NOVADAQ and PINPOINT PC9000Recalls & reported problems
Novadaq Technologies Inc · Video Processor/Illuminator (VPI) · Class 2
All models of NOVADAQ and PINPOINT (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00858701006049), received between Feb 26, 2021 and Aug 31, 2026 — classified as 2 injuries, 15 malfunctions.
The most frequently reported specific device problem is “No Device Output” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Device Output4
- Temperature Problem4
- No Display/Image3
- Poor Quality Image3
- Electrostatic Discharge2
- Failure to Power Up2
- Image Display Error/Artifact2
- Output Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- Electric Shock3
- Insufficient Information3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Shock1
- Undesired Nerve Stimulation1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
InjuryAug 31, 2026No Display/Image; Poor Quality Image
It was reported that the customer had an issue with setting up the picture sharpness and the picture was vanishing during the procedure, that was probably caused due to a camera cable connection problem. The medical procedure was converted to a conventional laparotomy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50534175 ↗InjuryAug 28, 2026No Display/Image; Poor Quality Image
It was reported that the customer had an issue with setting up the picture sharpness and the picture was vanishing during the procedure, that was probably caused due to a camera cable connection problem. The medical procedure was converted to a conventional laparotomy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50507209 ↗MalfunctionApr 22, 2026No Device Output; Overheating of Device
IT WAS REPORTED THAT LOUD POP NOISE DURING CASE AND IT STOPPED WORKING AND SMELT LIKE SOMETHING WAS BURNING. MY UNDERSTANDING IS THAT IT CAME FROM THE VPI. THEREFORE, THERE WAS A THERMAL EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24957016 ↗MalfunctionJan 17, 2024No Device Output; Temperature Problem; No Device Output; Temperature Problem
IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18530806 ↗MalfunctionJan 4, 2023Electrostatic Discharge; Electrostatic Discharge
IT WAS REPORTED THAT THE USER RECEIVED AN ELECTRICAL SHOCK.
Patient problems: Electric Shock; Undesired Nerve Stimulation
MAUDE report 16098456 ↗MalfunctionOct 31, 2022Image Display Error/Artifact; Optical Problem; Image Display Error/Artifact; Optical Problem
IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 15700410 ↗MalfunctionSep 24, 2022Image Display Error/Artifact; Temperature Problem; Image Display Error/Artifact; Temperature Problem
IT WAS REPORTED THAT THE LIGHT LEAD WAS VERY HOT, INDICATING A THERMAL EVENT. THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15478593 ↗MalfunctionAug 25, 2022Failure to Power Up; No Display/Image; Failure to Power Up
IT WAS REPORTED THAT THE DEVICE LOST IMAGE DURING A PROCEDURE AND COULD NOT BE POWERED BACK ON. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED ADVERSE CONSEQUENCES AS A RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15297972 ↗MalfunctionAug 18, 2022Failure to Power Up; Unexpected Shutdown
IT WAS REPORTED THAT THE DEVICE LOST IMAGE AND POWERED OFF DURING A PROCEDURE. THE USERS WERE UNABLE TO GET THE DEVICE TO POWER BACK ON, WHICH RESULTED IN A 15 MINUTE DELAY WHILE A REPLACEMENT WAS BROUGHT IN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WAS NO REPORT OF ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15252529 ↗MalfunctionMay 25, 2022Use of Device Problem; Temperature Problem
IT WAS REPORTED THAT THE HANDHELD DEVICE WAS LEFT ON A PATIENT IN ERROR. AS A RESULT, THE LIGHT FROM THE VIDEO PROCESSOR ILLUMINATOR CAUSED A SMALL HOLE TO BE BURNED INTO THE PATIENT DRAPE. THERE WAS NO ADVERSE CONSEQUENCES TO THE PATIENT AS A RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14496622 ↗MalfunctionNov 30, 2021Electrostatic Discharge; Unintended Electrical Shock; Electrostatic Discharge; Unintended Electrical Shock
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED ELECTRIC SHOCK.
Patient problems: Electric Shock
MAUDE report 12896329 ↗MalfunctionNov 10, 2021No Device Output; Unintended Electrical Shock; No Device Output; Unintended Electrical Shock
IT WAS REPORTED THAT THE USER EXPERIENCED ELECTRIC SHOCK.
Patient problems: Electric Shock; Shock
MAUDE report 12789401 ↗MalfunctionAug 24, 2021Thermal Decomposition of Device; Temperature Problem; Temperature Problem
IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12363416 ↗MalfunctionAug 19, 2021Output Problem; Output Problem
IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE DURING PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12345955 ↗MalfunctionJun 30, 2021Overheating of Device
IT WAS REPORTED THAT THE DEVICE OVERHEATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12093067 ↗MalfunctionApr 22, 2021No Device Output; Adverse Event Without Identified Device or Use Problem; No Device Output
IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 11710891 ↗MalfunctionFeb 26, 2021Poor Quality Image; Output Problem; Poor Quality Image; Output Problem
IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11387790 ↗
Other System, X-Ray, Angiographic devices
- ARTIS icono ceiling 11328100Siemens Healthcare GmbH · System, x-ray, angiographic9 recalls1 report
- Artis Q ceiling 10848281Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls1 death49 reports
- Artis Q floor 10848280Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls18 reports
- Artis zee biplane 10094141Siemens Healthcare GmbH · System, x-ray, angiographic3 recalls63 reports
- Artis Q.zen ceiling 10848354Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls10 reports
- Artis Q.zen floor 10848353Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls5 reports
- Artis zee floor MN 10094142Siemens Healthcare GmbH · System, x-ray, angiographic2 recalls0 reports
- iMRX System iMRX System AR-DR MR30RSIMRIS Imaging, Inc. · System, X-Ray, Angiographic1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page