z-2591-2021 · Sep 2, 2021 · Open, Classified
    MOSAIQ Oncology Information System

Other Accelerator, Linear, Medical devices

All Accelerator, Linear, Medical devices →

Sources and updates