SENSINEL™ CPM WEARABLE ADCP1100-WB-001Recalls & reported problems

Analog Devices, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00860006382322), received between Oct 16, 2025 and Dec 12, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 6 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component5
  • Insufficient Device Problem Information3

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING ROUTINE DATA MONITORING ON (B)(6) 2025, THE TEAM NOTICED AN ERRATIC PATTERN IN SOME MEASUREMENT SIGNALS. ON (B)(6) 2025, THE TEAM CONFIRMED INACCURATE READINGS IN 9 OUT OF 40 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (22.5%). TWO SIGNAL PATTERNS WERE OBSERVED: (1) A SHIFT IN THE ACOUSTIC SIGNAL AWAY FROM THE EXPECTED BASELINE AND (2) ABRUPT SIMULTANEOUS JUMPS/FALLS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE ANOMALIES DID NOT ALIGN WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, SUGGESTING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23789738 ↗
  • MalfunctionDec 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ON (B)(6) 2025, ANALOG DEVICES' TRIAGING TEAM ESCALATED THE QUALITY OF THIS PATIENT'S MEASUREMENT DATA. AS PART OF THE COMPLAINT INVESTIGATION, THE TEAM THEN NOTICED AN ERRATIC PATTERN IN SOME MEASUREMENT SIGNALS. ON (B)(6) 2025, THE TEAM CONFIRMED INACCURATE READINGS IN 4 OUT OF 35 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (11.43%). TWO SIGNAL PATTERNS WERE OBSERVED: (1) A SHIFT IN THE ACOUSTIC SIGNAL AWAY FROM THE EXPECTED BASELINE AND (2) ABRUPT SIMULTANEOUS JUMPS/FALLS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE ANOMALIES DID NOT ALIGN WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, SUGGESTING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23789675 ↗
  • MalfunctionOct 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING ROUTINE DATA MONITORING ON (B)(6) 2025, THE TEAM NOTICED AN ERRATIC PATTERN IN SOME MEASUREMENT SIGNALS. ON (B)(6) 2025, THE TEAM CONFIRMED INACCURATE READINGS IN 4 OUT OF 33 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (12.12%). TWO ABNORMAL SIGNAL PATTERNS WERE OBSERVED: (1) A SHIFT IN THE ACOUSTIC SIGNAL AWAY FROM THE EXPECTED BASELINE AND (2) ABRUPT SIMULTANEOUS JUMPS/FALLS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE ANOMALIES DID NOT ALIGN WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, SUGGESTING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23314406 ↗
  • MalfunctionOct 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING ROUTINE DATA MONITORING ON (B)(6) 2025, THE TEAM NOTICED AN ERRATIC PATTERN IN SOME MEASUREMENT SIGNALS. ON (B)(6) 2025, A RETROSPECTIVE REVIEW OF THE PATIENT DATA FOUND INACCURATE READINGS IN 5 OF 24 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (20.83%). TWO ABNORMAL SIGNAL PATTERNS WERE OBSERVED: (1) A SHIFT IN THE ACOUSTIC SIGNAL AWAY FROM ITS EXPECTED BASELINE AND (2) ABRUPT SIMULTANEOUS JUMPS/FALLS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE ANOMALIES DID NOT ALIGN WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, SUGGESTING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23314394 ↗
  • MalfunctionOct 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING ROUTINE DATA MONITORING ON (B)(6) 2025, THE TEAM NOTICED AN ERRATIC PATTERN IN SOME MEASUREMENT SIGNALS. ON (B)(6) 2025, THE TEAM CONFIRMED INACCURATE READINGS IN 4 OUT OF 18 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (22.22%). TWO ABNORMAL SIGNAL PATTERNS WERE OBSERVED: (1) A SHIFT IN THE ACOUSTIC SIGNAL AWAY FROM THE EXPECTED BASELINE AND (2) ABRUPT SIMULTANEOUS JUMPS/FALLS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE ANOMALIES DID NOT ALIGN WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, SUGGESTING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23314373 ↗
  • MalfunctionOct 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING ROUTINE MONITORING ON (B)(6) 2025, THE TEAM IDENTIFIED ERRATIC MEASUREMENT SIGNALS. ANALOG NOTIFIED THE PATIENT AND HEALTHCARE PROVIDER OF THE IDENTIFIED ISSUE AND ADVISED THE PATIENT TO DISCONTINUE USING THE DEVICE. ON (B)(6) 2025, A RETROSPECTIVE REVIEW OF THE PATIENT DATA FOUND INACCURATE READINGS IN 11 OF 42 RELATIVE TIDAL VOLUME MEASUREMENTS (26.19%) AND 41 OF 42 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (97.62%). TWO ABNORMAL PATTERNS WERE NOTED: (1) A SHIFT OF THE ACOUSTIC SIGNAL FROM ITS EXPECTED BASELINE AND (2) ABRUPT, SIMULTANEOUS JUMPS OR DROPS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE FINDINGS WERE INCONSISTENT WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, INDICATING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23313085 ↗
  • MalfunctionOct 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING ROUTINE DATA MONITORING ON (B)(6) 2025, THE TEAM NOTICED AN ERRATIC PATTERN IN SOME MEASUREMENT SIGNALS. ON (B)(6) 2025, THE TEAM CONFIRMED INACCURATE READINGS IN 26 OUT OF 35 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (74.29%). TWO ABNORMAL SIGNAL PATTERNS WERE OBSERVED: (1) A SHIFT IN THE ACOUSTIC SIGNAL AWAY FROM THE EXPECTED BASELINE AND (2) ABRUPT SIMULTANEOUS JUMPS/FALLS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE ANOMALIES DID NOT ALIGN WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, SUGGESTING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23312881 ↗
  • MalfunctionOct 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING ROUTINE DATA MONITORING ON (B)(6) 2025, THE TEAM NOTICED AN ERRATIC PATTERN IN SOME MEASUREMENT SIGNALS. THE TEAM CONFIRMED INACCURATE READINGS IN 1 OF 18 RELATIVE TIDAL VOLUME MEASUREMENTS (5.56%) AND 4 OF 9 DIASTOLIC HEART SOUND STRENGTH MEASUREMENTS (44.44%). TWO ABNORMAL SIGNAL PATTERNS WERE OBSERVED: (1) A SHIFT IN THE ACOUSTIC SIGNAL AWAY FROM ITS EXPECTED BASELINE AND (2) ABRUPT SIMULTANEOUS JUMPS/FALLS IN BOTH ACOUSTIC AND IMPEDANCE SIGNALS. THESE ANOMALIES DID NOT ALIGN WITH COMMON USER ERRORS OR POOR DEVICE CONTACT, SUGGESTING A POTENTIAL HARDWARE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23312798 ↗

Other Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) devices

All Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) devices →